Medical Devices Regulation

Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices, amending […]

32017R0745 · Health · Atom feed · on EUR-Lex · reflects the consolidated version of 2026-08-10

32017R074502017R0745-20170505

in force not stated · detected 2026-08-10

15 provisions touched — 15 substantive, 0 date-only, 15 disputed · 0 sentences quoted verbatim by the gate, 0 changes shipped without an explanation

MODIFIED Art. 2 — Definitions · applies from unchanged

Disputed — seen by the structural diff, not by corpus metadata.

In point (30), the definition of manufacturer changes the spelling of the term describing the name under which a device is marketed, from "trademark" to "trade mark". Art. 2, v1 Art. 2, v2

text before / after

32017R074502017R0745-20170505

Article 2DefinitionsFor the purposes of this Regulation, the following definitions apply:(1)medical device means any instrument, apparatus, appliance, software, implant, reagent, material or other article intended by the manufacturer to be used, alone or in combination, for human beings for one or more of the following specific medical purposes:diagnosis, prevention, monitoring, prediction, prognosis, treatment or alleviation of disease,diagnosis, monitoring, treatment, alleviation of, or compensation for, an injury or disability,investigation, replacement or modification of the anatomy or of a physiological or pathological process or state,providing information by means of in vitro examination of specimens derived from the human body, including organ, blood and tissue donations,and which does not achieve its principal intended action by pharmacological, immunological or metabolic means, in or on the human body, but which may be assisted in its function by such means.The following products shall also be deemed to be medical devices:devices for the control or support of conception;products specifically intended for the cleaning, disinfection or sterilisation of devices as referred to in Article 1(4) and of those referred to in the first paragraph of this point.(2)accessory for a medical device means an article which, whilst not being itself a medical device, is intended by its manufacturer to be used together with one or several particular medical device(s) to specifically enable the medical device(s) to be used in accordance with its/their intended purpose(s) or to specifically and directly assist the medical functionality of the medical device(s) in terms of its/their intended purpose(s);(3)custom-made device means any device specifically made in accordance with a written prescription of any person authorised by national law by virtue of that person's professional qualifications which gives, under that person's responsibility, specific design characteristics, and is intended for the sole use of a particular patient exclusively to meet their individual conditions and needs.However, mass-produced devices which need to be adapted to meet the specific requirements of any professional user and devices which are mass-produced by means of industrial manufacturing processes in accordance with the written prescriptions of any authorised person shall not be considered to be custom-made devices;(4)active device means any device, the operation of which depends on a source of energy other than that generated by the human body for that purpose, or by gravity, and which acts by changing the density of or converting that energy. Devices intended to transmit energy, substances or other elements between an active device and the patient, without any significant change, shall not be deemed to be active devices.Software shall also be deemed to be an active device;(5)implantable device means any device, including those that are partially or wholly absorbed, which is intended:to be totally introduced into the human body, orto replace an epithelial surface or the surface of the eye,by clinical intervention and which is intended to remain in place after the procedure.Any device intended to be partially introduced into the human body by clinical intervention and intended to remain in place after the procedure for at least 30 days shall also be deemed to be an implantable device;(6)invasive device means any device which, in whole or in part, penetrates inside the body, either through a body orifice or through the surface of the body;(7)generic device group means a set of devices having the same or similar intended purposes or a commonality of technology allowing them to be classified in a generic manner not reflecting specific characteristics;(8)single-use device means a device that is intended to be used on one individual during a single procedure;(9)falsified device means any device with a false presentation of its identity and/or of its source and/or its CE marking certificates or documents relating to CE marking procedures. This definition does not include unintentional non-compliance and is without prejudice to infringements of intellectual property rights;(10)procedure pack means a combination of products packaged together and placed on the market with the purpose of being used for a specific medical purpose;(11)system means a combination of products, either packaged together or not, which are intended to be inter-connected or combined to achieve a specific medical purpose;(12)intended purpose means the use for which a device is intended according to the data supplied by the manufacturer on the label, in the instructions for use or in promotional or sales materials or statements and as specified by the manufacturer in the clinical evaluation;(13)label means the written, printed or graphic information appearing either on the device itself, or on the packaging of each unit or on the packaging of multiple devices;(14)instructions for use means the information provided by the manufacturer to inform the user of a device's intended purpose and proper use and of any precautions to be taken;(15)Unique Device Identifier (UDI) means a series of numeric or alphanumeric characters that is created through internationally accepted device identification and coding standards and that allows unambiguous identification of specific devices on the market;(16)non-viable means having no potential for metabolism or multiplication;(17)derivative means a non-cellular substance extracted from human or animal tissue or cells through a manufacturing process. The final substance used for manufacturing of the device in this case does not contain any cells or tissues;(18)nanomaterial means a natural, incidental or manufactured material containing particles in an unbound state or as an aggregate or as an agglomerate and where, for 50 % or more of the particles in the number size distribution, one or more external dimensions is in the size range 1-100 nm;Fullerenes, graphene flakes and single-wall carbon nanotubes with one or more external dimensions below 1 nm shall also be deemed to be nanomaterials;(19)particle, for the purposes of the definition of nanomaterial in point (18), means a minute piece of matter with defined physical boundaries;(20)agglomerate, for the purposes of the definition of nanomaterial in point (18), means a collection of weakly bound particles or aggregates where the resulting external surface area is similar to the sum of the surface areas of the individual components;(21)aggregate, for the purposes of the definition of nanomaterial in point (18), means a particle comprising of strongly bound or fused particles;(22)performance means the ability of a device to achieve its intended purpose as stated by the manufacturer;(23)risk means the combination of the probability of occurrence of harm and the severity of that harm;(24)benefit-risk determination means the analysis of all assessments of benefit and risk of possible relevance for the use of the device for the intended purpose, when used in accordance with the intended purpose given by the manufacturer;(25)compatibility is the ability of a device, including software, when used together with one or more other devices in accordance with its intended purpose, to:(a)perform without losing or compromising the ability to perform as intended, and/or(b)integrate and/or operate without the need for modification or adaption of any part of the combined devices, and/or(c)be used together without conflict/interference or adverse reaction.(26)interoperability is the ability of two or more devices, including software, from the same manufacturer or from different manufacturers, to:(a)exchange information and use the information that has been exchanged for the correct execution of a specified function without changing the content of the data, and/or(b)communicate with each other, and/or(c)work together as intended.(27)making available on the market means any supply of a device, other than an investigational device, for distribution, consumption or use on the Union market in the course of a commercial activity, whether in return for payment or free of charge;(28)placing on the market means the first making available of a device, other than an investigational device, on the Union market;(29)putting into service means the stage at which a device, other than an investigational device, has been made available to the final user as being ready for use on the Union market for the first time for its intended purpose;(30)manufacturer means a natural or legal person who manufactures or fully refurbishes a device or has a device designed, manufactured or fully refurbished, and markets that device under its name or trademark;(31)fully trade mark;(31)fully refurbishing, for the purposes of the definition of manufacturer, means the complete rebuilding of a device already placed on the market or put into service, or the making of a new device from used devices, to bring it into conformity with this Regulation, combined with the assignment of a new lifetime to the refurbished device;(32)authorised representative means any natural or legal person established within the Union who has received and accepted a written mandate from a manufacturer, located outside the Union, to act on the manufacturer's behalf in relation to specified tasks with regard to the latter's obligations under this Regulation;(33)importer means any natural or legal person established within the Union that places a device from a third country on the Union market;(34)distributor means any natural or legal person in the supply chain, other than the manufacturer or the importer, that makes a device available on the market, up until the point of putting into service;(35)economic operator means a manufacturer, an authorised representative, an importer, a distributor or the person referred to in Article 22(1) and 22(3);(36)health institution means an organisation the primary purpose of which is the care or treatment of patients or the promotion of public health;(37)user means any healthcare professional or lay person who uses a device;(38)lay person means an individual who does not have formal education in a relevant field of healthcare or medical discipline;(39)reprocessing means a process carried out on a used device in order to allow its safe reuse including cleaning, disinfection, sterilisation and related procedures, as well as testing and restoring the technical and functional safety of the used device;(40)conformity assessment means the process demonstrating whether the requirements of this Regulation relating to a device have been fulfilled;(41)conformity assessment body means a body that performs third-party conformity assessment activities including calibration, testing, certification and inspection;(42)notified body means a conformity assessment body designated in accordance with this Regulation;(43)CE marking of conformity or CE marking means a marking by which a manufacturer indicates that a device is in conformity with the applicable requirements set out in this Regulation and other applicable Union harmonisation legislation providing for its affixing;(44)clinical evaluation means a systematic and planned process to continuously generate, collect, analyse and assess the clinical data pertaining to a device in order to verify the safety and performance, including clinical benefits, of the device when used as intended by the manufacturer;(45)clinical investigation means any systematic investigation involving one or more human subjects, undertaken to assess the safety or performance of a device;(46)investigational device means a device that is assessed in a clinical investigation;(47)clinical investigation plan means a document that describes the rationale, objectives, design, methodology, monitoring, statistical considerations, organisation and conduct of a clinical investigation;(48)clinical data means information concerning safety or performance that is generated from the use of a device and is sourced from the following:clinical investigation(s) of the device concerned,clinical investigation(s) or other studies reported in scientific literature, of a device for which equivalence to the device in question can be demonstrated,reports published in peer reviewed scientific literature on other clinical experience of either the device in question or a device for which equivalence to the device in question can be demonstrated,clinically relevant information coming from post-market surveillance, in particular the post-market clinical follow-up;(49)sponsor means any individual, company, institution or organisation which takes responsibility for the initiation, for the management and setting up of the financing of the clinical investigation;(50)subject means an individual who participates in a clinical investigation;(51)clinical evidence means clinical data and clinical evaluation results pertaining to a device of a sufficient amount and quality to allow a qualified assessment of whether the device is safe and achieves the intended clinical benefit(s), when used as intended by the manufacturer;(52)clinical performance means the ability of a device, resulting from any direct or indirect medical effects which stem from its technical or functional characteristics, including diagnostic characteristics, to achieve its intended purpose as claimed by the manufacturer, thereby leading to a clinical benefit for patients, when used as intended by the manufacturer;(53)clinical benefit means the positive impact of a device on the health of an individual, expressed in terms of a meaningful, measurable, patient-relevant clinical outcome(s), including outcome(s) related to diagnosis, or a positive impact on patient management or public health;(54)investigator means an individual responsible for the conduct of a clinical investigation at a clinical investigation site;(55)informed consent means a subject's free and voluntary expression of his or her willingness to participate in a particular clinical investigation, after having been informed of all aspects of the clinical investigation that are relevant to the subject's decision to participate or, in the case of minors and of incapacitated subjects, an authorisation or agreement from their legally designated representative to include them in the clinical investigation;(56)ethics committee means an independent body established in a Member State in accordance with the law of that Member State and empowered to give opinions for the purposes of this Regulation, taking into account the views of laypersons, in particular patients or patients' organisations;(57)adverse event means any untoward medical occurrence, unintended disease or injury or any untoward clinical signs, including an abnormal laboratory finding, in subjects, users or other persons, in the context of a clinical investigation, whether or not related to the investigational device;(58)serious adverse event means any adverse event that led to any of the following:(a)death,(b)serious deterioration in the health of the subject, that resulted in any of the following:(i)life-threatening illness or injury,(ii)permanent impairment of a body structure or a body function,(iii)hospitalisation or prolongation of patient hospitalisation,(iv)medical or surgical intervention to prevent life-threatening illness or injury or permanent impairment to a body structure or a body function,(v)chronic disease,(c)foetal distress, foetal death or a congenital physical or mental impairment or birth defect;(59)device deficiency means any inadequacy in the identity, quality, durability, reliability, safety or performance of an investigational device, including malfunction, use errors or inadequacy in information supplied by the manufacturer;(60)post-market surveillance means all activities carried out by manufacturers in cooperation with other economic operators to institute and keep up to date a systematic procedure to proactively collect and review experience gained from devices they place on the market, make available on the market or put into service for the purpose of identifying any need to immediately apply any necessary corrective or preventive actions;(61)market surveillance means the activities carried out and measures taken by competent authorities to check and ensure that devices comply with the requirements set out in the relevant Union harmonisation legislation and do not endanger health, safety or any other aspect of public interest protection;(62)recall means any measure aimed at achieving the return of a device that has already been made available to the end user;(63)withdrawal means any measure aimed at preventing a device in the supply chain from being further made available on the market;(64)incident means any malfunction or deterioration in the characteristics or performance of a device made available on the market, including use-error due to ergonomic features, as well as any inadequacy in the information supplied by the manufacturer and any undesirable side-effect;(65)serious incident means any incident that directly or indirectly led, might have led or might lead to any of the following:(a)the death of a patient, user or other person,(b)the temporary or permanent serious deterioration of a patient's, user's or other person's state of health,(c)a serious public health threat;(66)serious public health threat means an event which could result in imminent risk of death, serious deterioration in a person's state of health, or serious illness, that may require prompt remedial action, and that may cause significant morbidity or mortality in humans, or that is unusual or unexpected for the given place and time;(67)corrective action means action taken to eliminate the cause of a potential or actual non-conformity or other undesirable situation;(68)field safety corrective action means corrective action taken by a manufacturer for technical or medical reasons to prevent or reduce the risk of a serious incident in relation to a device made available on the market;(69)field safety notice means a communication sent by a manufacturer to users or customers in relation to a field safety corrective action;(70)harmonised standard means a European standard as defined in point (1)(c) of Article 2 of Regulation (EU) No 1025/2012;(71)common specifications (CS) means a set of technical and/or clinical requirements, other than a standard, that provides a means of complying with the legal obligations applicable to a device, process or system.

MODIFIED Art. 7 — Claims · applies from unchanged

Disputed — seen by the structural diff, not by corpus metadata.

The only change is the spelling of the word for registered marks, from a single word to two words. Art. 7, v1 Art. 7, v2

text before / after

32017R074502017R0745-20170505

Article 7ClaimsIn the labelling, instructions for use, making available, putting into service and advertising of devices, it shall be prohibited to use text, names, trademarks, trade marks, pictures and figurative or other signs that may mislead the user or the patient with regard to the device's intended purpose, safety and performance by:(a)ascribing functions and properties to the device which the device does not have;(b)creating a false impression regarding treatment or diagnosis, functions or properties which the device does not have;(c)failing to inform the user or the patient of a likely risk associated with the use of the device in line with its intended purpose;(d)suggesting uses for the device other than those stated to form part of the intended purpose for which the conformity assessment was carried out.

MODIFIED Art. 10 — General obligations of manufacturers · applies from unchanged

Disputed — seen by the structural diff, not by corpus metadata.

In paragraph 15, the cross-reference for where information on the identity of the person designing or manufacturing the device is to be submitted was changed from Article 30(1) to Article 29(4). Art. 10, v1 Art. 10, v2

text before / after

32017R074502017R0745-20170505

Article 10General obligations of manufacturers1.When placing their devices on the market or putting them into service, manufacturers shall ensure that they have been designed and manufactured in accordance with the requirements of this Regulation.2.Manufacturers shall establish, document, implement and maintain a system for risk management as described in Section 3 of Annex I.3.Manufacturers shall conduct a clinical evaluation in accordance with the requirements set out in Article 61 and Annex XIV, including a PMCF.4.Manufacturers of devices other than custom-made devices shall draw up and keep up to date technical documentation for those devices. The technical documentation shall be such as to allow the conformity of the device with the requirements of this Regulation to be assessed. The technical documentation shall include the elements set out in Annexes II and III.The Commission is empowered to adopt delegated acts in accordance with Article 115 amending, in the light of technical progress, the Annexes II and III.5.Manufacturers of custom-made devices shall draw up, keep up to date and keep available for competent authorities documentation in accordance with Section 2 of Annex XIII.6.Where compliance with the applicable requirements has been demonstrated following the applicable conformity assessment procedure, manufacturers of devices, other than custom-made or investigational devices, shall draw up an EU declaration of conformity in accordance with Article 19, and affix the CE marking of conformity in accordance with Article 20.7.Manufacturers shall comply with the obligations relating to the UDI system referred to in Article 27 and with the registration obligations referred to in Articles 29 and 31.8.Manufacturers shall keep the technical documentation, the EU declaration of conformity and, if applicable, a copy of any relevant certificate, including any amendments and supplements, issued in accordance with Article 56, available for the competent authorities for a period of at least 10 years after the last device covered by the EU declaration of conformity has been placed on the market. In the case of implantable devices, the period shall be at least 15 years after the last device has been placed on the market.Upon request by a competent authority, the manufacturer shall, as indicated therein, provide that technical documentation in its entirety or a summary thereof.A manufacturer with a registered place of business outside the Union shall, in order to allow its authorised representative to fulfil the tasks mentioned in Article 11(3), ensure that the authorised representative has the necessary documentation permanently available.9.Manufacturers shall ensure that procedures are in place to keep series production in conformity with the requirements of this Regulation. Changes in device design or characteristics and changes in the harmonised standards or CS by reference to which the conformity of a device is declared shall be adequately taken into account in a timely manner. Manufacturers of devices, other than investigational devices, shall establish, document, implement, maintain, keep up to date and continually improve a quality management system that shall ensure compliance with this Regulation in the most effective manner and in a manner that is proportionate to the risk class and the type of device.The quality management system shall cover all parts and elements of a manufacturer's organisation dealing with the quality of processes, procedures and devices. It shall govern the structure, responsibilities, procedures, processes and management resources required to implement the principles and actions necessary to achieve compliance with the provisions of this Regulation.The quality management system shall address at least the following aspects:(a)a strategy for regulatory compliance, including compliance with conformity assessment procedures and procedures for management of modifications to the devices covered by the system;(b)identification of applicable general safety and performance requirements and exploration of options to address those requirements;(c)responsibility of the management;(d)resource management, including selection and control of suppliers and sub-contractors;(e)risk management as set out in in Section 3 of Annex I;(f)clinical evaluation in accordance with Article 61 and Annex XIV, including PMCF;(g)product realisation, including planning, design, development, production and service provision;(h)verification of the UDI assignments made in accordance with Article 27(3) to all relevant devices and ensuring consistency and validity of information provided in accordance with Article 29;(i)setting-up, implementation and maintenance of a post-market surveillance system, in accordance with Article 83;(j)handling communication with competent authorities, notified bodies, other economic operators, customers and/or other stakeholders;(k)processes for reporting of serious incidents and field safety corrective actions in the context of vigilance;(l)management of corrective and preventive actions and verification of their effectiveness;(m)processes for monitoring and measurement of output, data analysis and product improvement.10.Manufacturers of devices shall implement and keep up to date the post-market surveillance system in accordance with Article 83.11.Manufacturers shall ensure that the device is accompanied by the information set out in Section 23 of Annex I in an official Union language(s) determined by the Member State in which the device is made available to the user or patient. The particulars on the label shall be indelible, easily legible and clearly comprehensible to the intended user or patient.12.Manufacturers who consider or have reason to believe that a device which they have placed on the market or put into service is not in conformity with this Regulation shall immediately take the necessary corrective action to bring that device into conformity, to withdraw it or to recall it, as appropriate. They shall inform the distributors of the device in question and, where applicable, the authorised representative and importers accordingly.Where the device presents a serious risk, manufacturers shall immediately inform the competent authorities of the Member States in which they made the device available and, where applicable, the notified body that issued a certificate for the device in accordance with Article 56, in particular, of the non-compliance and of any corrective action taken.13.Manufacturers shall have a system for recording and reporting of incidents and field safety corrective actions as described in Articles 87 and 88.14.Manufacturers shall, upon request by a competent authority, provide it with all the information and documentation necessary to demonstrate the conformity of the device, in an official Union language determined by the Member State concerned. The competent authority of the Member State in which the manufacturer has its registered place of business may require that the manufacturer provide samples of the device free of charge or, where that is impracticable, grant access to the device. Manufacturers shall cooperate with a competent authority, at its request, on any corrective action taken to eliminate or, if that is not possible, mitigate the risks posed by devices which they have placed on the market or put into service.If the manufacturer fails to cooperate or the information and documentation provided is incomplete or incorrect, the competent authority may, in order to ensure the protection of public health and patient safety, take all appropriate measures to prohibit or restrict the device's being made available on its national market, to withdraw the device from that market or to recall it until the manufacturer cooperates or provides complete and correct information.If a competent authority considers or has reason to believe that a device has caused damage, it shall, upon request, facilitate the provision of the information and documentation referred to in the first subparagraph to the potentially injured patient or user and, as appropriate, the patient's or user's successor in title, the patient's or user's health insurance company or other third parties affected by the damage caused to the patient or user, without prejudice to data protection rules and, unless there is an overriding public interest in disclosure, without prejudice to the protection of intellectual property rights.The competent authority need not comply with the obligation laid down in the third subparagraph where disclosure of the information and documentation referred to in the first subparagraph is ordinarily dealt with in the context of legal proceedings.15.Where manufacturers have their devices designed or manufactured by another legal or natural person the information on the identity of that person shall be part of the information to be submitted in accordance with Article 30(1).16.Natural 29(4).16.Natural or legal persons may claim compensation for damage caused by a defective device in accordance with applicable Union and national law.Manufacturers shall, in a manner that is proportionate to the risk class, type of device and the size of the enterprise, have measures in place to provide sufficient financial coverage in respect of their potential liability under Directive 85/374/EEC, without prejudice to more protective measures under national law.

MODIFIED Art. 74 — Clinical investigations regarding devices bearing the CE marking · applies from unchanged

Disputed — seen by the structural diff, not by corpus metadata.

The list of provisions applicable to PMCF investigations changed how Articles 75, 76 and 77 are referenced, listing them together as Articles 75, 76 and 77 rather than citing each separately with Article 75, Article 76, Article 77. Art. 74, v1 Art. 74, v2

The reference to Article 80(5) was expanded to also include Article 80(6). Art. 74, v2

text before / after

32017R074502017R0745-20170505

Article 74Clinical investigations regarding devices bearing the CE marking1.Where a clinical investigation is to be conducted to further assess, within the scope of its intended purpose, a device which already bears the CE marking in accordance with Article 20(1), (PMCF investigation), and where the investigation would involve submitting subjects to procedures additional to those performed under the normal conditions of use of the device and those additional procedures are invasive or burdensome, the sponsor shall notify the Member States concerned at least 30 days prior to its commencement by means of the electronic system referred to in Article 73. The sponsor shall include the documentation referred to in Chapter II of Annex XV as part of the notification. Points (b) to (k) and (m) of Article 62(4), Article Articles 75, Article 76, Article 76 and 77, and Article 80(5) and (6), and the relevant provisions of Annex XV shall apply to PMCF investigations.2.Where a clinical investigation is to be conducted to assess, outside the scope of its intended purpose, a device which already bears the CE marking in accordance with Article 20(1), Articles 62 to 81 shall apply.

MODIFIED Art. 78 — Coordinated assessment procedure for clinical investigations · applies from unknown (the text changed beyond its dates; the applicability binding is prose)

Disputed — seen by the structural diff, not by corpus metadata.

In paragraph 8(c), the cross-reference to considerations submitted under point (b) of paragraph 4 was changed to a reference to point (d) of paragraph 4. Art. 78, v1 Art. 78, v2

In paragraph 14, the date until which the coordinated procedure applies only to agreeing Member States was changed from 27 May 2027 to 25 May 2027, and the date from which all Member States must apply the procedure was changed from a phrase referring to after 27 May 2027 to a phrase referring to from 26 May 2027. Art. 78, v1 Art. 78, v2

text before / after

32017R074502017R0745-20170505

Article 78Coordinated assessment procedure for clinical investigations1.By means of the electronic system referred to in Article 73, the sponsor of a clinical investigation to be conducted in more than one Member State may submit, for the purpose of Article 70, a single application that, upon receipt, is transmitted electronically to all Member States in which the clinical investigation is to be conducted.2.The sponsor shall propose in the single application referred to in paragraph 1 that one of the Member States in which the clinical investigation is to be conducted acts as coordinating Member State. The Member States in which the clinical investigation is to be conducted shall, within six days of submission of the application, agree on one of them taking the role of the coordinating Member State. If they do not agree on a coordinating Member State, the coordinating Member State proposed by the sponsor shall assume that role.3.Under the direction of the coordinating Member State referred to in paragraph 2, the Member States concerned shall coordinate their assessment of the application, in particular of the documentation referred to in Chapter II of Annex XV.However, the completeness of the documentation referred to in Sections 1.13, 3.1.3, 4.2, 4.3 and 4.4 of Chapter II of Annex XV shall be assessed separately by each Member State concerned in accordance with Article 70(1) to (5).4.With regard to documentation other than that referred to in the second subparagraph of paragraph 3, the coordinating Member State shall:(a)within six days of receipt of the single application, notify the sponsor that it is the coordinating Member State (notification date);(b)for the purpose of the validation of the application, take into account any considerations submitted within seven days of the notification date by any Member State concerned;(c)within 10 days of the notification date, assess whether the clinical investigation falls within the scope of this Regulation and whether the application is complete, and shall notify the sponsor accordingly. Article 70(1) and (3) to (5) shall apply to the coordinating Member State in relation to that assessment;(d)establish the results of its assessment in a draft assessment report to be transmitted within 26 days of the validation date to the Member States concerned. By day 38 after the validation date, the other Member States concerned shall transmit their comments and proposals on the draft assessment report and the underlying application to the coordinating Member State which shall take due account of those comments and proposals in its finalisation of the final assessment report, to be transmitted within 45 days of the validation date to the sponsor and the other Member States concerned.The final assessment report shall be taken into account by all Member States concerned when deciding on the sponsor's application in accordance with Article 70(7).5.As regards the assessment of the documentation referred to in the second subparagraph of paragraph 3, each Member State concerned may request, on a single occasion, additional information from the sponsor. The sponsor shall submit the requested additional information within the period set by the Member State concerned, which shall not exceed 12 days from the receipt of the request. The expiry of the last deadline pursuant to point (d) of paragraph 4 shall be suspended from the date of the request until such time as the additional information has been received.6.For class IIb and class III devices, the coordinating Member State may also extend the periods referred to in paragraph 4 by a further 50 days, for the purpose of consulting with experts.7.The Commission may, by means of implementing acts, further specify the procedures and timescales for coordinated assessments to be taken into account by Member States concerned when deciding on the sponsor's application. Such implementing acts may also set out the procedures and timescales for coordinated assessment in the case of substantial modifications pursuant to paragraph 12 of this Article, in the case of reporting of adverse events pursuant to Article 80(4) and in the case of clinical investigations of combination products between medical devices and medicinal products, where the latter are under a concurrent coordinated assessment of a clinical trial under Regulation (EU) No 536/2014. Those implementing acts shall be adopted in accordance with the examination procedure referred to in Article 114(3).8.Where the conclusion of the coordinating Member State concerning the area of coordinated assessment is that the conduct of the clinical investigation is acceptable or acceptable subject to compliance with specific conditions, that conclusion shall be deemed to be the conclusion of all Member States concerned.Notwithstanding the first subparagraph, a Member State concerned may only disagree with the conclusion of the coordinating Member State concerning the area of coordinated assessment on the following grounds:(a)when it considers that participation in the clinical investigation would lead to a subject receiving treatment inferior to that received in normal clinical practice in that Member State concerned;(b)infringement of national law; or(c)considerations as regards subject safety and data reliability and robustness submitted under point (b) (d) of paragraph 4.Where one of the Member States concerned disagrees with the conclusion on the basis of the second subparagraph of this paragraph, it shall communicate its disagreement, together with a detailed justification, through the electronic system referred to in Article 73, to the Commission, to all other Member States concerned and to the sponsor.9.Where the conclusion of the coordinating Member State concerning the area of coordinated assessment is that the clinical investigation is not acceptable, that conclusion shall be deemed to be the conclusion of all Member States concerned.10.A Member State concerned shall refuse to authorise a clinical investigation if it disagrees with the conclusion of the coordinating Member State as regards any of the grounds referred to in the second subparagraph of paragraph 8, or if it finds, on duly justified grounds, that the aspects addressed in Sections 1.13, 3.1.3, 4.2, 4.3 and 4.4 of Chapter II of Annex XV are not complied with, or where an ethics committee has issued a negative opinion in relation to that clinical investigation, which is valid, in accordance with national law, for that entire Member State. That Member State shall provide for an appeal procedure in respect of such refusal.11.Each Member State concerned shall notify the sponsor through the electronic system referred to in Article 73 as to whether the clinical investigation is authorised, whether it is authorised subject to conditions, or whether authorisation has been refused. Notification shall be done by way of one single decision within five days of the transmission, pursuant to point (d) of paragraph 4, by the coordinating Member State of the final assessment report. Where an authorisation of a clinical investigation is subject to conditions, those conditions may only be such that, by their nature, they cannot be fulfilled at the time of that authorisation.12.Any substantial modifications as referred to in Article 75 shall be notified to the Member States concerned by means of the electronic system referred to in Article 73. Any assessment as to whether there are grounds for disagreement as referred to in the second subparagraph of paragraph 8 of this Article shall be carried out under the direction of the coordinating Member State, except for substantial modifications concerning Sections 1.13, 3.1.3, 4.2, 4.3 and 4.4 of Chapter II of Annex XV, which shall be assessed separately by each Member State concerned.13.The Commission shall provide administrative support to the coordinating Member State in the accomplishment of its tasks under this Chapter.14.The procedure set out in this Article shall, until 27 25 May 2027, be applied only by those of the Member States in which the clinical investigation is to be conducted which have agreed to apply it. After 27 From 26 May 2027, all Member States shall be required to apply that procedure.

MODIFIED Art. 84 — Post-market surveillance plan · applies from unchanged

Disputed — seen by the structural diff, not by corpus metadata.

The reference to the Annex III requirements changes from Section 1.1 to Section 1. Art. 84, v1 Art. 84, v2

text before / after

32017R074502017R0745-20170505

Article 84Post-market surveillance planThe post-market surveillance system referred to in Article 83 shall be based on a post-market surveillance plan, the requirements for which are set out in Section 1.1 1 of Annex III. For devices other than custom-made devices, the post-market surveillance plan shall be part of the technical documentation specified in Annex II.

MODIFIED Art. 88 — Trend reporting · applies from unchanged

Disputed — seen by the structural diff, not by corpus metadata.

The reference to the benefit-risk analysis sections of Annex I was changed from Sections 1 and 5 to Sections 1 and 8. Art. 88, v1 Art. 88, v2

text before / after

32017R074502017R0745-20170505

Article 88Trend reporting1.Manufacturers shall report, by means of the electronic system referred to in Article 92, any statistically significant increase in the frequency or severity of incidents that are not serious incidents or that are expected undesirable side-effects that could have a significant impact on the benefit-risk analysis referred to in Sections 1 and 5 8 of Annex I and which have led or may lead to risks to the health or safety of patients, users or other persons that are unacceptable when weighed against the intended benefits. The significant increase shall be established in comparison to the foreseeable frequency or severity of such incidents in respect of the device, or category or group of devices, in question during a specific period as specified in the technical documentation and product information.The manufacturer shall specify how to manage the incidents referred to in the first subparagraph and the methodology used for determining any statistically significant increase in the frequency or severity of such incidents, as well as the observation period, in the post-market surveillance plan referred to in Article 84.2.The competent authorities may conduct their own assessments on the trend reports referred to in paragraph 1 and require the manufacturer to adopt appropriate measures in accordance with this Regulation in order to ensure the protection of public health and patient safety. Each competent authority shall inform the Commission, the other competent authorities and the notified body that issued the certificate, of the results of such assessment and of the adoption of such measures.

MODIFIED Art. 120 — Transitional provisions · applies from unknown (the text changed beyond its dates; the applicability binding is prose)

Disputed — seen by the structural diff, not by corpus metadata.

Paragraph 3 now expressly extends the derogation to class I devices under Directive 93/42/EEC whose declaration of conformity was drawn up before 26 May 2020 and whose conformity assessment under this Regulation requires notified body involvement, alongside devices holding a still-valid certificate, and it sets a placing-on-market or putting-into-service cutoff of 26 May 2024 while changing the compliance start reference from the date of application of the Regulation to 26 May 2020. Art. 120, v2

Paragraph 4 changes the final continued-availability date from 27 May 2025 to 26 May 2025 and rephrases the reference to devices placed on the market from 26 May 2020 as being placed pursuant to paragraph 3 rather than by virtue of a certificate under paragraph 2. Art. 120, v2

Paragraph 8 adds Article 11(5) of Directive 90/385/EEC and Article 16(5) of Directive 93/42/EEC to the derogated provisions and adds Article 31(1) of this Regulation to the compliance conditions, while paragraph 10 removes the reference to point (f) of Article 1(6), leaving only point (g), and paragraph 6 corrects "prior 26 May 2020" to "prior to 26 May 2020." Art. 120, v2

text before / after

32017R074502017R0745-20170505

Article 120Transitional provisions1.From 26 May 2020, any publication of a notification in respect of a notified body in accordance with Directives 90/385/EEC and 93/42/EEC shall become void.2.Certificates issued by notified bodies in accordance with Directives 90/385/EEC and 93/42/EEC prior to 25 May 2017 shall remain valid until the end of the period indicated on the certificate, except for certificates issued in accordance with Annex 4 to Directive 90/385/EEC or Annex IV to Directive 93/42/EEC which shall become void at the latest on 27 May 2022.Certificates issued by notified bodies in accordance with Directives 90/385/EEC and 93/42/EEC from 25 May 2017 shall remain valid until the end of the period indicated on the certificate, which shall not exceed five years from its issuance. They shall however become void at the latest on 27 May 2024.3.By way of derogation from Article 5 of this Regulation, a device with which is a class I device pursuant to Directive 93/42/EEC, for which the declaration of conformity was drawn up prior to 26 May 2020 and for which the conformity assessment procedure pursuant to this Regulation requires the involvement of a notified body, or which has a certificate that was issued in accordance with Directive 90/385/EEC or Directive 93/42/EEC and which that is valid by virtue of paragraph 2 of this Article Article, may only be placed on the market or put into service until 26 May 2024, provided that from the date of application of this Regulation 26 May 2020 it continues to comply with either of those Directives, and provided there are no significant changes in the design and intended purpose. However, the requirements of this Regulation relating to post-market surveillance, market surveillance, vigilance, registration of economic operators and of devices shall apply in place of the corresponding requirements in those Directives.Without prejudice to Chapter IV and paragraph 1 of this Article, the notified body that issued the certificate referred to in the first subparagraph shall continue to be responsible for the appropriate surveillance in respect of all of the applicable requirements relating to the devices it has certified.4.Devices lawfully placed on the market pursuant to Directives 90/385/EEC and 93/42/EEC prior to 26 May 2020, and devices placed on the market from 26 May 2020 by virtue of a certificate as referred pursuant to in paragraph 2 3 of this Article, may continue to be made available on the market or put into service until 27 26 May 2025.5.By way of derogation from Directives 90/385/EEC and 93/42/EEC, devices which comply with this Regulation may be placed on the market prior to 26 May 2020.6.By way of derogation from Directives 90/385/EEC and 93/42/EEC, conformity assessment bodies which comply with this Regulation may be designated and notified prior to 26 May 2020. Notified bodies which are designated and notified in accordance with this Regulation may carry out the conformity assessment procedures laid down in this Regulation and issue certificates in accordance with this Regulation prior to 26 May 2020.7.As regards devices subject to the consultation procedure laid down in Article 54, paragraph 5 of this Article shall apply provided that the necessary appointments to the MDCG and expert panels have been made.8.By way of derogation from Article 10a and 10a, point (a) of Article 10b(1) and Article 11(5) of Directive 90/385/EEC and Article 14(1) and (2) and (2), points (a) and (b) of Article 14a(1) and Article 16(5) of Directive 93/42/EEC, manufacturers, authorised representatives, importers and notified bodies which, during the period starting on the later of the dates referred to in point (d) of Article 123(3) and ending 18 months later, comply with Article 29(4) Articles 29(4), 31(1) and Article 56(5) of this Regulation shall be considered to comply with the laws and regulations adopted by Member States in accordance with, respectively, Article 10a of Directive 90/385/EEC or Article 14(1) and (2) of Directive 93/42/EEC and 93/42/EEC, with, respectively, point (a) of Article 10b(1) of Directive 90/385/EEC or points (a) and (b) of Article 14a(1) of Directive 93/42/EEC and with, respectively, Article 11(5) of Directive 90/385/EEC or Article 16(5) of Directive 93/42/EEC, as specified in Decision 2010/227/EU.9.Authorisations granted by the competent authorities of the Member States in accordance with Article 9(9) of Directive 90/385/EEC or Article 11(13) of Directive 93/42/EEC shall keep the validity indicated in the authorisation.10.Devices falling within the scope of this Regulation in accordance with points (f) and point (g) of Article 1(6) which have been legally placed on the market or put into service in accordance with the rules in force in the Member States prior to 26 May 2020 may continue to be placed on the market and put into service in the Member States concerned.11.Clinical investigations which have started to be conducted in accordance with Article 10 of Directive 90/385/EEC or Article 15 of Directive 93/42/EEC prior to 26 May 2020 may continue to be conducted. As of 26 May 2020, however, the reporting of serious adverse events and device deficiencies shall be carried out in accordance with this Regulation.12.Until the Commission has designated, pursuant to Article 27(2), issuing entities, GS1, HIBCC and ICCBBA shall be considered to be designated issuing entities.

MODIFIED Art. 122 — Repeal · applies from unchanged

Disputed — seen by the structural diff, not by corpus metadata.

The second exception listed in Article 122(1) now also includes Article 11(5) of Directive 90/385/EEC among the provisions repealed 18 months after the later of the dates referred to in point (d) of Article 123(3). Art. 122, v2

The fourth exception listed in Article 122(1) now also includes Article 16(5) of Directive 93/42/EEC among the provisions repealed 18 months after the later of the dates referred to in point (d) of Article 123(3). Art. 122, v2

text before / after

32017R074502017R0745-20170505

Article 122RepealWithout prejudice to Articles 120(3) and (4) of this Regulation, and without prejudice to the obligations of the Member States and manufacturers as regards vigilance and to the obligations of manufacturers as regards the making available of documentation, under Directives 90/385/EEC and 93/42/EEC, those Directives are repealed with effect from 26 May 2020, with the exception of:Articles 8 and 10, points (b) and (c) of Article 10b(1), Article 10b(2) and Article 10b(3) of Directive 90/385/EEC, and the obligations relating to vigilance and clinical investigations provided for in the corresponding Annexes, which are repealed with effect from the later of the dates referred to in point (d) of Article 123(3) of this Regulation;Article 10a and 10a, point (a) of Article 10b(1) and Article 11(5) of Directive 90/385/EEC, and the obligations relating to registration of devices and economic operators, and to certificate notifications, provided for in the corresponding Annexes, which are repealed with effect from 18 months after the later of the dates referred to in point (d) of Article 123(3) of this Regulation;Article 10, points (c) and (d) of Article 14a(1), Article 14a(2), Article 14a(3) and Article 15 of Directive 93/42/EEC, and the obligations relating to vigilance and clinical investigations provided for in the corresponding Annexes, which are repealed with effect from the later of the dates referred to in point (d) of Article 123(3) of this Regulation; andArticle 14(1) and (2) and (2), points (a) and (b) of Article 14a(1) and Article 16(5) of Directive 93/42/EEC, and the obligations relating to registration of devices and economic operators, and to certificate notifications, provided for in the corresponding Annexes, which are repealed with effect from 18 months after the later of the dates referred to in point (d) of Article 123(3) of this Regulation.As regards the devices referred to in Article 120 (3) and (4) of this Regulation, the Directives referred to in the first paragraph shall continue to apply until 27 May 2025 to the extent necessary for the application of those paragraphs.Notwithstanding the first paragraph, Regulations (EU) No 207/2012 and (EU) No 722/2012 shall remain in force and continue to apply unless and until repealed by implementing acts adopted by the Commission pursuant to this Regulation.References to the repealed Directives shall be understood as references to this Regulation and shall be read in accordance with the correlation table laid down in Annex XVII to this Regulation.

MODIFIED Annex I — ANNEX I · applies from unchanged

Disputed — seen by the structural diff, not by corpus metadata.

In section 23.2(h), the reference to the UDI carrier changes from citing Article 27(4) together with Part C of Annex VII to citing Article 27(4) together with Part C of Annex VI. Annex I, v1 Annex I, v2

text before / after

32017R074502017R0745-20170505

ANNEX IGENERAL SAFETY AND PERFORMANCE REQUIREMENTSCHAPTER IGENERAL REQUIREMENTS1.Devices shall achieve the performance intended by their manufacturer and shall be designed and manufactured in such a way that, during normal conditions of use, they are suitable for their intended purpose. They shall be safe and effective and shall not compromise the clinical condition or the safety of patients, or the safety and health of users or, where applicable, other persons, provided that any risks which may be associated with their use constitute acceptable risks when weighed against the benefits to the patient and are compatible with a high level of protection of health and safety, taking into account the generally acknowledged state of the art.2.The requirement in this Annex to reduce risks as far as possible means the reduction of risks as far as possible without adversely affecting the benefit-risk ratio.3.Manufacturers shall establish, implement, document and maintain a risk management system.Risk management shall be understood as a continuous iterative process throughout the entire lifecycle of a device, requiring regular systematic updating. In carrying out risk management manufacturers shall:(a)establish and document a risk management plan for each device;(b)identify and analyse the known and foreseeable hazards associated with each device;(c)estimate and evaluate the risks associated with, and occurring during, the intended use and during reasonably foreseeable misuse;(d)eliminate or control the risks referred to in point (c) in accordance with the requirements of Section 4;(e)evaluate the impact of information from the production phase and, in particular, from the post-market surveillance system, on hazards and the frequency of occurrence thereof, on estimates of their associated risks, as well as on the overall risk, benefit-risk ratio and risk acceptability; and(f)based on the evaluation of the impact of the information referred to in point (e), if necessary amend control measures in line with the requirements of Section 4.4.Risk control measures adopted by manufacturers for the design and manufacture of the devices shall conform to safety principles, taking account of the generally acknowledged state of the art. To reduce risks, Manufacturers shall manage risks so that the residual risk associated with each hazard as well as the overall residual risk is judged acceptable. In selecting the most appropriate solutions, manufacturers shall, in the following order of priority:(a)eliminate or reduce risks as far as possible through safe design and manufacture;(b)where appropriate, take adequate protection measures, including alarms if necessary, in relation to risks that cannot be eliminated; and(c)provide information for safety (warnings/precautions/contra-indications) and, where appropriate, training to users.Manufacturers shall inform users of any residual risks.5.In eliminating or reducing risks related to use error, the manufacturer shall:(a)reduce as far as possible the risks related to the ergonomic features of the device and the environment in which the device is intended to be used (design for patient safety), and(b)give consideration to the technical knowledge, experience, education, training and use environment, where applicable, and the medical and physical conditions of intended users (design for lay, professional, disabled or other users).6.The characteristics and performance of a device shall not be adversely affected to such a degree that the health or safety of the patient or the user and, where applicable, of other persons are compromised during the lifetime of the device, as indicated by the manufacturer, when the device is subjected to the stresses which can occur during normal conditions of use and has been properly maintained in accordance with the manufacturer's instructions.7.Devices shall be designed, manufactured and packaged in such a way that their characteristics and performance during their intended use are not adversely affected during transport and storage, for example, through fluctuations of temperature and humidity, taking account of the instructions and information provided by the manufacturer.8.All known and foreseeable risks, and any undesirable side-effects, shall be minimised and be acceptable when weighed against the evaluated benefits to the patient and/or user arising from the achieved performance of the device during normal conditions of use.9.For the devices referred to in Annex XVI, the general safety requirements set out in Sections 1 and 8 shall be understood to mean that the device, when used under the conditions and for the purposes intended, does not present a risk at all or presents a risk that is no more than the maximum acceptable risk related to the product's use which is consistent with a high level of protection for the safety and health of persons.CHAPTER IIREQUIREMENTS REGARDING DESIGN AND MANUFACTURE10.Chemical, physical and biological properties10.1.Devices shall be designed and manufactured in such a way as to ensure that the characteristics and performance requirements referred to in Chapter I are fulfilled. Particular attention shall be paid to:(a)the choice of materials and substances used, particularly as regards toxicity and, where relevant, flammability;(b)the compatibility between the materials and substances used and biological tissues, cells and body fluids, taking account of the intended purpose of the device and, where relevant, absorption, distribution, metabolism and excretion;(c)the compatibility between the different parts of a device which consists of more than one implantable part;(d)the impact of processes on material properties;(e)where appropriate, the results of biophysical or modelling research the validity of which has been demonstrated beforehand;(f)the mechanical properties of the materials used, reflecting, where appropriate, matters such as strength, ductility, fracture resistance, wear resistance and fatigue resistance;(g)surface properties; and(h)the confirmation that the device meets any defined chemical and/or physical specifications.10.2.Devices shall be designed, manufactured and packaged in such a way as to minimise the risk posed by contaminants and residues to patients, taking account of the intended purpose of the device, and to the persons involved in the transport, storage and use of the devices. Particular attention shall be paid to tissues exposed to those contaminants and residues and to the duration and frequency of exposure.10.3.Devices shall be designed and manufactured in such a way that they can be used safely with the materials and substances, including gases, with which they enter into contact during their intended use; if the devices are intended to administer medicinal products they shall be designed and manufactured in such a way as to be compatible with the medicinal products concerned in accordance with the provisions and restrictions governing those medicinal products and that the performance of both the medicinal products and of the devices is maintained in accordance with their respective indications and intended use.10.4.Substances10.4.1.Design and manufacture of devicesDevices shall be designed and manufactured in such a way as to reduce as far as possible the risks posed by substances or particles, including wear debris, degradation products and processing residues, that may be released from the device.Devices, or those parts thereof or those materials used therein that:are invasive and come into direct contact with the human body,(re)administer medicines, body liquids or other substances, including gases, to/from the body, ortransport or store such medicines, body fluids or substances, including gases, to be (re)administered to the body,shall only contain the following substances in a concentration that is above 0,1 % weight by weight (w/w) where justified pursuant to Section 10.4.2:(a)substances which are carcinogenic, mutagenic or toxic to reproduction (CMR), of category 1A or 1B, in accordance with Part 3 of Annex VI to Regulation (EC) No 1272/2008 of the European Parliament and of the CouncilRegulation (EC) No 1272/2008 of the European Parliament and of the Council of 16 December 2008 on classification, labelling and packaging of substances and mixtures, amending and repealing Directives 67/548/EEC and 1999/45/EC, and amending Regulation (EC) No 1907/2006 ( OJ L 353, 31.12.2008, p. 1)., or(b)substances having endocrine-disrupting properties for which there is scientific evidence of probable serious effects to human health and which are identified either in accordance with the procedure set out in Article 59 of Regulation (EC) No 1907/2006 of the European Parliament and of the CouncilRegulation (EC) No 1907/2006 of the European Parliament and of the Council of 18 December 2006 concerning the Registration, Evaluation, Authorisation and Restriction of Chemicals (REACH) (OJ L 396, 30.12.2006, p. 1). or, once a delegated act has been adopted by the Commission pursuant to the first subparagraph of Article 5(3) of Regulation (EU) No 528/2012 of the European Parliament and the CouncilRegulation (EU) No 528/2012 of the European Parliament and the Council of 22 May 2012 concerning the making available on the market of and use of biocidal products (OJ L 167, 27.6.2012, p. 1)., in accordance with the criteria that are relevant to human health amongst the criteria established therein.10.4.2.Justification regarding the presence of CMR and/or endocrine-disrupting substancesThe justification for the presence of such substances shall be based upon:(a)an analysis and estimation of potential patient or user exposure to the substance;(b)an analysis of possible alternative substances, materials or designs, including, where available, information about independent research, peer-reviewed studies, scientific opinions from relevant scientific committees and an analysis of the availability of such alternatives;(c)argumentation as to why possible substance and/ or material substitutes, if available, or design changes, if feasible, are inappropriate in relation to maintaining the functionality, performance and the benefit-risk ratios of the product; including taking into account if the intended use of such devices includes treatment of children or treatment of pregnant or breastfeeding women or treatment of other patient groups considered particularly vulnerable to such substances and/or materials; and(d)where applicable and available, the latest relevant scientific committee guidelines in accordance with Sections 10.4.3. and 10.4.4.10.4.3.Guidelines on phthalatesFor the purposes of Section 10.4., the Commission shall, as soon as possible and by 26 May 2018, provide the relevant scientific committee with a mandate to prepare guidelines that shall be ready before 26 May 2020. The mandate for the committee shall encompass at least a benefit-risk assessment of the presence of phthalates which belong to either of the groups of substances referred to in points (a) and (b) of Section 10.4.1. The benefit-risk assessment shall take into account the intended purpose and context of the use of the device, as well as any available alternative substances and alternative materials, designs or medical treatments. When deemed appropriate on the basis of the latest scientific evidence, but at least every five years, the guidelines shall be updated.10.4.4.Guidelines on other CMR and endocrine-disrupting substancesSubsequently, the Commission shall mandate the relevant scientific committee to prepare guidelines as referred to in Section 10.4.3. also for other substances referred to in points (a) and (b) of Section 10.4.1., where appropriate.10.4.5.LabellingWhere devices, parts thereof or materials used therein as referred to in Section 10.4.1. contain substances referred to in points (a) or (b) of Section 10.4.1. in a concentration above 0,1 % weight by weight (w/w), the presence of those substances shall be labelled on the device itself and/or on the packaging for each unit or, where appropriate, on the sales packaging, with the list of such substances. If the intended use of such devices includes treatment of children or treatment of pregnant or breastfeeding women or treatment of other patient groups considered particularly vulnerable to such substances and/or materials, information on residual risks for those patient groups and, if applicable, on appropriate precautionary measures shall be given in the instructions for use.10.5.Devices shall be designed and manufactured in such a way as to reduce as far as possible the risks posed by the unintentional ingress of substances into the device taking into account the device and the nature of the environment in which it is intended to be used.10.6.Devices shall be designed and manufactured in such a way as to reduce as far as possible the risks linked to the size and the properties of particles which are or can be released into the patient's or user's body, unless they come into contact with intact skin only. Special attention shall be given to nanomaterials.11.Infection and microbial contamination11.1.Devices and their manufacturing processes shall be designed in such a way as to eliminate or to reduce as far as possible the risk of infection to patients, users and, where applicable, other persons. The design shall:(a)reduce as far as possible and appropriate the risks from unintended cuts and pricks, such as needle stick injuries,(b)allow easy and safe handling,(c)reduce as far as possible any microbial leakage from the device and/or microbial exposure during use, and(d)prevent microbial contamination of the device or its content such as specimens or fluids.11.2.Where necessary devices shall be designed to facilitate their safe cleaning, disinfection, and/or re-sterilisation.11.3.Devices labelled as having a specific microbial state shall be designed, manufactured and packaged to ensure that they remain in that state when placed on the market and remain so under the transport and storage conditions specified by the manufacturer.11.4.Devices delivered in a sterile state shall be designed, manufactured and packaged in accordance with appropriate procedures, to ensure that they are sterile when placed on the market and that, unless the packaging which is intended to maintain their sterile condition is damaged, they remain sterile, under the transport and storage conditions specified by the manufacturer, until that packaging is opened at the point of use. It shall be ensured that the integrity of that packaging is clearly evident to the final user.11.5.Devices labelled as sterile shall be processed, manufactured, packaged and, sterilised by means of appropriate, validated methods.11.6.Devices intended to be sterilised shall be manufactured and packaged in appropriate and controlled conditions and facilities.11.7.Packaging systems for non-sterile devices shall maintain the integrity and cleanliness of the product and, where the devices are to be sterilised prior to use, minimise the risk of microbial contamination; the packaging system shall be suitable taking account of the method of sterilisation indicated by the manufacturer.11.8.The labelling of the device shall distinguish between identical or similar devices placed on the market in both a sterile and a non-sterile condition additional to the symbol used to indicate that devices are sterile.12.Devices incorporating a substance considered to be a medicinal product and devices that are composed of substances or of combinations of substances that are absorbed by or locally dispersed in the human body.12.1.In the case of devices referred to in the first subparagraph of Article 1(8), the quality, safety and usefulness of the substance which, if used separately, would be considered to be a medicinal product within the meaning of point (2) of Article 1 of Directive 2001/83/EC, shall be verified by analogy with the methods specified in Annex I to Directive 2001/83/EC, as required by the applicable conformity assessment procedure under this Regulation.12.2.Devices that are composed of substances or of combinations of substances that are intended to be introduced into the human body, and that are absorbed by or locally dispersed in the human body shall comply, where applicable and in a manner limited to the aspects not covered by this Regulation, with the relevant requirements laid down in Annex I to Directive 2001/83/EC for the evaluation of absorption, distribution, metabolism, excretion, local tolerance, toxicity, interaction with other devices, medicinal products or other substances and potential for adverse reactions, as required by the applicable conformity assessment procedure under this Regulation.13.Devices incorporating materials of biological origin13.1.For devices manufactured utilising derivatives of tissues or cells of human origin which are non-viable or are rendered non-viable covered by this Regulation in accordance with point (g) of Article 1(6), the following shall apply:(a)donation, procurement and testing of the tissues and cells shall be done in accordance with Directive 2004/23/EC;(b)processing, preservation and any other handling of those tissues and cells or their derivatives shall be carried out so as to provide safety for patients, users and, where applicable, other persons. In particular, safety with regard to viruses and other transmissible agents shall be addressed by appropriate methods of sourcing and by implementation of validated methods of elimination or inactivation in the course of the manufacturing process;(c)the traceability system for those devices shall be complementary and compatible with the traceability and data protection requirements laid down in Directive 2004/23/EC and in Directive 2002/98/EC.13.2.For devices manufactured utilising tissues or cells of animal origin, or their derivatives, which are non-viable or rendered non-viable the following shall apply:(a)where feasible taking into account the animal species, tissues and cells of animal origin, or their derivatives, shall originate from animals that have been subjected to veterinary controls that are adapted to the intended use of the tissues. Information on the geographical origin of the animals shall be retained by manufacturers;(b)sourcing, processing, preservation, testing and handling of tissues, cells and substances of animal origin, or their derivatives, shall be carried out so as to provide safety for patients, users and, where applicable, other persons. In particular safety with regard to viruses and other transmissible agents shall be addressed by implementation of validated methods of elimination or viral inactivation in the course of the manufacturing process, except when the use of such methods would lead to unacceptable degradation compromising the clinical benefit of the device;(c)in the case of devices manufactured utilising tissues or cells of animal origin, or their derivatives, as referred to in Regulation (EU) No 722/2012 the particular requirements laid down in that Regulation shall apply.13.3.For devices manufactured utilising non-viable biological substances other than those referred to in Sections 13.1 and 13.2, the processing, preservation, testing and handling of those substances shall be carried out so as to provide safety for patients, users and, where applicable, other persons, including in the waste disposal chain. In particular, safety with regard to viruses and other transmissible agents shall be addressed by appropriate methods of sourcing and by implementation of validated methods of elimination or inactivation in the course of the manufacturing process.14.Construction of devices and interaction with their environment14.1.If the device is intended for use in combination with other devices or equipment the whole combination, including the connection system shall be safe and shall not impair the specified performance of the devices. Any restrictions on use applying to such combinations shall be indicated on the label and/or in the instructions for use. Connections which the user has to handle, such as fluid, gas transfer, electrical or mechanical coupling, shall be designed and constructed in such a way as to minimise all possible risks, such as misconnection.14.2.Devices shall be designed and manufactured in such a way as to remove or reduce as far as possible:(a)the risk of injury, in connection with their physical features, including the volume/pressure ratio, dimensional and where appropriate ergonomic features;(b)risks connected with reasonably foreseeable external influences or environmental conditions, such as magnetic fields, external electrical and electromagnetic effects, electrostatic discharge, radiation associated with diagnostic or therapeutic procedures, pressure, humidity, temperature, variations in pressure and acceleration or radio signal interferences;(c)the risks associated with the use of the device when it comes into contact with materials, liquids, and substances, including gases, to which it is exposed during normal conditions of use;(d)the risks associated with the possible negative interaction between software and the IT environment within which it operates and interacts;(e)the risks of accidental ingress of substances into the device;(f)the risks of reciprocal interference with other devices normally used in the investigations or for the treatment given; and(g)risks arising where maintenance or calibration are not possible (as with implants), from ageing of materials used or loss of accuracy of any measuring or control mechanism.14.3.Devices shall be designed and manufactured in such a way as to minimise the risks of fire or explosion during normal use and in single fault condition. Particular attention shall be paid to devices the intended use of which includes exposure to or use in association with flammable or explosive substances or substances which could cause combustion.14.4.Devices shall be designed and manufactured in such a way that adjustment, calibration, and maintenance can be done safely and effectively.14.5.Devices that are intended to be operated together with other devices or products shall be designed and manufactured in such a way that the interoperability and compatibility are reliable and safe.14.6Any measurement, monitoring or display scale shall be designed and manufactured in line with ergonomic principles, taking account of the intended purpose, users and the environmental conditions in which the devices are intended to be used.14.7.Devices shall be designed and manufactured in such a way as to facilitate their safe disposal and the safe disposal of related waste substances by the user, patient or other person. To that end, manufacturers shall identify and test procedures and measures as a result of which their devices can be safely disposed after use. Such procedures shall be described in the instructions for use.15.Devices with a diagnostic or measuring function15.1.Diagnostic devices and devices with a measuring function, shall be designed and manufactured in such a way as to provide sufficient accuracy, precision and stability for their intended purpose, based on appropriate scientific and technical methods. The limits of accuracy shall be indicated by the manufacturer.15.2.The measurements made by devices with a measuring function shall be expressed in legal units conforming to the provisions of Council Directive 80/181/EECCouncil Directive 80/181/EEC of 20 December 1979 on the approximation of the laws of the Member States relating to units of measurement and on the repeal of Directive 71/354/EEC (OJ L 39, 15.2.1980, p. 40)..16.Protection against radiation16.1.General(a)Devices shall be designed, manufactured and packaged in such a way that exposure of patients, users and other persons to radiation is reduced as far as possible, and in a manner that is compatible with the intended purpose, whilst not restricting the application of appropriate specified levels for therapeutic and diagnostic purposes.(b)The operating instructions for devices emitting hazardous or potentially hazardous radiation shall contain detailed information as to the nature of the emitted radiation, the means of protecting the patient and the user, and on ways of avoiding misuse and of reducing the risks inherent to installation as far as possible and appropriate. Information regarding the acceptance and performance testing, the acceptance criteria, and the maintenance procedure shall also be specified.16.2.Intended radiation(a)Where devices are designed to emit hazardous, or potentially hazardous, levels of ionizing and/or non-ionizing radiation necessary for a specific medical purpose the benefit of which is considered to outweigh the risks inherent to the emission, it shall be possible for the user to control the emissions. Such devices shall be designed and manufactured to ensure reproducibility of relevant variable parameters within an acceptable tolerance.(b)Where devices are intended to emit hazardous, or potentially hazardous, ionizing and/or non-ionizing radiation, they shall be fitted, where possible, with visual displays and/or audible warnings of such emissions.16.3.Devices shall be designed and manufactured in such a way that exposure of patients, users and other persons to the emission of unintended, stray or scattered radiation is reduced as far as possible. Where possible and appropriate, methods shall be selected which reduce the exposure to radiation of patients, users and other persons who may be affected.16.4.Ionising radiation(a)Devices intended to emit ionizing radiation shall be designed and manufactured taking into account the requirements of the Directive 2013/59/Euratom laying down basic safety standards for protection against the dangers arising from exposure to ionising radiation.(b)Devices intended to emit ionising radiation shall be designed and manufactured in such a way as to ensure that, where possible, taking into account the intended use, the quantity, geometry and quality of the radiation emitted can be varied and controlled, and, if possible, monitored during treatment.(c)Devices emitting ionising radiation intended for diagnostic radiology shall be designed and manufactured in such a way as to achieve an image and/or output quality that are appropriate to the intended medical purpose whilst minimising radiation exposure of the patient and user.(d)Devices that emit ionising radiation and are intended for therapeutic radiology shall be designed and manufactured in such a way as to enable reliable monitoring and control of the delivered dose, the beam type, energy and, where appropriate, the quality of radiation.17.Electronic programmable systems — devices that incorporate electronic programmable systems and software that are devices in themselves17.1.Devices that incorporate electronic programmable systems, including software, or software that are devices in themselves, shall be designed to ensure repeatability, reliability and performance in line with their intended use. In the event of a single fault condition, appropriate means shall be adopted to eliminate or reduce as far as possible consequent risks or impairment of performance.17.2.For devices that incorporate software or for software that are devices in themselves, the software shall be developed and manufactured in accordance with the state of the art taking into account the principles of development life cycle, risk management, including information security, verification and validation.17.3.Software referred to in this Section that is intended to be used in combination with mobile computing platforms shall be designed and manufactured taking into account the specific features of the mobile platform (e.g. size and contrast ratio of the screen) and the external factors related to their use (varying environment as regards level of light or noise).17.4.Manufacturers shall set out minimum requirements concerning hardware, IT networks characteristics and IT security measures, including protection against unauthorised access, necessary to run the software as intended.18.Active devices and devices connected to them18.1.For non-implantable active devices, in the event of a single fault condition, appropriate means shall be adopted to eliminate or reduce as far as possible consequent risks.18.2.Devices where the safety of the patient depends on an internal power supply shall be equipped with a means of determining the state of the power supply and an appropriate warning or indication for when the capacity of the power supply becomes critical. If necessary, such warning or indication shall be given prior to the power supply becoming critical.18.3.Devices where the safety of the patient depends on an external power supply shall include an alarm system to signal any power failure.18.4.Devices intended to monitor one or more clinical parameters of a patient shall be equipped with appropriate alarm systems to alert the user of situations which could lead to death or severe deterioration of the patient's state of health.18.5.Devices shall be designed and manufactured in such a way as to reduce as far as possible the risks of creating electromagnetic interference which could impair the operation of the device in question or other devices or equipment in the intended environment.18.6.Devices shall be designed and manufactured in such a way as to provide a level of intrinsic immunity to electromagnetic interference such that is adequate to enable them to operate as intended.18.7.Devices shall be designed and manufactured in such a way as to avoid, as far as possible, the risk of accidental electric shocks to the patient, user or any other person, both during normal use of the device and in the event of a single fault condition in the device, provided the device is installed and maintained as indicated by the manufacturer.18.8.Devices shall be designed and manufactured in such a way as to protect, as far as possible, against unauthorised access that could hamper the device from functioning as intended.19.Particular requirements for active implantable devices19.1.Active implantable devices shall be designed and manufactured in such a way as to remove or minimize as far as possible:(a)risks connected with the use of energy sources with particular reference, where electricity is used, to insulation, leakage currents and overheating of the devices,(b)risks connected with medical treatment, in particular those resulting from the use of defibrillators or high-frequency surgical equipment, and(c)risks which may arise where maintenance and calibration are impossible, including:excessive increase of leakage currents,ageing of the materials used,excess heat generated by the device,decreased accuracy of any measuring or control mechanism.19.2.Active implantable devices shall be designed and manufactured in such a way as to ensureif applicable, the compatibility of the devices with the substances they are intended to administer, andthe reliability of the source of energy.19.3.Active implantable devices and, if appropriate, their component parts shall be identifiable to allow any necessary measure to be taken following the discovery of a potential risk in connection with the devices or their component parts.19.4.Active implantable devices shall bear a code by which they and their manufacturer can be unequivocally identified (particularly with regard to the type of device and its year of manufacture); it shall be possible to read this code, if necessary, without the need for a surgical operation.20.Protection against mechanical and thermal risks20.1.Devices shall be designed and manufactured in such a way as to protect patients and users against mechanical risks connected with, for example, resistance to movement, instability and moving parts.20.2.Devices shall be designed and manufactured in such a way as to reduce to the lowest possible level the risks arising from vibration generated by the devices, taking account of technical progress and of the means available for limiting vibrations, particularly at source, unless the vibrations are part of the specified performance.20.3.Devices shall be designed and manufactured in such a way as to reduce to the lowest possible level the risks arising from the noise emitted, taking account of technical progress and of the means available to reduce noise, particularly at source, unless the noise emitted is part of the specified performance.20.4.Terminals and connectors to the electricity, gas or hydraulic and pneumatic energy supplies which the user or other person has to handle, shall be designed and constructed in such a way as to minimise all possible risks.20.5.Errors likely to be made when fitting or refitting certain parts which could be a source of risk shall be made impossible by the design and construction of such parts or, failing this, by information given on the parts themselves and/or their housings.The same information shall be given on moving parts and/or their housings where the direction of movement needs to be known in order to avoid a risk.20.6.Accessible parts of devices (excluding the parts or areas intended to supply heat or reach given temperatures) and their surroundings shall not attain potentially dangerous temperatures under normal conditions of use.21.Protection against the risks posed to the patient or user by devices supplying energy or substances21.1.Devices for supplying the patient with energy or substances shall be designed and constructed in such a way that the amount to be delivered can be set and maintained accurately enough to ensure the safety of the patient and of the user.21.2.Devices shall be fitted with the means of preventing and/or indicating any inadequacies in the amount of energy delivered or substances delivered which could pose a danger. Devices shall incorporate suitable means to prevent, as far as possible, the accidental release of dangerous levels of energy or substances from an energy and/or substance source.21.3.The function of the controls and indicators shall be clearly specified on the devices. Where a device bears instructions required for its operation or indicates operating or adjustment parameters by means of a visual system, such information shall be understandable to the user and, as appropriate, the patient.22.Protection against the risks posed by medical devices intended by the manufacturer for use by lay persons22.1.Devices for use by lay persons shall be designed and manufactured in such a way that they perform appropriately for their intended purpose taking into account the skills and the means available to lay persons and the influence resulting from variation that can be reasonably anticipated in the lay person's technique and environment. The information and instructions provided by the manufacturer shall be easy for the lay person to understand and apply.22.2.Devices for use by lay persons shall be designed and manufactured in such a way as to:ensure that the device can be used safely and accurately by the intended user at all stages of the procedure, if necessary after appropriate training and/or information,reduce, as far as possible and appropriate, the risk from unintended cuts and pricks such as needle stick injuries, andreduce as far as possible the risk of error by the intended user in the handling of the device and, if applicable, in the interpretation of the results.22.3.Devices for use by lay persons shall, where appropriate, include a procedure by which the lay person:can verify that, at the time of use, the device will perform as intended by the manufacturer, andif applicable, is warned if the device has failed to provide a valid result.CHAPTER IIIREQUIREMENTS REGARDING THE INFORMATION SUPPLIED WITH THE DEVICE23.Label and instructions for use23.1.General requirements regarding the information supplied by the manufacturerEach device shall be accompanied by the information needed to identify the device and its manufacturer, and by any safety and performance information relevant to the user, or any other person, as appropriate. Such information may appear on the device itself, on the packaging or in the instructions for use, and shall, if the manufacturer has a website, be made available and kept up to date on the website, taking into account the following:(a)The medium, format, content, legibility, and location of the label and instructions for use shall be appropriate to the particular device, its intended purpose and the technical knowledge, experience, education or training of the intended user(s). In particular, instructions for use shall be written in terms readily understood by the intended user and, where appropriate, supplemented with drawings and diagrams.(b)The information required on the label shall be provided on the device itself. If this is not practicable or appropriate, some or all of the information may appear on the packaging for each unit, and/or on the packaging of multiple devices.(c)Labels shall be provided in a human-readable format and may be supplemented by machine-readable information, such as radio-frequency identification (RFID) or bar codes.(d)Instructions for use shall be provided together with devices. By way of exception, instructions for use shall not be required for class I and class IIa devices if such devices can be used safely without any such instructions and unless otherwise provided for elsewhere in this Section.(e)Where multiple devices are supplied to a single user and/or location, a single copy of the instructions for use may be provided if so agreed by the purchaser who in any case may request further copies to be provided free of charge.(f)Instructions for use may be provided to the user in non-paper format (e.g. electronic) to the extent, and only under the conditions, set out in Regulation (EU) No 207/2012 or in any subsequent implementing rules adopted pursuant to this Regulation.(g)Residual risks which are required to be communicated to the user and/or other person shall be included as limitations, contra-indications, precautions or warnings in the information supplied by the manufacturer.(h)Where appropriate, the information supplied by the manufacturer shall take the form of internationally recognised symbols. Any symbol or identification colour used shall conform to the harmonised standards or CS. In areas for which no harmonised standards or CS exist, the symbols and colours shall be described in the documentation supplied with the device.23.2.Information on the labelThe label shall bear all of the following particulars:(a)the name or trade name of the device;(b)the details strictly necessary for a user to identify the device, the contents of the packaging and, where it is not obvious for the user, the intended purpose of the device;(c)the name, registered trade name or registered trade mark of the manufacturer and the address of its registered place of business;(d)if the manufacturer has its registered place of business outside the Union, the name of the authorised representative and address of the registered place of business of the authorised representative;(e)where applicable, an indication that the device contains or incorporates:a medicinal substance, including a human blood or plasma derivative, ortissues or cells, or their derivatives, of human origin, ortissues or cells of animal origin, or their derivatives, as referred to in Regulation (EU) No 722/2012;(f)where applicable, information labelled in accordance with Section 10.4.5.;(g)the lot number or the serial number of the device preceded by the words LOT NUMBER or SERIAL NUMBER or an equivalent symbol, as appropriate;(h)the UDI carrier referred to in Article 27(4) and Part C of Annex VII;(i)an VI;(i)an unambiguous indication of t the time limit for using or implanting the device safely, expressed at least in terms of year and month, where this is relevant;(j)where there is no indication of the date until when it may be used safely, the date of manufacture. This date of manufacture may be included as part of the lot number or serial number, provided the date is clearly identifiable;(k)an indication of any special storage and/or handling condition that applies;(l)if the device is supplied sterile, an indication of its sterile state and the sterilisation method;(m)warnings or precautions to be taken that need to be brought to the immediate attention of the user of the device, and to any other person. This information may be kept to a minimum in which case more detailed information shall appear in the instructions for use, taking into account the intended users;(n)if the device is intended for single use, an indication of that fact. A manufacturer's indication of single use shall be consistent across the Union;(o)if the device is a single-use device that has been reprocessed, an indication of that fact, the number of reprocessing cycles already performed, and any limitation as regards the number of reprocessing cycles;(p)if the device is custom-made, the words custom-made device;(q)an indication that the device is a medical device. If the device is intended for clinical investigation only, the words exclusively for clinical investigation;(r)in the case of devices that are composed of substances or of combinations of substances that are intended to be introduced into the human body via a body orifice or applied to the skin and that are absorbed by or locally dispersed in the human body, the overall qualitative composition of the device and quantitative information on the main constituent or constituents responsible for achieving the principal intended action;(s)for active implantable devices, the serial number, and for other implantable devices, the serial number or the lot number.23.3.Information on the packaging which maintains the sterile condition of a device (sterile packaging)The following particulars shall appear on the sterile packaging:(a)an indication permitting the sterile packaging to be recognised as such,(b)a declaration that the device is in a sterile condition,(c)the method of sterilisation,(d)the name and address of the manufacturer,(e)a description of the device,(f)if the device is intended for clinical investigations, the words exclusively for clinical investigations,(g)if the device is custom-made, the words custom-made device,(h)the month and year of manufacture,(i)an unambiguous indication of the time limit for using or implanting the device safely expressed at least in terms of year and month, and(j)an instruction to check the instructions for use for what to do if the sterile packaging is damaged or unintentionally opened before use.23.4.Information in the instructions for useThe instructions for use shall contain all of the following particulars:(a)the particulars referred to in points (a), (c), (e), (f), (k), (l), (n) and (r) of Section 23.2;(b)the device's intended purpose with a clear specification of indications, contra-indications, the patient target group or groups, and of the intended users, as appropriate;(c)where applicable, a specification of the clinical benefits to be expected.(d)where applicable, links to the summary of safety and clinical performance referred to in Article 32;(e)the performance characteristics of the device;(f)where applicable, information allowing the healthcare professional to verify if the device is suitable and select the corresponding software and accessories;(g)any residual risks, contra-indications and any undesirable side-effects, including information to be conveyed to the patient in this regard;(h)specifications the user requires to use the device appropriately, e.g. if the device has a measuring function, the degree of accuracy claimed for it;(i)details of any preparatory treatment or handling of the device before it is ready for use or during its use, such as sterilisation, final assembly, calibration, etc., including the levels of disinfection required to ensure patient safety and all available methods for achieving those levels of disinfection;(j)any requirements for special facilities, or special training, or particular qualifications of the device user and/or other persons;(k)the information needed to verify whether the device is properly installed and is ready to perform safely and as intended by the manufacturer, together with, where relevant:details of the nature, and frequency, of preventive and regular maintenance, and of any preparatory cleaning or disinfection,identification of any consumable components and how to replace them,information on any necessary calibration to ensure that the device operates properly and safely during its intended lifetime, andmethods for eliminating the risks encountered by persons involved in installing, calibrating or servicing devices;(l)if the device is supplied sterile, instructions in the event of the sterile packaging being damaged or unintentionally opened before use;(m)if the device is supplied non-sterile with the intention that it is sterilised before use, the appropriate instructions for sterilisation;(n)if the device is reusable, information on the appropriate processes for allowing reuse, including cleaning, disinfection, packaging and, where appropriate, the validated method of re-sterilisation appropriate to the Member State or Member States in which the device has been placed on the market. Information shall be provided to identify when the device should no longer be reused, e.g. signs of material degradation or the maximum number of allowable reuses;(o)an indication, if appropriate, that a device can be reused only if it is reconditioned under the responsibility of the manufacturer to comply with the general safety and performance requirements;(p)if the device bears an indication that it is for single use, information on known characteristics and technical factors known to the manufacturer that could pose a risk if the device were to be re-used. This information shall be based on a specific section of the manufacturer's risk management documentation, where such characteristics and technical factors shall be addressed in detail. If in accordance with point (d) of Section 23.1. no instructions for use are required, this information shall be made available to the user upon request;(q)for devices intended for use together with other devices and/or general purpose equipment:information to identify such devices or equipment, in order to obtain a safe combination, and/orinformation on any known restrictions to combinations of devices and equipment;(r)if the device emits radiation for medical purposes:detailed information as to the nature, type and where appropriate, the intensity and distribution of the emitted radiation,the means of protecting the patient, user, or other person from unintended radiation during use of the device;(s)information that allows the user and/or patient to be informed of any warnings, precautions, contra-indications, measures to be taken and limitations of use regarding the device. That information shall, where relevant, allow the user to brief the patient about any warnings, precautions, contra-indications, measures to be taken and limitations of use regarding the device. The information shall cover, where appropriate:warnings, precautions and/or measures to be taken in the event of malfunction of the device or changes in its performance that may affect safety,warnings, precautions and/or measures to be taken as regards the exposure to reasonably foreseeable external influences or environmental conditions, such as magnetic fields, external electrical and electromagnetic effects, electrostatic discharge, radiation associated with diagnostic or therapeutic procedures, pressure, humidity, or temperature,warnings, precautions and/or measures to be taken as regards the risks of interference posed by the reasonably foreseeable presence of the device during specific diagnostic investigations, evaluations, or therapeutic treatment or other procedures such as electromagnetic interference emitted by the device affecting other equipment,if the device is intended to administer medicinal products, tissues or cells of human or animal origin, or their derivatives, or biological substances, any limitations or incompatibility in the choice of substances to be delivered,warnings, precautions and/or limitations related to the medicinal substance or biological material that is incorporated into the device as an integral part of the device; andprecautions related to materials incorporated into the device that contain or consist of CMR substances or endocrine-disrupting substances, or that could result in sensitisation or an allergic reaction by the patient or user;(t)in the case of devices that are composed of substances or of combinations of substances that are intended to be introduced into the human body and that are absorbed by or locally dispersed in the human body, warnings and precautions, where appropriate, related to the general profile of interaction of the device and its products of metabolism with other devices, medicinal products and other substances as well as contra-indications, undesirable side-effects and risks relating to overdose;(u)in the case of implantable devices, the overall qualitative and quantitative information on the materials and substances to which patients can be exposed;(v)warnings or precautions to be taken in order to facilitate the safe disposal of the device, its accessories and the consumables used with it, if any. This information shall cover, where appropriate:infection or microbial hazards such as explants, needles or surgical equipment contaminated with potentially infectious substances of human origin, andphysical hazards such as from sharps.If in accordance with the point (d) of Section 23.1 no instructions for use are required, this information shall be made available to the user upon request;(w)for devices intended for use by lay persons, the circumstances in which the user should consult a healthcare professional;(x)for the devices covered by this Regulation pursuant to Article 1(2), information regarding the absence of a clinical benefit and the risks related to use of the device;(y)date of issue of the instructions for use or, if they have been revised, date of issue and identifier of the latest revision of the instructions for use;(z)a notice to the user and/or patient that any serious incident that has occurred in relation to the device should be reported to the manufacturer and the competent authority of the Member State in which the user and/or patient is established;(aa)information to be supplied to the patient with an implanted device in accordance with Article 18;(ab)for devices that incorporate electronic programmable systems, including software, or software that are devices in themselves, minimum requirements concerning hardware, IT networks characteristics and IT security measures, including protection against unauthorised access, necessary to run the software as intended.

MODIFIED Annex III — ANNEX III · applies from unchanged

Disputed — seen by the structural diff, not by corpus metadata.

The numbering of the two listed elements changes from 1.1 and 1.2 to 1 and 2, while the surrounding text of the post-market surveillance plan and the PSUR/post-market surveillance report descriptions remains the same. Annex III, v1 Annex III, v2

Within the post-market surveillance plan's list of methods and protocols, the phrase referring to events subject to the trend report is changed to refer instead to incidents subject to the trend report. Annex III, v2

text before / after

32017R074502017R0745-20170505

ANNEX IIITECHNICAL DOCUMENTATION ON POST-MARKET SURVEILLANCEThe technical documentation on post-market surveillance to be drawn up by the manufacturer in accordance with Articles 83 to 86 shall be presented in a clear, organised, readily searchable and unambiguous manner and shall include in particular the elements described in this Annex.1.1.The Annex.1.The post-market surveillance plan drawn up in accordance with Article 84.The manufacturer shall prove in a post-market surveillance plan that it complies with the obligation referred to in Article 83.(a)The post-market surveillance plan shall address the collection and utilization of available information, in particular:information concerning serious incidents, including information from PSURs, and field safety corrective actions;records referring to non-serious incidents and data on any undesirable side-effects;information from trend reporting;relevant specialist or technical literature, databases and/or registers;information, including feedbacks and complaints, provided by users, distributors and importers; andpublicly available information about similar medical devices.(b)The post-market surveillance plan shall cover at least:a proactive and systematic process to collect any information referred to in point (a). The process shall allow a correct characterisation of the performance of the devices and shall also allow a comparison to be made between the device and similar products available on the market;effective and appropriate methods and processes to assess the collected data;suitable indicators and threshold values that shall be used in the continuous reassessment of the benefit-risk analysis and of the risk management as referred to in Section 3 of Annex I;effective and appropriate methods and tools to investigate complaints and analyse market-related experience collected in the field;methods and protocols to manage the events incidents subject to the trend report as provided for in Article 88, including the methods and protocols to be used to establish any statistically significant increase in the frequency or severity of incidents as well as the observation period;methods and protocols to communicate effectively with competent authorities, notified bodies, economic operators and users;reference to procedures to fulfil the manufacturers obligations laid down in Articles 83, 84 and 86;systematic procedures to identify and initiate appropriate measures including corrective actions;effective tools to trace and identify devices for which corrective actions might be necessary; anda PMCF plan as referred to in Part B of Annex XIV, or a justification as to why a PMCF is not applicable.1.2.The applicable.2.The PSUR referred to in Article 86 and the post-market surveillance report referred to in Article 85.

MODIFIED Annex VII — ANNEX VII · applies from unchanged

Disputed — seen by the structural diff, not by corpus metadata.

In section 4.5.2(a), the sampling plan requirement changed from ensuring that all devices covered by the certificate are sampled over the period of validity of the certificate to ensuring that the entire range of devices covered by the certificate is sampled over that period. Annex VII, v1 Annex VII, v2

The rest of section 4, section 4.5 and section 4.5.2 shown in both texts otherwise reads the same, and both texts are cut off before the end of the annex, so no further comparison can be made beyond that point. Annex VII, v1 Annex VII, v2

text before / after

32017R074502017R0745-20170505

ANNEX VIIREQUIREMENTS TO BE MET BY NOTIFIED BODIES1.ORGANISATIONAL AND GENERAL REQUIREMENTS1.1.Legal status and organisational structure1.1.1.Each notified body shall be established under the national law of a Member State, or under the law of a third country with which the Union has concluded an agreement in this respect. Its legal personality and status shall be fully documented. Such documentation shall include information about ownership and the legal or natural persons exercising control over the notified body.1.1.2.If the notified body is a legal entity that is part of a larger organisation, the activities of that organisation as well as its organisational structure and governance, and the relationship with the notified body shall be clearly documented. In such cases, the requirements of Section 1.2 are applicable to both the notified body and the organisation to which it belongs.1.1.3.If a notified body wholly or partly owns legal entities established in a Member State or in a third country or is owned by another legal entity, the activities and responsibilities of those entities, as well as their legal and operational relationships with the notified body, shall be clearly defined and documented. Personnel of those entities performing conformity assessment activities under this Regulation shall be subject to the applicable requirements of this Regulation.1.1.4.The organisational structure, allocation of responsibilities, reporting lines and operation of the notified body shall be such that they ensure that there is confidence in the performance by the notified body and in the results of the conformity assessment activities it conducts.1.1.5.The notified body shall clearly document its organisational structure and the functions, responsibilities and authority of its top-level management and of other personnel who may have an influence upon the performance by the notified body and upon the results of its conformity assessment activities.1.1.6.The notified body shall identify the persons in top-level management that have overall authority and responsibility for each of the following:the provision of adequate resources for conformity assessment activities;the development of procedures and policies for the operation of the notified body;the supervision of implementation of the procedures, policies and quality management systems of the notified body;the supervision of the notified body's finances;the activities and decisions taken by the notified body, including contractual agreements;the delegation of authority to personnel and/or committees, where necessary, for the performance of defined activities;the interaction with the authority responsible for notified bodies and the obligations regarding communications with other competent authorities, the Commission and other notified bodies.1.2.Independence and impartiality1.2.1.The notified body shall be a third-party body that is independent of the manufacturer of the device in relation to which it performs conformity assessment activities. The notified body shall also be independent of any other economic operator having an interest in the device as well as of any competitors of the manufacturer. This does not preclude the notified body from carrying out conformity assessment activities for competing manufacturers.1.2.2.The notified body shall be organised and operated so as to safeguard the independence, objectivity and impartiality of its activities. The notified body shall document and implement a structure and procedures for safeguarding impartiality and for promoting and applying the principles of impartiality throughout its organisation, personnel and assessment activities. Such procedures shall provide for the identification, investigation and resolution of any case in which a conflict of interest may arise, including involvement in consultancy services in the field of devices prior to taking up employment with the notified body. The investigation, outcome and its resolution shall be documented.1.2.3.The notified body, its top-level management and the personnel responsible for carrying out the conformity assessment tasks shall not:(a)be the designer, manufacturer, supplier, installer, purchaser, owner or maintainer of devices which they assess, nor the authorised representative of any of those parties. Such restriction shall not preclude the purchase and use of assessed devices that are necessary for the operations of the notified body and the conduct of the conformity assessment, or the use of such devices for personal purposes;(b)be involved in the design, manufacture or construction, marketing, installation and use, or maintenance of the devices for which they are designated, nor represent the parties engaged in those activities;(c)engage in any activity that may conflict with their independence of judgement or integrity in relation to conformity assessment activities for which they are designated;(d)offer or provide any service which may jeopardise the confidence in their independence, impartiality or objectivity. In particular, they shall not offer or provide consultancy services to the manufacturer, its authorised representative, a supplier or a commercial competitor as regards the design, construction, marketing or maintenance of devices or processes under assessment, and(e)be linked to any organisation which itself provides consultancy services as referred to in point (d). Such restriction does not preclude general training activities that are not client specific and that relate to regulation of devices or to related standards.1.2.4.Involvement in consultancy services in the field of devices prior to taking up employment with a notified body shall be fully documented at the time of employment and potential conflicts of interest shall be monitored and resolved in accordance with this Annex. Personnel who were formerly employed by a specific client, or provided consultancy services in the field of devices to that specific client prior to taking up employment with a notified body, shall not be assigned for conformity assessment activities for that specific client or companies belonging to the same group for a period of three years.1.2.5.The impartiality of notified bodies, of their top-level management and of the assessment personnel shall be guaranteed. The level of the remuneration of the top-level management and assessment personnel of a notified body and subcontractors, involved in assessment activities shall not depend on the results of the assessments. Notified bodies shall make publicly available the declarations of interest of their top-level management.1.2.6.If a notified body is owned by a public entity or institution, independence and absence of any conflict of interest shall be ensured and documented between, on the one hand, the authority responsible for notified bodies and/or the competent authority and, on the other hand, the notified body.1.2.7.The notified body shall ensure and document that the activities of its subsidiaries or subcontractors, or of any associated body, including the activities of its owners do not affect its independence, impartiality or the objectivity of its conformity assessment activities.1.2.8.The notified body shall operate in accordance with a set of consistent, fair and reasonable terms and conditions, taking into account the interests of small and medium-sized enterprises as defined in Recommendation 2003/361/EC in relation to fees.1.2.9.The requirements laid down in this Section in no way preclude exchanges of technical information and regulatory guidance between a notified body and a manufacturer applying for conformity assessment.1.3.Confidentiality1.3.1.The notified body shall have documented procedures in place ensuring that its personnel, committees, subsidiaries, subcontractors, and any associated body or personnel of external bodies respect the confidentiality of the information which comes into its possession during the performance of conformity assessment activities, except when disclosure is required by law.1.3.2.The personnel of a notified body shall observe professional secrecy in carrying out their tasks under this Regulation or any provision of national law giving effect to it, except in relation to the authorities responsible for notified bodies, competent authorities for medical devices in the Member States or the Commission. Proprietary rights shall be protected. The notified body shall have documented procedures in place in respect of the requirements of this Section.1.4.Liability1.4.1.The notified body shall take out appropriate liability insurance for its conformity assessment activities, unless liability is assumed by the Member State in question in accordance with national law or that Member State is directly responsible for the conformity assessment.1.4.2.The scope and overall financial value of the liability insurance shall correspond to the level and geographic scope of activities of the notified body and be commensurate with the risk profile of the devices certified by the notified body. The liability insurance shall cover cases where the notified body may be obliged to withdraw, restrict or suspend certificates.1.5.Financial requirementsThe notified body shall have at its disposal the financial resources required to conduct its conformity assessment activities within its scope of designation and related business operations. It shall document and provide evidence of its financial capacity and its long-term economic viability, taking into account, where relevant, any specific circumstances during an initial start-up phase.1.6.Participation in coordination activities1.6.1.The notified body shall participate in, or ensure that its assessment personnel is informed of, any relevant standardisation activities and in the activities of the notified body coordination group referred to in Article 49 and that its assessment and decision-making personnel are informed of all relevant legislation, guidance and best practice documents adopted in the framework of this Regulation.1.6.2.The notified body shall take into consideration guidance and best practice documents.2.QUALITY MANAGEMENT REQUIREMENTS2.1.The notified body shall establish, document, implement, maintain and operate a quality management system that is appropriate to the nature, area and scale of its conformity assessment activities and is capable of supporting and demonstrating the consistent fulfilment of the requirements of this Regulation.2.2.The quality management system of the notified body shall address at least the following:management system structure and documentation, including policies and objectives for its activities;policies for assignment of activities and responsibilities to personnel;assessment and decision-making processes in accordance with the tasks, responsibilities and role of the notified body's personnel and top-level management;the planning, conduct, evaluation and, if necessary, adaptation of its conformity assessment procedures;control of documents;control of records;management reviews;internal audits;corrective and preventive actions;complaints and appeals; andcontinuous training.Where documents are used in various languages, the notified body shall ensure and control that they have the same content.2.3.The top-level management of the notified body shall ensure that the quality management system is fully understood, implemented and maintained throughout the notified body organisation including subsidiaries and subcontractors involved in conformity assessment activities pursuant to this Regulation.2.4.The notified body shall require all personnel to formally commit themselves by a signature or equivalent to comply with the procedures defined by the notified body. That commitment shall cover aspects relating to confidentiality and to independence from commercial and other interests, and any existing or prior association with clients. The personnel shall be required to complete written statements indicating their compliance with confidentiality, independence and impartiality principles.3.RESOURCE REQUIREMENTS3.1.General3.1.1.Notified bodies shall be capable of carrying out all the tasks falling to them under this Regulation with the highest degree of professional integrity and the requisite competence in the specific field, whether those tasks are carried out by notified bodies themselves or on their behalf and under their responsibility.In particular, notified bodies shall have the necessary personnel and possess or have access to all equipment, facilities and competence needed to perform properly the technical, scientific and administrative tasks entailed in the conformity assessment activities in relation to which they have been designated.Such requirement presupposes at all times and for each conformity assessment procedure and each type of devices in relation to which they have been designated, that the notified body has permanent availability of sufficient administrative, technical and scientific personnel who possess experience and knowledge relating to the relevant devices and the corresponding technologies. Such personnel shall be in sufficient numbers to ensure that the notified body in question can perform the conformity assessment tasks, including the assessment of the medical functionality, clinical evaluations and the performance and safety of devices, for which it has been designated, having regard to the requirements of this Regulation, in particular, those set out in Annex I.A notified body's cumulative competences shall be such as to enable it to assess the types of devices for which it is designated. The notified body shall have sufficient internal competence to critically evaluate assessments conducted by external expertise. Tasks which a notified body is precluded from subcontracting are set out in Section 4.1.Personnel involved in the management of the operation of a notified body's conformity assessment activities for devices shall have appropriate knowledge to set up and operate a system for the selection of assessment and verification staff, for verification of their competence, for authorisation and allocation of their tasks, for organisation of their initial and ongoing training and for the assignment of their duties and the monitoring of those staff, in order to ensure that personnel who carry out and perform assessment and verification operations are competent to fulfil the tasks required of them.The notified body shall identify at least one individual within its top-level management as having overall responsibility for all conformity assessment activities in relation to devices.3.1.2.The notified body shall ensure that personnel involved in conformity assessment activities maintain their qualification and expertise by implementing a system for exchange of experience and a continuous training and education programme.3.1.3.The notified body shall clearly document the extent and limits of duties and responsibilities and the level of authorisation of the personnel, including any subcontractors and external experts, involved in conformity assessment activities and inform those personnel accordingly.3.2.Qualification criteria in relation to personnel3.2.1.The Notified Body shall establish and document qualification criteria and procedures for selection and authorisation of persons involved in conformity assessment activities, including as regards knowledge, experience and other competence required, and the required initial and ongoing training. The qualification criteria shall address the various functions within the conformity assessment process, such as auditing, product evaluation or testing, technical documentation review and decision-making, as well as the devices, technologies and areas, such as biocompatibility, sterilisation, tissues and cells of human and animal origin and clinical evaluation, covered by the scope of designation.3.2.2.The qualification criteria referred to in Section 3.2.1 shall refer to the scope of a notified body's designation in accordance with the scope description used by the Member State for the notification referred to in Article 42(3), providing a sufficient level of detail for the required qualification within the subdivisions of the scope description.Specific qualification criteria shall be defined at least for the assessment of:the pre-clinical evaluation,clinical evaluation,tissues and cells of human and animal origin,functional safety,software,packaging,devices that incorporate as an integral part a medicinal product,devices that are composed of substances or of combinations of substances that are absorbed by or locally dispersed in the human body andthe different types of sterilisation processes.3.2.3.The personnel responsible for establishing qualification criteria and for authorising other personnel to perform specific conformity assessment activities shall be employed by the notified body itself and shall not be external experts or subcontracted. They shall have proven knowledge and experience in all of the following:Union devices legislation and relevant guidance documents;the conformity assessment procedures provided for in this Regulation;a broad base of knowledge of device technologies and the design and manufacture of devices;the notified body's quality management system, related procedures and the required qualification criteria;training relevant to personnel involved in conformity assessment activities in relation to devices;adequate experience in conformity assessments under this Regulation or previously applicable law within a notified body.3.2.4.The notified body shall have permanent availability of personnel with relevant clinical expertise and where possible such personnel shall be employed by the notified body itself. Such personnel shall be integrated throughout the notified body's assessment and decision-making process in order to:identify when specialist input is required for the assessment of the clinical evaluation conducted by the manufacturer and identify appropriately qualified experts;appropriately train external clinical experts in the relevant requirements of this Regulation, CS, guidance and harmonised standards and ensure that the external clinical experts are fully aware of the context and implications of their assessment and the advice they provide;be able to review and scientifically challenge the clinical data contained within the clinical evaluation, and any associated clinical investigations, and appropriately guide external clinical experts in the assessment of the clinical evaluation presented by the manufacturer;be able to scientifically evaluate and, if necessary, challenge the clinical evaluation presented, and the results of the external clinical experts' assessment of the manufacturer's clinical evaluation;be able to ascertain the comparability and consistency of the assessments of clinical evaluations conducted by clinical experts;be able to make an assessment of the manufacturer's clinical evaluation and a clinical judgement of the opinion provided by any external expert and make a recommendation to the notified body's decision maker; andbe able to draw up records and reports demonstrating that the relevant conformity assessment activities have been appropriately carried out.3.2.5.The personnel responsible for carrying out product-related reviews (product reviewers), such as technical documentation reviews or type examination, including aspects such as clinical evaluation, biological safety, sterilisation and software validation, shall have all of the following proven qualifications:successful completion of a university or a technical college degree or equivalent qualification in relevant studies, e.g. medicine, pharmacy, engineering or other relevant sciences;four years' professional experience in the field of healthcare products or related activities, such as in manufacturing, auditing or research, of which two years shall be in the design, manufacture, testing or use of the device or technology to be assessed or related to the scientific aspects to be assessed;knowledge of device legislation, including the general safety and performance requirements set out in Annex I;appropriate knowledge and experience of relevant harmonised standards, CS and guidance documents;appropriate knowledge and experience of risk management and related device standards and guidance documents;appropriate knowledge and experience of clinical evaluation;appropriate knowledge of the devices which they are assessing;appropriate knowledge and experience of the conformity assessment procedures laid down in Annexes IX to XI, in particular of the aspects of those procedures for which they are responsible, and adequate authorisation for carrying out those assessments;the ability to draw up records and reports demonstrating that the relevant conformity assessment activities have been appropriately carried out.3.2.6.The personnel responsible for carrying out audits of the manufacturer's quality management system (site auditors) shall have all of the following proven qualifications:successful completion of a university or a technical college degree or equivalent qualification in relevant studies, such as medicine, pharmacy, engineering or other relevant sciences;four years' professional experience in the field of healthcare products or related activities, such as in manufacturing, auditing or research, of which two years shall be in the area of quality management;appropriate knowledge of devices legislation as well as related harmonised standards, CS and guidance documents;appropriate knowledge and experience of risk management and related device standards and guidance documents;appropriate knowledge of quality management systems and related standards and guidance documents;appropriate knowledge and experience of the conformity assessment procedures laid down in Annexes IX to XI, in particular of the aspects of those procedures for which they are responsible, and adequate authorisation for carrying out those audits;training in auditing techniques enabling them to challenge quality management systems;the ability to draw up records and reports demonstrating that the relevant conformity assessment activities have been appropriately carried out.3.2.7.The personnel with overall responsibility for final reviews and decision-making on certification shall be employed by the notified body itself and shall not be external experts or be subcontracted. Those personnel shall, as a group, have proven knowledge and comprehensive experience of all of the following:devices legislation and relevant guidance documents;the device conformity assessments relevant to this Regulation;the types of qualifications, experience and expertise relevant to device conformity assessment;a broad base of knowledge of device technologies, including sufficient experience of conformity assessment of devices being reviewed for certification, the device industry and the design and manufacture of devices;the notified body's quality management system, related procedures and the required qualifications for personnel involved;the ability to draw up records and reports demonstrating that the conformity assessment activities have been appropriately carried out.3.3.Documentation of qualification, training and authorisation of personnel3.3.1.The notified body shall have a procedure in place to fully document the qualification of each member of personnel involved in conformity assessment activities and the satisfaction of the qualification criteria referred to in Section 3.2. Where in exceptional circumstances the fulfilment of the qualification criteria set out in Section 3.2. cannot be fully demonstrated, the notified body shall justify to the authority responsible for notified bodies the authorisation of those members of personnel to carry out specific conformity assessment activities.3.3.2.For all of its personnel referred to in Sections 3.2.3 to 3.2.7, the notified body shall establish and maintain up to date:a matrix detailing the authorisations and responsibilities of the personnel in respect of conformity assessment activities; andrecords attesting to the required knowledge and experience for the conformity assessment activity for which they are authorised. The records shall contain a rationale for defining the scope of the responsibilities for each of the assessment personnel and records of the conformity assessment activities carried out by each of them.3.4.Subcontractors and external experts3.4.1.Notified bodies may, without prejudice to Section 3.2, subcontract certain clearly defined component parts of a conformity assessment activity.The subcontracting of the auditing of quality management systems or of product related reviews as a whole shall not be permitted; nevertheless parts of those activities may be conducted by subcontractors and external auditors and experts working on behalf of the notified body. The notified body in question shall retain full responsibility for being able to produce appropriate evidence of the competence of subcontractors and experts to fulfil their specific tasks, for making a decision based on a subcontractor's assessment and for the work conducted by subcontractors and experts on its behalf.The following activities may not be subcontracted by notified bodies:review of the qualifications and monitoring of the performance of external experts;auditing and certification activities where the subcontracting in question is to auditing or certification organisations;allocation of work to external experts for specific conformity assessment activities; andfinal review and decision making functions.3.4.2.Where a notified body subcontracts certain conformity assessment activities either to an organisation or an individual, it shall have a policy describing the conditions under which subcontracting may take place, and shall ensure that:the subcontractor meets the relevant requirements of this Annex;subcontractors and external experts do not further subcontract work to organisations or personnel; andthe natural or legal person that applied for conformity assessment has been informed of the requirements referred to in the first and second indent.Any subcontracting or consultation of external personnel shall be properly documented, shall not involve any intermediaries and shall be subject to a written agreement covering, among other things, confidentiality and conflicts of interest. The notified body in question shall take full responsibility for the tasks performed by subcontractors.3.4.3.Where subcontractors or external experts are used in the context of a conformity assessment, in particular regarding novel, invasive and implantable devices or technologies, the notified body in question shall have internal competence in each product area for which it is designated that is adequate for the purpose of leading the overall conformity assessment, verifying the appropriateness and validity of expert opinions and making decisions on certification.3.5.Monitoring of competences, training and exchange of experience3.5.1.The notified body shall establish procedures for the initial evaluation and on-going monitoring of the competence, conformity assessment activities and performance of all internal and external personnel, and subcontractors, involved in conformity assessment activities.3.5.2.Notified bodies shall review at regular intervals, the competence of their personnel, identify training needs and draw up a training plan to maintain the required level of qualification and knowledge of individual personnel. That review shall at a minimum, verify that personnel:are aware of Union and national law in force on devices, relevant harmonised standards, CS, guidance documents and the results of the coordination activities referred to in Section 1.6; andtake part in the internal exchange of experience and the continuous training and education programme referred to in Section 3.1.2.4.PROCESS REQUIREMENTS4.1.GeneralThe notified body shall have in place documented processes and sufficiently detailed procedures for the conduct of each conformity assessment activity for which it is designated, comprising the individual steps from pre-application activities up to decision making and surveillance and taking into account, when necessary, the respective specificities of the devices.The requirements laid down in Sections 4.3, 4.4, 4.7 and 4.8 shall be fulfilled as part of the internal activities of notified bodies and shall not be subcontracted.4.2.Notified body quotations and pre-application activitiesThe notified body shall:(a)publish a publicly available description of the application procedure by which manufacturers can obtain certification from it. That description shall include which languages are acceptable for submission of documentation and for any related correspondence;(b)have documented procedures relating to, and documented details about, fees charged for specific conformity assessment activities and any other financial conditions relating to notified bodies' assessment activities for devices;(c)have documented procedures in relation to advertising of their conformity assessment services. Those procedures shall ensure that advertising or promotional activities in no way imply or are capable of leading to an inference that their conformity assessment will offer manufacturers earlier market access or be quicker, easier or less stringent than that of other notified bodies;(d)have documented procedures requiring the review of pre-application information, including the preliminary verification that the product is covered by this Regulation and its classification, prior to issuing any quotation to the manufacturer relating to a specific conformity assessment; and(e)ensure that all contracts relating to the conformity assessment activities covered by this Regulation are concluded directly between the manufacturer and the notified body and not with any other organisation.4.3.Application review and contractThe notified body shall require a formal application signed by a manufacturer or an authorised representative containing all of the information and the manufacturer's declarations required by the relevant conformity assessment as referred to in Annexes IX to XI.The contract between a notified body and a manufacturer shall take the form of a written agreement signed by both parties. It shall be kept by the notified body. This contract shall have clear terms and conditions and contain obligations that enable the notified body to act as required under this Regulation, including an obligation on the manufacturer to inform the notified body of vigilance reports, the right of the notified body to suspend, restrict or withdraw certificates issued and the duty of the notified body to fulfil its information obligations.The notified body shall have documented procedures to review applications, addressing:(a)the completeness of those applications with respect to the requirements of the relevant conformity assessment procedure, as referred to in the corresponding Annex, under which approval has been sought,(b)the verification of the qualification of products covered by those applications as devices and their respective classifications,(c)whether the conformity assessment procedures chosen by the applicant are applicable to the device in question under this Regulation,(d)the ability of the notified body to assess the application based on its designation, and(e)the availability of sufficient and appropriate resources.The outcome of each review of an application shall be documented. Refusals or withdrawals of applications shall be notified to the electronic system referred to in Article 57 and shall be accessible to other notified bodies.4.4.Allocation of resourcesThe notified body shall have documented procedures to ensure that all conformity assessment activities are conducted by appropriately authorised and qualified personnel who are sufficiently experienced in the evaluation of the devices, systems and processes and related documentation that are subject to conformity assessment.For each application, the notified body shall determine the resources needed and identify one individual responsible for ensuring that the assessment of that application is conducted in accordance with the relevant procedures and for ensuring that the appropriate resources including personnel are utilised for each of the tasks of the assessment. The allocation of tasks required to be carried out as part of the conformity assessment and any changes subsequently made to this allocation shall be documented.4.5.Conformity assessment activities4.5.1.GeneralThe notified body and its personnel shall carry out the conformity assessment activities with the highest degree of professional integrity and the requisite technical and scientific competence in the specific fields.The notified body shall have expertise, facilities and documented procedures that are sufficient to effectively conduct the conformity assessment activities for which the notified body in question is designated, taking account of the relevant requirements set out in Annexes IX to XI, and in particular all of the following requirements:appropriately plan the conduct of each individual project,ensure that the composition of the assessment teams is such that there is sufficient experience in relation to the technology concerned, and that there is continuous objectivity and independence, and to provide for rotation of the members of the assessment team at appropriate intervals,specify the rationale for fixing time limits for completion of conformity assessment activities,assess the manufacturer's technical documentation and the solutions adopted to meet the requirements laid down in Annex I,review the manufacturer's procedures and documentation relating to the evaluation of pre-clinical aspects,review the manufacturer's procedures and documentation relating to clinical evaluation,address the interface between the manufacturer's risk management process and its appraisal and analysis of the pre-clinical and clinical evaluation and to evaluate their relevance for the demonstration of conformity with the relevant requirements in Annex I,carry out the specific procedures referred to in Sections 5.2 to 5.4 of Annex IX,in the case of class IIa or class IIb devices, assess the technical documentation of devices selected on a representative basis,plan and periodically carry out appropriate surveillance audits and assessments, carry out or request certain tests to verify the proper functioning of the quality management system and to perform unannounced on site audits,relating to the sampling of devices, verify that the manufactured device is in conformity with the technical documentation; such requirements shall define the relevant sampling criteria and testing procedure prior to sampling,evaluate and verify a manufacturer's compliance with relevant Annexes.The notified body shall, where relevant, take into consideration available CS, guidance and best practice documents and harmonised standards, even if the manufacturer does not claim to be in compliance.4.5.2.Quality management system auditing(a)As part of the assessment of the quality management system, a notified body shall prior to an audit and in accordance with its documented procedures:assess the documentation submitted in accordance with the relevant conformity assessment Annex, and draw up an audit programme which clearly identifies the number and sequence of activities required to demonstrate complete coverage of a manufacturer's quality management system and to determine whether it meets the requirements of this Regulation,identify links between, and allocation of responsibilities among, the various manufacturing sites, and identify relevant suppliers and/or subcontractors of the manufacturer, and consider the need to specifically audit any of those suppliers or subcontractors or both,clearly define, for each audit identified in the audit programme, the objectives, criteria and scope of the audit, and draw up an audit plan that adequately addresses and takes account of the specific requirements for the devices, technologies and processes involved,draw up and keep up to date, for class IIa and class IIb devices, a sampling plan for the assessment of technical documentation as referred to in Annexes II and III covering the range of such devices covered by the manufacturer's application. That plan shall ensure that all the entire range of devices covered by the certificate are is sampled over the period of validity of the certificate, andselect and assign appropriately qualified and authorised personnel for conducting the individual audits. The respective roles, responsibilities and authorities of the team members shall be clearly defined and documented.(b)Based on the audit programme it has drawn up, the notified body shall, in accordance with its documented procedures:audit the manufacturer's quality management system, in order to verify that the quality management system ensures that the devices covered conform to the relevant provisions of this Regulation which apply to devices at every stage, from design through final quality control to ongoing surveillance, and shall determine whether the requirements of this Regulation are met,based on relevant technical documentation and in order to determine whether the manufacturer meets the requirements referred to in the relevant conformity assessment Annex, review and audit the manufacturer's processes and subsystems, in particular for:design and development,production and process controls,product documentation,purchasing controls including verification of purchased devices,corrective and preventive actions, including for post-market surveillance, andPMCF,and review and audit requirements and provisions adopted by the manufacturer, including those in relation to fulfilling the general safety and performance requirements set out in Annex I.The documentation shall be sampled in such a manner as to reflect the risks associated with the intended use of the device, the complexity of the manufacturing technologies, the range and classes of devices produced and any available post-market surveillance information,if not already covered by the audit programme, audit the control of processes on the premises of the manufacturer's suppliers, when the conformity of finished devices is significantly influenced by the activity of suppliers and, in particular when the manufacturer cannot demonstrate sufficient control over its suppliers,conduct assessments of the technical documentation based on its sampling plan and taking account of Sections 4.5.4. and 4.5.5. for pre-clinical and clinical evaluations, andthe notified body shall ensure that audit findings are appropriately and consistently classified in accordance with the requirements of this Regulation and with relevant standards, or with best practice documents developed or adopted by the MDCG.4.5.3.Product verificationAssessment of the technical documentationFor assessment of the technical documentation conducted in accordance with Chapter II of Annex IX, notified bodies shall have sufficient expertise, facilities and documented procedures for:the allocation of appropriately qualified and authorised personnel for the examination of individual aspects such as use of the device, biocompatibility, clinical evaluation, risk management, and sterilisation, andthe assessment of conformity of the design with this Regulation, and for taking account of Sections 4.5.4. to 4.5.6. That assessment shall include examination of the implementation by manufacturers of incoming, in-process and final checks and the results thereof. If further tests or other evidence is required for the assessment of conformity with the requirements of this Regulation, the notified body in question shall carry out adequate physical or laboratory tests in relation to the device or request the manufacturer to carry out such tests.Type-examinationsThe notified body shall have documented procedures, sufficient expertise and facilities for the type-examination of devices in accordance with Annex X including the capacity to:examine and assess the technical documentation taking account of Sections 4.5.4. to 4.5.6., and verify that the type has been manufactured in conformity with that documentation;establish a test plan identifying all relevant and critical parameters which need to be tested by the notified body or under its responsibility;document its rationale for the selection of those parameters;carry out the appropriate examinations and tests in order to verify that the solutions adopted by the manufacturer meet the general safety and performance requirements set out in Annex I. Such examinations and tests shall include all tests necessary to verify that the manufacturer has in fact applied the relevant standards it has opted to use;agree with the applicant as to where the necessary tests will be performed if they are not to be carried out directly by the notified body; andassume full responsibility for test results. Test reports submitted by the manufacturer shall only be taken into account if they have been issued by conformity assessment bodies which are competent and independent of the manufacturer.Verification by examination and testing of every productThe notified body shall:(a)have documented procedures, sufficient expertise and facilities for the verification by examination and testing of every product in accordance with Part B of Annex XI;(b)establish a test plan identifying all relevant and critical parameters which need to be tested by the notified body or under its responsibility in order to:verify, for class IIb devices, the conformity of the device with the type described in the EU type-examination certificate and with the requirements of this Regulation which apply to those devices,confirm, for class IIa devices, the conformity with the technical documentation referred to in Annexes II and III and with the requirements of this Regulation which apply to those devices;(c)document its rationale for the selection of the parameters referred to in point (b);(d)have documented procedures to carry out the appropriate assessments and tests in order to verify the conformity of the device with the requirements of this Regulation by examining and testing every product as specified in Section 15 of Annex XI;(e)have documented procedures providing for the reaching of an agreement with the applicant concerning when and where necessary tests that are not to be carried out by the notified body itself are to be performed; and(f)assume full responsibility for test results in accordance with documented procedures; test reports submitted by the manufacturer shall only be taken into account if they have been issued by conformity assessment bodies which are competent and independent of the manufacturer.4.5.4.Pre-clinical evaluation assessmentThe notified body shall have documented procedures in place for the review of the manufacturer's procedures and documentation relating to the evaluation of pre-clinical aspects. The notified body shall examine, validate and verify that the manufacturer's procedures and documentation adequately address:(a)the planning, conduct, assessment, reporting and, where appropriate, updating of the pre-clinical evaluation, in particular ofthe scientific pre-clinical literature search, andthe pre-clinical testing, for example laboratory testing, simulated use testing, computer modelling, the use of animal models,(b)the nature and duration of body contact and the specific associated biological risks,(c)the interface with the risk management process, and(d)the appraisal and analysis of the available pre-clinical data and its relevance with regard to demonstrating conformity with the relevant requirements in Annex I.The notified body's assessment of pre-clinical evaluation procedures and documentation shall address the results of literature searches and all validation, verification and testing performed and conclusions drawn, and shall typically include considering the use of alternative materials and substances and take account of the packaging, stability, including shelf life, of the finished device. Where no new testing has been undertaken by a manufacturer or where there are deviations from procedures, the notified body in question shall critically examine the justification presented by the manufacturer.4.5.5.Clinical evaluation assessmentThe notified body shall have documented procedures in place relating to the assessment of a manufacturer's procedures and documentation relating to clinical evaluation both for initial conformity assessment and on an ongoing basis. The notified body shall examine, validate and verify that manufacturers' procedures and documentation adequately address:the planning, conduct, assessment, reporting and updating of the clinical evaluation as referred to in Annex XIV,post-market surveillance and PMCF,the interface with the risk management process,the appraisal and analysis of the available data and its relevance with regard to demonstrating conformity with the relevant requirements in Annex I, andthe conclusions drawn with regard to the clinical evidence and drawing up of the clinical evaluation report.These procedures referred to in the first paragraph shall take into consideration available CS, guidance and best practice documents.The notified body's assessment of clinical evaluations as referred to in Annex XIV shall cover:the intended use specified by the manufacturer and claims for the device defined by it,the planning of the clinical evaluation,the methodology for the literature search,relevant documentation from the literature search,the clinical investigation,validity of equivalence claimed in relation to other devices, the demonstration of equivalence, the suitability and conclusions data from equivalent and similar devices,post-market surveillance and PMCF,the clinical evaluation report, andjustifications in relation to non-performance of clinical investigations or PMCF.In relation to clinical data from clinical investigations included within the clinical evaluation, the notified body in question shall ensure that the conclusions drawn by the manufacturer are valid in the light of the approved clinical investigation plan.The notified body shall ensure that the clinical evaluation adequately addresses the relevant safety and performance requirements provided for in Annex I, that it is appropriately aligned with the risk management requirements, that it is conducted in accordance with Annex XIV and that it is appropriately reflected in the information provided relating to the device.4.5.6.Specific ProceduresThe notified body shall have documented procedures, sufficient expertise and facilities for the procedures referred to in Sections 5 and 6 of Annex IX, Section 6 of Annex X and Section 16 of Annex XI, for which they are designated.In the case of devices manufactured utilising tissues or cells of animal origin or their derivatives, such as from TSE susceptible species, as referred to in Regulation (EU) No 722/2012, the notified body shall have documented procedures in place that fulfil the requirements laid down in that Regulation, including for the preparation of a summary evaluation report for the relevant competent authority.4.6.ReportingThe notified body shall:ensure that all steps of the conformity assessment are documented so that the conclusions of the assessment are clear and demonstrate compliance with the requirements of this Regulation and can represent objective evidence of such compliance to persons that are not themselves involved in the assessment, for example personnel in designating authorities,ensure that records that are sufficient to provide a discernible audit trail are available for quality management system audits,clearly document the conclusions of its assessment of clinical evaluation in a clinical evaluation assessment report, andfor each specific project, provide a detailed report which shall be based on a standard format containing a minimum set of elements determined by the MDCG.The report of the notified body shall:clearly document the outcome of its assessment and draw clear conclusions from the verification of the manufacturer's conformity with the requirements of this Regulation,make a recommendation for a final review and for a final decision to be taken by the notified body; this recommendation shall be signed off by the member of personnel responsible in the notified body, andbe provided to the manufacturer in question.4.7.Final reviewThe notified body shall prior to making a final decision:ensure that the personnel assigned for the final review and decision-making on specific projects are appropriately authorised and are different from the personnel who have conducted the assessments,verify that the report or reports and supporting documentation needed for decision making, including concerning resolution of non-conformities noted during assessment, are complete and sufficient with respect to the scope of the application, andverify whether there are any unresolved non-conformities preventing issuance of a certificate.4.8.Decisions and CertificationsThe notified body shall have documented procedures for decision-making including as regards the allocation of responsibilities for the issuance, suspension, restriction and withdrawal of certificates. Those procedures shall include the notification requirements laid down in Chapter V of this Regulation. The procedures shall allow the notified body in question to:decide, based on the assessment documentation and additional information available, whether the requirements of this Regulation are fulfilled,decide, based on the results of its assessment of the clinical evaluation and risk management, whether the post-market surveillance plan, including the PMCF plan, is adequate,decide on specific milestones for further review by the notified body of the up to date clinical evaluation,decide whether specific conditions or provisions need to be defined for the certification,decide, based on the novelty, risk classification, clinical evaluation and conclusions from the risk analysis of the device, on a period of certification not exceeding five years,clearly document decision making and approval steps including approval by signature of the members of personnel responsible,clearly document responsibilities and mechanisms for communication of decisions, in particular, where the final signatory of a certificate differs from the decision maker or decision makers or does not fulfil the requirements laid down in Section 3.2.7,issue a certificate or certificates in accordance with the minimum requirements laid down in Annex XII for a period of validity not exceeding five years and shall indicate whether there are specific conditions or limitations associated with the certification,issue a certificate or certificates for the applicant alone and shall not issue certificates covering multiple entities, andensure that the manufacturer is notified of the outcome of the assessment and the resultant decision and that they are entered into the electronic system referred to in Article 57.4.9.Changes and modificationsThe notified body shall have documented procedures and contractual arrangements with manufacturers in place relating to the manufacturers' information obligations and the assessment of changes to:the approved quality management system or systems or to the product-range covered,the approved design of a device,the intended use of or claims made for the device,the approved type of a device, andany substance incorporated in or utilised for the manufacturing of a device and being subject to the specific procedures in accordance with Section 4.5.6.The procedures and contractual arrangements referred to in the first paragraph shall include measures for checking the significance of the changes referred to in the first paragraph.In accordance with its documented procedures, the notified body in question shall:ensure that manufacturers submit for prior approval plans for changes as referred to in the first paragraph and relevant information relating to such changes,assess the changes proposed and verify whether, after these changes, the quality management system, or the design of a device or type of a device, still meets the requirements of this Regulation, andnotify the manufacturer of its decision and provide a report or as applicable a supplementary report, which shall contain the justified conclusions of its assessment.4.10.Surveillance activities and post-certification monitoringThe notified body shall have documented procedures:defining how and when surveillance activities of manufacturers are to be conducted. Those procedures shall include arrangements for unannounced on-site audits of manufacturers and, where applicable, subcontractors and suppliers carrying out product tests and the monitoring of compliance with any conditions binding manufacturers and associated with certification decisions, such as updates to clinical data at defined intervals,for screening relevant sources of scientific and clinical data and post-market information relating to the scope of their designation. Such information shall be taken into account in the planning and conduct of surveillance activities, andto review vigilance data to which they have access under Article 92(2) in order to estimate its impact, if any, on the validity of existing certificates. The results of the evaluation and any decisions taken shall be thoroughly documented.The notified body in question shall, upon receipt of information about vigilance cases from a manufacturer or competent authorities, decide which of the following options to apply:not to take action on the basis that the vigilance case is clearly not related to the certification granted,observe the manufacturer's and competent authority's activities and the results of the manufacturer's investigation so as to determine whether the certification granted is at risk or whether adequate corrective action has been taken,perform extraordinary surveillance measures, such as document reviews, short-notice or unannounced audits and product testing, where it is likely that the certification granted is at risk,increase the frequency of surveillance audits,review specific products or processes on the occasion of the next audit of the manufacturer, ortake any other relevant measure.In relation to surveillance audits of manufacturers, the notified body shall have documented procedures to:conduct surveillance audits of the manufacturer on at least an annual basis which shall be planned and conducted in line with the relevant requirements in Section 4.5,ensure adequate assessment of the manufacturer's documentation on, and application of the provisions on, vigilance, the post-market surveillance, and PMCF,sample and test devices and technical documentation, during audits, according to pre-defined sampling criteria and testing procedures to ensure that the manufacturer continuously applies the approved quality management system,ensure that the manufacturer complies with the documentation and information obligations laid down in the relevant Annexes and that its procedures take into account best practices in the implementation of quality management systems,ensure that the manufacturer does not use quality management system or device approvals in a misleading manner,gather sufficient information to determine if the quality management system continues to comply with the requirements of this Regulation,ask the manufacturer, if non-conformities are detected, for corrections, corrective actions and, where applicable, preventive actions, andwhere necessary, impose specific restrictions on the relevant certificate, or suspend or withdraw it.The notified body shall, if listed as part of the conditions for certification:conduct an in-depth review of the clinical evaluation as most recently updated by the manufacturer based on the manufacturer's post-market surveillance, on its PMCF and on clinical literature relevant to the condition being treated with the device or on clinical literature relevant to similar devices,clearly document the outcome of the in-depth review and address any specific concerns to the manufacturer or impose any specific conditions on it, andensure that the clinical evaluation as most recently updated, is appropriately reflected in the instructions for use and, where applicable, the summary of safety and performance.4.11.Re-certificationThe notified body shall have documented procedures in place relating to the re-certification reviews and the renewal of certificates. Re-certification of approved quality management systems or EU technical documentation assessment certificates or EU type-examination certificates shall occur at least every five years.The notified body shall have documented procedures relating to renewals of EU technical documentation assessment certificates and EU type-examination certificates and those procedures shall require the manufacturer in question to submit a summary of changes and scientific findings for the device, including:(a)all changes to the originally approved device, including changes not yet notified,(b)experience gained from post-market surveillance,(c)experience from risk management,(d)experience from updating the proof of compliance with the general safety and performance requirements set out in Annex I,(e)experience from reviews of the clinical evaluation, including the results of any clinical investigations and PMCF,(f)changes to the requirements, to components of the device or to the scientific or regulatory environment,(g)changes to applied or new harmonised standards, CS or equivalent documents, and(h)changes in medical, scientific and technical knowledge, such as:new treatments,changes in test methods,new scientific findings on materials and components, including findings on their biocompatibility,experience from studies on comparable devices,data from registers and registries,experience from clinical investigations with comparable devices.The notified body shall have documented procedures to assess the information referred to in the second paragraph and shall pay particular attention to clinical data from post-market surveillance and PMCF activities undertaken since the previous certification or re-certification, including appropriate updates to manufacturers' clinical evaluation reports.For the decision on re-certification, the notified body in question shall use the same methods and principles as for the initial certification decision. If necessary, separate forms shall be established for re-certification taking into account the steps taken for certification such as application and application review.

MODIFIED Annex VIII — ANNEX VIII · applies from unchanged

Disputed — seen by the structural diff, not by corpus metadata.

Section 3.2 no longer mentions accessories for a product listed in Annex XVI, referring only to accessories for a medical device being classified in their own right separately from the device with which they are used. Annex VIII, v1 Annex VIII, v2

text before / after

32017R074502017R0745-20170505

ANNEX VIIICLASSIFICATION RULESCHAPTER IDEFINITIONS SPECIFIC TO CLASSIFICATION RULES1.DURATION OF USE1.1.Transient means normally intended for continuous use for less than 60 minutes.1.2.Short term means normally intended for continuous use for between 60 minutes and 30 days.1.3.Long term means normally intended for continuous use for more than 30 days.2.INVASIVE AND ACTIVE DEVICES2.1.Body orifice means any natural opening in the body, as well as the external surface of the eyeball, or any permanent artificial opening, such as a stoma.2.2.Surgically invasive device means:(a)an invasive device which penetrates inside the body through the surface of the body, including through mucous membranes of body orifices with the aid or in the context of a surgical operation; and(b)a device which produces penetration other than through a body orifice.2.3.Reusable surgical instrument means an instrument intended for surgical use in cutting, drilling, sawing, scratching, scraping, clamping, retracting, clipping or similar procedures, without a connection to an active device and which is intended by the manufacturer to be reused after appropriate procedures such as cleaning, disinfection and sterilisation have been carried out.2.4.Active therapeutic device means any active device used, whether alone or in combination with other devices, to support, modify, replace or restore biological functions or structures with a view to treatment or alleviation of an illness, injury or disability.2.5.Active device intended for diagnosis and monitoring means any active device used, whether alone or in combination with other devices, to supply information for detecting, diagnosing, monitoring or treating physiological conditions, states of health, illnesses or congenital deformities.2.6.Central circulatory system means the following blood vessels: arteriae pulmonales, aorta ascendens, arcus aortae, aorta descendens to the bifurcatio aortae, arteriae coronariae, arteria carotis communis, arteria carotis externa, arteria carotis interna, arteriae cerebrales, truncus brachiocephalicus, venae cordis, venae pulmonales, vena cava superior and vena cava inferior.2.7.Central nervous system means the brain, meninges and spinal cord.2.8.Injured skin or mucous membrane means an area of skin or a mucous membrane presenting a pathological change or change following disease or a wound.CHAPTER IIIMPLEMENTING RULES3.1.Application of the classification rules shall be governed by the intended purpose of the devices.3.2.If the device in question is intended to be used in combination with another device, the classification rules shall apply separately to each of the devices. Accessories for a medical device and for a product listed in Annex XVI shall be classified in their own right separately from the device with which they are used.3.3.Software, which drives a device or influences the use of a device, shall fall within the same class as the device.If the software is independent of any other device, it shall be classified in its own right.3.4.If the device is not intended to be used solely or principally in a specific part of the body, it shall be considered and classified on the basis of the most critical specified use.3.5.If several rules, or if, within the same rule, several sub-rules, apply to the same device based on the device's intended purpose, the strictest rule and sub-rule resulting in the higher classification shall apply.3.6.In calculating the duration referred to in Section 1, continuous use shall mean:(a)the entire duration of use of the same device without regard to temporary interruption of use during a procedure or temporary removal for purposes such as cleaning or disinfection of the device. Whether the interruption of use or the removal is temporary shall be established in relation to the duration of the use prior to and after the period when the use is interrupted or the device removed; and(b)the accumulated use of a device that is intended by the manufacturer to be replaced immediately with another of the same type.3.7.A device is considered to allow direct diagnosis when it provides the diagnosis of the disease or condition in question by itself or when it provides decisive information for the diagnosis.CHAPTER IIICLASSIFICATION RULES4.NON-INVASIVE DEVICES4.1.Rule 1All non-invasive devices are classified as class I, unless one of the rules set out hereinafter applies.4.2.Rule 2All non-invasive devices intended for channelling or storing blood, body liquids, cells or tissues, liquids or gases for the purpose of eventual infusion, administration or introduction into the body are classified as class IIa:if they may be connected to a class IIa, class IIb or class III active device; orif they are intended for use for channelling or storing blood or other body liquids or for storing organs, parts of organs or body cells and tissues, except for blood bags; blood bags are classified as class IIb.In all other cases, such devices are classified as class I.4.3.Rule 3All non-invasive devices intended for modifying the biological or chemical composition of human tissues or cells, blood, other body liquids or other liquids intended for implantation or administration into the body are classified as class IIb, unless the treatment for which the device is used consists of filtration, centrifugation or exchanges of gas, heat, in which case they are classified as class IIa.All non-invasive devices consisting of a substance or a mixture of substances intended to be used in vitro in direct contact with human cells, tissues or organs taken from the human body or used in vitro with human embryos before their implantation or administration into the body are classified as class III.4.4.Rule 4All non-invasive devices which come into contact with injured skin or mucous membrane are classified as:class I if they are intended to be used as a mechanical barrier, for compression or for absorption of exudates;class IIb if they are intended to be used principally for injuries to skin which have breached the dermis or mucous membrane and can only heal by secondary intent;class IIa if they are principally intended to manage the micro-environment of injured skin or mucous membrane; andclass IIa in all other cases.This rule applies also to the invasive devices that come into contact with injured mucous membrane.5.INVASIVE DEVICES5.1.Rule 5All invasive devices with respect to body orifices, other than surgically invasive devices, which are not intended for connection to an active device or which are intended for connection to a class I active device are classified as:class I if they are intended for transient use;class IIa if they are intended for short-term use, except if they are used in the oral cavity as far as the pharynx, in an ear canal up to the ear drum or in the nasal cavity, in which case they are classified as class I; andclass IIb if they are intended for long-term use, except if they are used in the oral cavity as far as the pharynx, in an ear canal up to the ear drum or in the nasal cavity and are not liable to be absorbed by the mucous membrane, in which case they are classified as class IIa.All invasive devices with respect to body orifices, other than surgically invasive devices, intended for connection to a class IIa, class IIb or class III active device, are classified as class IIa.5.2.Rule 6All surgically invasive devices intended for transient use are classified as class IIa unless they:are intended specifically to control, diagnose, monitor or correct a defect of the heart or of the central circulatory system through direct contact with those parts of the body, in which case they are classified as class III;are reusable surgical instruments, in which case they are classified as class I;are intended specifically for use in direct contact with the heart or central circulatory system or the central nervous system, in which case they are classified as class III;are intended to supply energy in the form of ionising radiation in which case they are classified as class IIb;have a biological effect or are wholly or mainly absorbed in which case they are classified as class IIb; orare intended to administer medicinal products by means of a delivery system, if such administration of a medicinal product is done in a manner that is potentially hazardous taking account of the mode of application, in which case they are classified as class IIb.5.3.Rule 7All surgically invasive devices intended for short-term use are classified as class IIa unless they:are intended specifically to control, diagnose, monitor or correct a defect of the heart or of the central circulatory system through direct contact with those parts of the body, in which case they are classified as class III;are intended specifically for use in direct contact with the heart or central circulatory system or the central nervous system, in which case they are classified as class III;are intended to supply energy in the form of ionizing radiation in which case they are classified as class IIb;have a biological effect or are wholly or mainly absorbed in which case they are classified as class III;are intended to undergo chemical change in the body in which case they are classified as class IIb, except if the devices are placed in the teeth; orare intended to administer medicines, in which case they are classified as class IIb.5.4.Rule 8All implantable devices and long-term surgically invasive devices are classified as class IIb unless they:are intended to be placed in the teeth, in which case they are classified as class IIa;are intended to be used in direct contact with the heart, the central circulatory system or the central nervous system, in which case they are classified as class III;have a biological effect or are wholly or mainly absorbed, in which case they are classified as class III;are intended to undergo chemical change in the body in which case they are classified as class III, except if the devices are placed in the teeth;are intended to administer medicinal products, in which case they are classified as class III;are active implantable devices or their accessories, in which cases they are classified as class III;are breast implants or surgical meshes, in which cases they are classified as class III;are total or partial joint replacements, in which case they are classified as class III, with the exception of ancillary components such as screws, wedges, plates and instruments; orare spinal disc replacement implants or are implantable devices that come into contact with the spinal column, in which case they are classified as class III with the exception of components such as screws, wedges, plates and instruments.6.ACTIVE DEVICES6.1.Rule 9All active therapeutic devices intended to administer or exchange energy are classified as class IIa unless their characteristics are such that they may administer energy to or exchange energy with the human body in a potentially hazardous way, taking account of the nature, the density and site of application of the energy, in which case they are classified as class IIb.All active devices intended to control or monitor the performance of active therapeutic class IIb devices, or intended directly to influence the performance of such devices are classified as class IIb.All active devices intended to emit ionizing radiation for therapeutic purposes, including devices which control or monitor such devices, or which directly influence their performance, are classified as class IIb.All active devices that are intended for controlling, monitoring or directly influencing the performance of active implantable devices are classified as class III.6.2.Rule 10Active devices intended for diagnosis and monitoring are classified as class IIa:if they are intended to supply energy which will be absorbed by the human body, except for devices intended to illuminate the patient's body, in the visible spectrum, in which case they are classified as class I;if they are intended to image in vivo distribution of radiopharmaceuticals; orif they are intended to allow direct diagnosis or monitoring of vital physiological processes, unless they are specifically intended for monitoring of vital physiological parameters and the nature of variations of those parameters is such that it could result in immediate danger to the patient, for instance variations in cardiac performance, respiration, activity of the central nervous system, or they are intended for diagnosis in clinical situations where the patient is in immediate danger, in which cases they are classified as class IIb.Active devices intended to emit ionizing radiation and intended for diagnostic or therapeutic radiology, including interventional radiology devices and devices which control or monitor such devices, or which directly influence their performance, are classified as class IIb.6.3.Rule 11Software intended to provide information which is used to take decisions with diagnosis or therapeutic purposes is classified as class IIa, except if such decisions have an impact that may cause:death or an irreversible deterioration of a person's state of health, in which case it is in class III; ora serious deterioration of a person's state of health or a surgical intervention, in which case it is classified as class IIb.Software intended to monitor physiological processes is classified as class IIa, except if it is intended for monitoring of vital physiological parameters, where the nature of variations of those parameters is such that it could result in immediate danger to the patient, in which case it is classified as class IIb.All other software is classified as class I.6.4.Rule 12All active devices intended to administer and/or remove medicinal products, body liquids or other substances to or from the body are classified as class IIa, unless this is done in a manner that is potentially hazardous, taking account of the nature of the substances involved, of the part of the body concerned and of the mode of application in which case they are classified as class IIb.6.5.Rule 13All other active devices are classified as class I.7.SPECIAL RULES7.1.Rule 14All devices incorporating, as an integral part, a substance which, if used separately, can be considered to be a medicinal product, as defined in point 2 of Article 1 of Directive 2001/83/EC, including a medicinal product derived from human blood or human plasma, as defined in point 10 of Article 1 of that Directive, and that has an action ancillary to that of the devices, are classified as class III.7.2.Rule 15All devices used for contraception or prevention of the transmission of sexually transmitted diseases are classified as class IIb, unless they are implantable or long term invasive devices, in which case they are classified as class III.7.3.Rule 16All devices intended specifically to be used for disinfecting, cleaning, rinsing or, where appropriate, hydrating contact lenses are classified as class IIb.All devices intended specifically to be used for disinfecting or sterilising medical devices are classified as class IIa, unless they are disinfecting solutions or washer-disinfectors intended specifically to be used for disinfecting invasive devices, as the end point of processing, in which case they are classified as class IIb.This rule does not apply to devices that are intended to clean devices other than contact lenses by means of physical action only.7.4.Rule 17Devices specifically intended for recording of diagnostic images generated by X-ray radiation are classified as class IIa.7.5.Rule 18All devices manufactured utilising tissues or cells of human or animal origin, or their derivatives, which are non-viable or rendered non-viable, are classified as class III, unless such devices are manufactured utilising tissues or cells of animal origin, or their derivatives, which are non-viable or rendered non-viable and are devices intended to come into contact with intact skin only.7.6.Rule 19All devices incorporating or consisting of nanomaterial are classified as:class III if they present a high or medium potential for internal exposure;class IIb if they present a low potential for internal exposure; andclass IIa if they present a negligible potential for internal exposure.7.7.Rule 20All invasive devices with respect to body orifices, other than surgically invasive devices, which are intended to administer medicinal products by inhalation are classified as class IIa, unless their mode of action has an essential impact on the efficacy and safety of the administered medicinal product or they are intended to treat life-threatening conditions, in which case they are classified as class IIb.7.8.Rule 21Devices that are composed of substances or of combinations of substances that are intended to be introduced into the human body via a body orifice or applied to the skin and that are absorbed by or locally dispersed in the human body are classified as:class III if they, or their products of metabolism, are systemically absorbed by the human body in order to achieve the intended purpose;class III if they achieve their intended purpose in the stomach or lower gastrointestinal tract and they, or their products of metabolism, are systemically absorbed by the human body;class IIa if they are applied to the skin or if they are applied in the nasal or oral cavity as far as the pharynx, and achieve their intended purpose on those cavities; andclass IIb in all other cases.7.9.Rule 22Active therapeutic devices with an integrated or incorporated diagnostic function which significantly determines the patient management by the device, such as closed loop systems or automated external defibrillators, are classified as class III.

MODIFIED Annex IX — ANNEX IX · applies from unchanged

Disputed — seen by the structural diff, not by corpus metadata.

Section 2.3 now refers to representative-sample device selection "as specified in Section 4" instead of citing Sections 4.4 to 4.8, and Section 3.5 makes the same change, also updating its internal cross-reference from the second paragraph to the third paragraph of Section 2.3. Annex IX, v1 Annex IX, v2

The heading of Section 3 drops the phrase "applicable to class IIa, class IIb and class III devices", leaving it simply titled "Surveillance assessment". Annex IX, v1 Annex IX, v2

Section 4.3 changes the wording from the notified body examining the application by using staff employed by it, to the notified body assessing the technical documentation using staff, removing the words "employed by it". Annex IX, v1 Annex IX, v2

text before / after

32017R074502017R0745-20170505

ANNEX IXCONFORMITY ASSESSMENT BASED ON A QUALITY MANAGEMENT SYSTEM AND ON ASSESSMENT OF TECHNICAL DOCUMENTATIONCHAPTER IQUALITY MANAGEMENT SYSTEM1.The manufacturer shall establish, document and implement a quality management system as described in Article 10(9) and maintain its effectiveness throughout the life cycle of the devices concerned. The manufacturer shall ensure the application of the quality management system as specified in Section 2 and shall be subject to audit, as laid down in Sections 2.3 and 2.4, and to surveillance as specified in Section 3.2.Quality management system assessment2.1.The manufacturer shall lodge an application for assessment of its quality management system with a notified body. The application shall include:the name of the manufacturer and address of its registered place of business and any additional manufacturing site covered by the quality management system, and, if the manufacturer's application is lodged by its authorised representative, the name of the authorised representative and the address of the authorised representative's registered place of business,all relevant information on the device or group of devices covered by the quality management system,a written declaration that no application has been lodged with any other notified body for the same device-related quality management system, or information about any previous application for the same device-related quality management system,a draft of an EU declaration of conformity in accordance with Article 19 and Annex IV for the device model covered by the conformity assessment procedure,the documentation on the manufacturer's quality management system,a documented description of the procedures in place to fulfil the obligations arising from the quality management system and required under this Regulation and the undertaking by the manufacturer in question to apply those procedures,a description of the procedures in place to ensure that the quality management system remains adequate and effective, and the undertaking by the manufacturer to apply those procedures,the documentation on the manufacturer's post-market surveillance system and, where applicable, on the PMCF plan, and the procedures put in place to ensure compliance with the obligations resulting from the provisions on vigilance set out in Articles 87 to 92,a description of the procedures in place to keep up to date the post-market surveillance system, and, where applicable, the PMCF plan, and the procedures ensuring compliance with the obligations resulting from the provisions on vigilance set out in Articles 87 to 92, as well as the undertaking by the manufacturer to apply those procedures,documentation on the clinical evaluation plan, anda description of the procedures in place to keep up to date the clinical evaluation plan, taking into account the state of the art.2.2.Implementation of the quality management system shall ensure compliance with this Regulation. All the elements, requirements and provisions adopted by the manufacturer for its quality management system shall be documented in a systematic and orderly manner in the form of a quality manual and written policies and procedures such as quality programmes, quality plans and quality records.Moreover, the documentation to be submitted for the assessment of the quality management system shall include an adequate description of, in particular:(a)the manufacturer's quality objectives;(b)the organisation of the business and in particular:the organisational structures with the assignment of staff responsibilities in relation to critical procedures, the responsibilities of the managerial staff and their organisational authority,the methods of monitoring whether the operation of the quality management system is efficient and in particular the ability of that system to achieve the desired design and device quality, including control of devices which fail to conform,where the design, manufacture and/or final verification and testing of the devices, or parts of any of those processes, is carried out by another party, the methods of monitoring the efficient operation of the quality management system and in particular the type and extent of control applied to the other party, andwhere the manufacturer does not have a registered place of business in a Member State, the draft mandate for the designation of an authorised representative and a letter of intention from the authorised representative to accept the mandate;(c)the procedures and techniques for monitoring, verifying, validating and controlling the design of the devices and the corresponding documentation as well as the data and records arising from those procedures and techniques. Those procedures and techniques shall specifically cover:the strategy for regulatory compliance, including processes for identification of relevant legal requirements, qualification, classification, handling of equivalence, choice of and compliance with conformity assessment procedures,identification of applicable general safety and performance requirements and solutions to fulfil those requirements, taking applicable CS and, where opted for, harmonised standards or other adequate solutions into account,risk management as referred to in Section 3 of Annex I,the clinical evaluation, pursuant to Article 61 and Annex XIV, including post-market clinical follow-up,solutions for fulfilling the applicable specific requirements regarding design and construction, including appropriate pre-clinical evaluation, in particular the requirements of Chapter II of Annex I,solutions for fulfilling the applicable specific requirements regarding the information to be supplied with the device, in particular the requirements of Chapter III of Annex I,the device identification procedures drawn up and kept up to date from drawings, specifications or other relevant documents at every stage of manufacture, andmanagement of design or quality management system changes; and(d)the verification and quality assurance techniques at the manufacturing stage and in particular the processes and procedures which are to be used, particularly as regards sterilisation and the relevant documents; and(e)the appropriate tests and trials which are to be carried out before, during and after manufacture, the frequency with which they are to take place, and the test equipment to be used; it shall be possible to trace back adequately the calibration of that test equipment.In addition, the manufacturer shall grant the notified body access to the technical documentation referred to in Annexes II and III.2.3.AuditThe notified body shall audit the quality management system to determine whether it meets the requirements referred to in Section 2.2. Where the manufacturer uses a harmonised standard or CS related to a quality management system, the notified body shall assess conformity with those standards or CS. The notified body shall assume that a quality management system which satisfies the relevant harmonised standards or CS conforms to the requirements covered by those standards or CS, unless it duly substantiates not doing so.The audit team of the notified body shall include at least one member with past experience of assessments of the technology concerned in accordance with Sections 4.3. to 4.5. of Annex VII. In circumstances where such experience is not immediately obvious or applicable, the notified body shall provide a documented rationale for the composition of that team. The assessment procedure shall include an audit on the manufacturer's premises and, if appropriate, on the premises of the manufacturer's suppliers and/or subcontractors to verify the manufacturing and other relevant processes.Moreover, in the case of class IIa and class IIb devices, the quality management system assessment shall be accompanied by the assessment of technical documentation for devices selected on a representative basis as specified in accordance with Sections 4.4 to 4.8. Section 4. In choosing representative samples, the notified body shall take into account the published guidance developed by the MDCG pursuant to Article 105 and in particular the novelty of the technology, similarities in design, technology, manufacturing and sterilisation methods, the intended purpose and the results of any previous relevant assessments such as with regard to physical, chemical, biological or clinical properties, that have been carried out in accordance with this Regulation. The notified body in question shall document its rationale for the samples taken.If the quality management system conforms to the relevant provisions of this Regulation, the notified body shall issue an EU quality management system certificate. The notified body shall notify the manufacturer of its decision to issue the certificate. The decision shall contain the conclusions of the audit and a reasoned report.2.4.The manufacturer in question shall inform the notified body which approved the quality management system of any plan for substantial changes to the quality management system, or the device-range covered. The notified body shall assess the changes proposed, determine the need for additional audits and verify whether after those changes the quality management system still meets the requirements referred to in Section 2.2. It shall notify the manufacturer of its decision which shall contain the conclusions of the assessment, and where applicable, conclusions of additional audits. The approval of any substantial change to the quality management system or the device-range covered shall take the form of a supplement to the EU quality management system certificate.3.Surveillance assessment applicable to class IIa, class IIb and class III devices3.1.The assessment3.1.The aim of surveillance is to ensure that the manufacturer duly fulfils the obligations arising from the approved quality management system.3.2.The manufacturer shall give authorisation to the notified body to carry out all the necessary audits, including on-site audits, and supply it with all relevant information, in particular:the documentation on its quality management system,documentation on any findings and conclusions resulting from the application of the post-market surveillance plan, including the PMCF plan, for a representative sample of devices, and of the provisions on vigilance set out in Articles 87 to 92,the data stipulated in the part of the quality management system relating to design, such as the results of analyses, calculations, tests and the solutions adopted regarding the risk-management as referred to in Section 4 of Annex I, andthe data stipulated in the part of the quality management system relating to manufacture, such as quality control reports and test data, calibration data, and records on the qualifications of the personnel concerned.3.3.Notified bodies shall periodically, at least once every 12 months, carry out appropriate audits and assessments to make sure that the manufacturer in question applies the approved quality management system and the post-market surveillance plan. Those audits and assessments shall include audits on the premises of the manufacturer and, if appropriate, of the manufacturer's suppliers and/or subcontractors. At the time of such on-site audits, the notified body shall, where necessary, carry out or ask for tests in order to check that the quality management system is working properly. It shall provide the manufacturer with a surveillance audit report and, if a test has been carried out, with a test report.3.4.The notified body shall randomly perform at least once every five years unannounced audits on the site of the manufacturer and, where appropriate, of the manufacturer's suppliers and/or subcontractors, which may be combined with the periodic surveillance assessment referred to in Section 3.3. or be performed in addition to that surveillance assessment. The notified body shall establish a plan for such unannounced on-site audits but shall not disclose it to the manufacturer.Within the context of such unannounced on-site audits, the notified body shall test an adequate sample of the devices produced or an adequate sample from the manufacturing process to verify that the manufactured device is in conformity with the technical documentation, with the exception of the devices referred to in the second subparagraph of Article 52(8). Prior to unannounced on-site audits, the notified body shall specify the relevant sampling criteria and testing procedure.Instead of, or in addition to, sampling referred to in the second paragraph, the notified body shall take samples of devices from the market to verify that the manufactured device is in conformity with the technical documentation, with the exception of the devices referred to in the second subparagraph of Article 52(8). Prior to the sampling, the notified body in question shall specify the relevant sampling criteria and testing procedure.The notified body shall provide the manufacturer in question with an on-site audit report which shall include, if applicable, the result of the sample test.3.5.In the case of class IIa and class IIb devices, the surveillance assessment shall also include an assessment of the technical documentation as referred to specified in Sections 4.4 to 4.8 Section 4 for the device or devices concerned on the basis of further representative samples chosen in accordance with the rationale documented by the notified body in accordance with the second third paragraph of Section 2.3.In the case of class III devices, the surveillance assessment shall also include a test of the approved parts and/or materials that are essential for the integrity of the device, including, where appropriate, a check that the quantities of produced or purchased parts and/or materials correspond to the quantities of finished devices.3.6.The notified body shall ensure that the composition of the assessment team is such that there is sufficient experience with the evaluation of the devices, systems and processes concerned, continuous objectivity and neutrality; this shall include a rotation of the members of the assessment team at appropriate intervals. As a general rule, a lead auditor shall neither lead nor attend audits for more than three consecutive years in respect of the same manufacturer.3.7.If the notified body finds a divergence between the sample taken from the devices produced or from the market and the specifications laid down in the technical documentation or the approved design, it shall suspend or withdraw the relevant certificate or impose restrictions on it.CHAPTER IIASSESSMENT OF THE TECHNICAL DOCUMENTATION4.Assessment of the technical documentation applicable to class III devices and to the class IIb devices referred to in the second subparagraph of Article 52(4)4.1.In addition to the obligations laid down in Section 2, the manufacturer shall lodge with the notified body an application for assessment of the technical documentation relating to the device which it plans to place on the market or put into service and which is covered by the quality management system referred to in Section 2.4.2.The application shall describe the design, manufacture and performance of the device in question. It shall include the technical documentation as referred to in Annexes II and III.4.3.The notified body shall examine assess the application by technical documentation using staff, employed by it, staff with proven knowledge and experience regarding the technology concerned and its clinical application. The notified body may require the application to be completed by having further tests carried out or requesting further evidence to be provided to allow assessment of conformity with the relevant requirements of the Regulation. The notified body shall carry out adequate physical or laboratory tests in relation to the device or request the manufacturer to carry out such tests.4.4.The notified body shall review the clinical evidence presented by the manufacturer in the clinical evaluation report and the related clinical evaluation that was conducted. The notified body shall employ device reviewers with sufficient clinical expertise and, if necessary, use external clinical experts with direct and current experience relating to the device in question or the clinical condition in which it is utilised, for the purposes of that review.4.5.The notified body shall, in circumstances in which the clinical evidence is based partly or totally on data from devices which are claimed to be equivalent to the device under assessment, assess the suitability of using such data, taking into account factors such as new indications and innovation. The notified body shall clearly document its conclusions on the claimed equivalence, and on the relevance and adequacy of the data for demonstrating conformity. For any characteristic of the device claimed as innovative by the manufacturer or for new indications, the notified body shall assess to what extent specific claims are supported by specific pre-clinical and clinical data and risk analysis.4.6.The notified body shall verify that the clinical evidence and the clinical evaluation are adequate and shall verify the conclusions drawn by the manufacturer on the conformity with the relevant general safety and performance requirements. That verification shall include consideration of the adequacy of the benefit-risk determination, the risk management, the instructions for use, the user training and the manufacturer's post-market surveillance plan, and include a review of the need for, and the adequacy of, the PMCF plan proposed, where applicable.4.7.Based on its assessment of the clinical evidence, the notified body shall consider the clinical evaluation and the benefit-risk determination, and whether specific milestones need to be defined to allow the notified body to review updates to the clinical evidence that result from post-market surveillance and PMCF data.4.8.The notified body shall clearly document the outcome of its assessment in the clinical evaluation assessment report.4.9.The notified body shall provide the manufacturer with a report on the technical documentation assessment, including a clinical evaluation assessment report. If the device conforms to the relevant provisions of this Regulation, the notified body shall issue an EU technical documentation assessment certificate. The certificate shall contain the conclusions of the technical documentation assessment, the conditions of the certificate's validity, the data needed for identification of the approved design, and, where appropriate, a description of the intended purpose of the device.4.10.Changes to the approved device shall require approval from the notified body which issued the EU technical documentation assessment certificate where such changes could affect the safety and performance of the device or the conditions prescribed for use of the device. Where the manufacturer plans to introduce any of the above-mentioned changes it shall inform the notified body which issued the EU technical documentation assessment certificate thereof. The notified body shall assess the planned changes and decide whether the planned changes require a new conformity assessment in accordance with Article 52 or whether they could be addressed by means of a supplement to the EU technical documentation assessment certificate. In the latter case, the notified body shall assess the changes, notify the manufacturer of its decision and, where the changes are approved, provide it with a supplement to the EU technical documentation assessment certificate.5.Specific additional procedures5.1.Assessment procedure for certain class III and class IIb devices(a)For class III implantable devices, and for class IIb active devices intended to administer and/or remove a medicinal product as referred to in Section 6.4. of Annex VIII (Rule 12), the notified body shall, having verified the quality of clinical data supporting the clinical evaluation report of the manufacturer referred to in Article 61(12), prepare a clinical evaluation assessment report which sets out its conclusions concerning the clinical evidence provided by the manufacturer, in particular concerning the benefit-risk determination, the consistency of that evidence with the intended purpose, including the medical indication or indications and the PMCF plan referred to in Article 10(3) and Part B of Annex XIV.The notified body shall transmit its clinical evaluation assessment report, along with the manufacturer's clinical evaluation documentation, referred to in points (c) and (d) of Section 6.1 of Annex II, to the Commission.The Commission shall immediately transmit those documents to the relevant expert panel referred to in Article 106.(b)The notified body may be requested to present its conclusions as referred to in point (a) to the expert panel concerned.(c)The expert panel shall decide, under the supervision of the Commission, on the basis of all of the following criteria:(i)the novelty of the device or of the related clinical procedure involved, and the possible major clinical or health impact thereof;(ii)a significantly adverse change in the benefit-risk profile of a specific category or group of devices due to scientifically valid health concerns in respect of components or source material or in respect of the impact on health in the case of failure of the device;(iii)a significantly increased rate of serious incidents reported in accordance with Article 87 in respect of a specific category or group of devices,whether to provide a scientific opinion on the clinical evaluation assessment report of the notified body based on the clinical evidence provided by the manufacturer, in particular concerning the benefit-risk determination, the consistency of that evidence with the medical indication or indications and the PMCF plan. That scientific opinion shall be provided within a period of 60 days, starting on the day of receipt of the documents from the Commission as referred to in point (a). The reasons for the decision to provide a scientific opinion on the basis of the criteria in points (i), (ii) and (iii) shall be included in the scientific opinion. Where the information submitted is not sufficient for the expert panel to reach a conclusion, this shall be stated in the scientific opinion.(d)The expert panel may decide, under the supervision of the Commission, on the basis of the criteria laid down in point (c) not to provide a scientific opinion, in which case it shall inform the notified body as soon as possible and in any event within 21 days of receipt of the documents as referred to in point (a) from the Commission. The expert panel shall within that time limit provide the notified body and the Commission with the reasons for its decision, whereupon the notified body may proceed with the certification procedure of that device.(e)The expert panel shall within 21 days of receipt of the documents from the Commission notify the Commission, through Eudamed whether it intends to provide a scientific opinion, pursuant to point (c), or whether it intends not to provide a scientific opinion, pursuant to point (d).(f)Where no opinion has been delivered within a period of 60 days, the notified body may proceed with the certification procedure of the device in question.(g)The notified body shall give due consideration to the views expressed in the scientific opinion of the expert panel. Where the expert panel finds that the level of clinical evidence is not sufficient or otherwise gives rise to serious concerns about the benefit-risk determination, the consistency of that evidence with the intended purpose, including the medical indication(s), and with the PMCF plan, the notified body shall, if necessary, advise the manufacturer to restrict the intended purpose of the device to certain groups of patients or certain medical indications and/or to impose a limit on the duration of validity of the certificate, to undertake specific PMCF studies, to adapt the instructions for use or the summary of safety and performance, or to impose other restrictions in its conformity assessment report, as appropriate. The notified body shall provide a full justification where it has not followed the advice of the expert panel in its conformity assessment report and the Commission shall without prejudice to Article 109 make both the scientific opinion of the expert panel and the written justification provided by the notified body publicly available via Eudamed.(h)The Commission, after consultation with the Member States and relevant scientific experts shall provide guidance for expert panels for consistent interpretation of the criteria in point (c) before 26 May 2020.5.2.Procedure in the case of devices incorporating a medicinal substance(a)Where a device incorporates, as an integral part, a substance which, if used separately, may be considered to be a medicinal product within the meaning of point 2 of Article 1 of Directive 2001/83/EC, including a medicinal product derived from human blood or human plasma and that has an action ancillary to that of the device, the quality, safety and usefulness of the substance shall be verified by analogy with the methods specified in Annex I to Directive 2001/83/EC.(b)Before issuing an EU technical documentation assessment certificate, the notified body shall, having verified the usefulness of the substance as part of the device and taking account of the intended purpose of the device, seek a scientific opinion from one of the competent authorities designated by the Member States in accordance with Directive 2001/83/EC or from the EMA, either of which to be referred to in this Section as the medicinal products authority consulted depending on which has been consulted under this point, on the quality and safety of the substance including the benefit or risk of the incorporation of the substance into the device. Where the device incorporates a human blood or plasma derivative or a substance that, if used separately, may be considered to be a medicinal product falling exclusively within the scope of the Annex to Regulation (EC) No 726/2004, the notified body shall seek the opinion of the EMA.(c)When issuing its opinion, the medicinal products authority consulted shall take into account the manufacturing process and the data relating to the usefulness of incorporation of the substance into the device as determined by the notified body.(d)The medicinal products authority consulted shall provide its opinion to the notified body within 210 days of receipt of all the necessary documentation.(e)The scientific opinion of the medicinal products authority consulted, and any possible update of that opinion, shall be included in the documentation of the notified body concerning the device. The notified body shall give due consideration to the views expressed in the scientific opinion when making its decision. The notified body shall not deliver the certificate if the scientific opinion is unfavourable and shall convey its final decision to the medicinal products authority consulted.(f)Before any change is made with respect to an ancillary substance incorporated in a device, in particular related to its manufacturing process, the manufacturer shall inform the notified body of the changes. That notified body shall seek the opinion of the medicinal products authority consulted, in order to confirm that the quality and safety of the ancillary substance remain unchanged. The medicinal products authority consulted shall take into account the data relating to the usefulness of incorporation of the substance into the device as determined by the notified body, in order to ensure that the changes have no negative impact on the risk or benefit previously established concerning the incorporation of the substance into the device. The medicinal products authority consulted shall provide its opinion within 60 days after receipt of all the necessary documentation regarding the changes. The notified body shall not deliver the supplement to the EU technical documentation assessment certificate if the scientific opinion provided by the medicinal products authority consulted is unfavourable. The notified body shall convey its final decision to the medicinal products authority consulted.(g)Where the medicinal products authority consulted obtains information on the ancillary substance, which could have an impact on the risk or benefit previously established concerning the incorporation of the substance into the device, it shall advise the notified body as to whether this information has an impact on the risk or benefit previously established concerning the incorporation of the substance into the device. The notified body shall take that advice into account in reconsidering its assessment of the conformity assessment procedure.5.3.Procedure in the case of devices manufactured utilising, or incorporating, tissues or cells of human or animal origin, or their derivatives, that are non-viable or rendered non-viable5.3.1.Tissues or cells of human origin or their derivatives(a)For devices manufactured utilising derivatives of tissues or cells of human origin that are covered by this Regulation in accordance with point (g) of Article 1(6) and for devices that incorporate, as an integral part, tissues or cells of human origin, or their derivatives, covered by Directive 2004/23/EC, that have an action ancillary to that of the device, the notified body shall, prior to issuing an EU technical documentation assessment certificate, seek a scientific opinion from one of the competent authorities designated by the Member States in accordance with Directive 2004/23/EC (human tissues and cells competent authority) on the aspects relating to the donation, procurement and testing of tissues or cells of human origin or their derivatives. The notified body shall submit a summary of the preliminary conformity assessment which provides, among other things, information about the non-viability of the human tissues or cells in question, their donation, procurement and testing and the risk or benefit of the incorporation of the tissues or cells of human origin or their derivatives into the device.(b)Within 120 days of receipt of all the necessary documentation, the human tissues and cells competent authority shall provide to the notified body its opinion.(c)The scientific opinion of the human tissues and cells competent authority, and any possible update, shall be included in the documentation of the notified body concerning the device. The notified body shall give due consideration to the views expressed in the scientific opinion of the human tissues and cells competent authority when making its decision. The notified body shall not deliver the certificate if that scientific opinion is unfavourable. It shall convey its final decision to the human tissues and cells competent authority concerned.(d)Before any change is made with respect to non-viable tissues or cells of human origin or their derivatives incorporated in a device, in particular relating to their donation, testing or procurement, the manufacturer shall inform the notified body of the intended changes. The notified body shall consult the authority that was involved in the initial consultation, in order to confirm that the quality and safety of the tissues or cells of human origin or their derivatives incorporated in the device are maintained. The human tissues and cells competent authority concerned shall take into account the data relating to the usefulness of incorporation of the tissues or cells of human origin or their derivatives into the device as determined by the notified body, in order to ensure that the changes have no negative impact on the established benefit-risk ratio of the addition of the tissues or cells of human origin or their derivatives in the device. It shall provide its opinion within 60 days of receipt of all the necessary documentation regarding the intended changes. The notified body shall not deliver a supplement to the EU technical documentation assessment certificate if the scientific opinion is unfavourable and shall convey its final decision to the human tissues and cells competent authority concerned.5.3.2.Tissues or cells of animal origin or their derivativesIn the case of devices manufactured utilising animal tissue which is rendered non-viable or utilising non-viable products derived from animal tissue, as referred to in Regulation (EU) No 722/2012, the notified body shall apply the relevant requirements laid down in that Regulation.5.4.Procedure in the case of devices that are composed of substances or of combinations of substances that are absorbed by or locally dispersed in the human body(a)The quality and safety of devices that are composed of substances or of combinations of substances that are intended to be introduced into the human body via a body orifice or applied to the skin and that are absorbed by, or locally dispersed in, the human body, shall be verified where applicable and only in respect of the requirements not covered by this Regulation, in accordance with the relevant requirements laid down in Annex I to Directive 2001/83/EC for the evaluation of absorption, distribution, metabolism, excretion, local tolerance, toxicity, interaction with other devices, medicinal products or other substances and potential for adverse reactions.(b)In addition, for devices, or their products of metabolism, that are systemically absorbed by the human body in order to achieve their intended purpose, the notified body shall seek a scientific opinion from one of the competent authorities designated by the Member States in accordance with Directive 2001/83/EC or from the EMA, either of which to be referred to in this Section as the medicinal products authority consulted depending on which has been consulted under this point, on the compliance of the device with the relevant requirements laid down in Annex I to Directive 2001/83/EC.(c)The opinion of the medicinal products authority consulted shall be drawn up within 150 days of receipt of all the necessary documentation.(d)The scientific opinion of the medicinal products authority consulted, and any possible update, shall be included in the documentation of the notified body concerning the device. The notified body shall give due consideration to the views expressed in the scientific opinion when making its decision and shall convey its final decision to the medicinal products authority consulted.6.Batch verification in the case of devices incorporating, as an integral part, a medicinal substance which, if used separately, would be considered to be a medicinal product derived from human blood or human plasma as referred to in Article 1(8)Upon completing the manufacture of each batch of devices that incorporate, as an integral part, a medicinal substance which, if used separately, would be considered to be a medicinal product derived from human blood or human plasma as referred to in the first subparagraph of Article 1(8), the manufacturer shall inform the notified body of the release of the batch of devices and send it the official certificate concerning the release of the batch of human blood or plasma derivative used in the device, issued by a Member State laboratory or a laboratory designated for that purpose by a Member State in accordance with Article 114(2) of Directive 2001/83/EC.CHAPTER IIIADMINISTRATIVE PROVISIONS7.The manufacturer or, where the manufacturer does not have a registered place of business in a Member State, its authorised representative shall, for a period ending no sooner than 10 years, and in the case of implantable devices no sooner than 15 years, after the last device has been placed on the market, keep at the disposal of the competent authorities:the EU declaration of conformity,the documentation referred to in the fifth indent of Section 2.1 and in particular the data and records arising from the procedures referred to in point (c) of the second paragraph of Section 2.2,information on the changes referred to in Section 2.4,the documentation referred to in Section 4.2, andthe decisions and reports from the notified body as referred to in this Annex.8.Each Member State shall require that the documentation referred to in Section 7 is kept at the disposal of competent authorities for the period indicated in that Section in case a manufacturer, or its authorised representative, established within its territory goes bankrupt or ceases its business activity prior to the end of that period.

MODIFIED Annex XV — ANNEX XV · applies from unchanged

Disputed — seen by the structural diff, not by corpus metadata.

In Section 2.5 of Chapter II, the phrase describing undesirable effects was changed to refer to undesirable side-effects. Annex XV, v1 Annex XV, v2

text before / after

32017R074502017R0745-20170505

ANNEX XVCLINICAL INVESTIGATIONSCHAPTER IGENERAL REQUIREMENTS1.Ethical principlesEach step in the clinical investigation, from the initial consideration of the need for and justification of the study to the publication of the results, shall be carried out in accordance with recognised ethical principles.2.Methods2.1.Clinical investigations shall be performed on the basis of an appropriate plan of investigation reflecting the latest scientific and technical knowledge and defined in such a way as to confirm or refute the manufacturer's claims regarding the safety, performance and aspects relating to benefit-risk of devices as referred to in Article 62(1); the clinical investigations shall include an adequate number of observations to guarantee the scientific validity of the conclusions. The rationale for the design and chosen statistical methodology shall be presented as further described in Section 3.6 of Chapter II of this Annex.2.2.The procedures used to perform the clinical investigation shall be appropriate to the device under investigation.2.3.The research methodologies used to perform the clinical investigation shall be appropriate to the device under investigation.2.4.Clinical investigations shall be performed in accordance with the clinical investigation plan by a sufficient number of intended users and in a clinical environment that is representative of the intended normal conditions of use of the device in the target patient population. Clinical investigations shall be in line with the clinical evaluation plan as referred to in Part A of Annex XIV.2.5.All the appropriate technical and functional features of the device, in particular those involving safety and performance, and their expected clinical outcomes shall be appropriately addressed in the investigational design. A list of the technical and functional features of the device and the related expected clinical outcomes shall be provided.2.6.The endpoints of the clinical investigation shall address the intended purpose, clinical benefits, performance and safety of the device. The endpoints shall be determined and assessed using scientifically valid methodologies. The primary endpoint shall be appropriate to the device and clinically relevant.2.7.Investigators shall have access to the technical and clinical data regarding the device. Personnel involved in the conduct of an investigation shall be adequately instructed and trained in the proper use of the investigational device, and as regards the clinical investigation plan and good clinical practice. This training shall be verified and where necessary arranged by the sponsor and documented appropriately.2.8.The clinical investigation report, signed by the investigator, shall contain a critical evaluation of all the data collected during the clinical investigation, and shall include any negative findings.CHAPTER IIDOCUMENTATION REGARDING THE APPLICATION FOR CLINICAL INVESTIGATIONFor investigational devices covered by Article 62, the sponsor shall draw up and submit the application in accordance with Article 70 accompanied by the following documents:1.Application formThe application form shall be duly filled in, containing information regarding:1.1.name, address and contact details of the sponsor and, if applicable, name, address and contact details of its contact person or legal representative in accordance with Article 62(2) established in the Union;1.2.if different from those in Section 1.1, name, address and contact details of the manufacturer of the device intended for clinical investigation and, if applicable, of its authorised representative;1.3.title of the clinical investigation;1.4.status of the clinical investigation application (i.e. first submission, resubmission, significant amendment);1.5.details and/or reference to the clinical evaluation plan;1.6.If the application is a resubmission with regard to a device for which an application has been already submitted, the date or dates and reference number or numbers of the earlier application or in the case of significant amendment, reference to the original application. The sponsor shall identify all of the changes from the previous application together with a rationale for those changes, in particular, whether any changes have been made to address conclusions of previous competent authority or ethics committee reviews;1.7.if the application is submitted in parallel with an application for a clinical trial in accordance with Regulation (EU) No 536/2014, reference to the official registration number of the clinical trial;1.8.identification of the Member States and third countries in which the clinical investigation is to be conducted as part of a multicentre or multinational study at the time of application;1.9.a brief description of the investigational device, its classification and other information necessary for the identification of the device and device type;1.10.information as to whether the device incorporates a medicinal substance, including a human blood or plasma derivative or whether it is manufactured utilising non-viable tissues or cells of human or animal origin, or their derivatives;1.11.summary of the clinical investigation plan including the objective or objectives of the clinical investigation, the number and gender of subjects, criteria for subject selection, whether there are subjects under 18 years of age, design of the investigation such as controlled and/or randomised studies, planned dates of commencement and of completion of the clinical investigation;1.12.if applicable, information regarding a comparator device, its classification and other information necessary for the identification of the comparator device;1.13.evidence from the sponsor that the clinical investigator and the investigational site are capable of conducting the clinical investigation in accordance with the clinical investigation plan;1.14.details of the anticipated start date and duration of the investigation;1.15.details to identify the notified body, if already involved at the stage of application for a clinical investigation;1.16.confirmation that the sponsor is aware that the competent authority may contact the ethics committee that is assessing or has assessed the application; and1.17.the statement referred to in Section 4.1.2.Investigator's BrochureThe investigator's brochure (IB) shall contain the clinical and non-clinical information on the investigational device that is relevant for the investigation and available at the time of application. Any updates to the IB or other relevant information that is newly available shall be brought to the attention of the investigators in a timely manner. The IB shall be clearly identified and contain in particular the following information:2.1.Identification and description of the device, including information on the intended purpose, the risk classification and applicable classification rule pursuant to Annex VIII, design and manufacturing of the device and reference to previous and similar generations of the device.2.2.Manufacturer's instructions for installation, maintenance, maintaining hygiene standards and for use, including storage and handling requirements, as well as, to the extent that such information is available, information to be placed on the label, and instructions for use to be provided with the device when placed on the market. In addition, information relating to any relevant training required.2.3.Pre-clinical evaluation based on relevant pre-clinical testing and experimental data, in particular regarding in-design calculations, in vitro tests, ex vivo tests, animal tests, mechanical or electrical tests, reliability tests, sterilisation validation, software verification and validation, performance tests, evaluation of biocompatibility and biological safety, as applicable.2.4.Existing clinical data, in particular:from relevant scientific literature available relating to the safety, performance, clinical benefits to patients, design characteristics and intended purpose of the device and/or of equivalent or similar devices;other relevant clinical data available relating to the safety, performance, clinical benefits to patients, design characteristics and intended purpose of equivalent or similar devices of the same manufacturer, including length of time on the market and a review of performance, clinical benefit and safety-related issues and any corrective actions taken.2.5.Summary of the benefit-risk analysis and the risk management, including information regarding known or foreseeable risks, any undesirable effects, side-effects, contraindications and warnings.2.6.In the case of devices that incorporate a medicinal substance, including a human blood or plasma derivative or devices manufactured utilising non-viable tissues or cells of human or animal origin, or their derivatives, detailed information on the medicinal substance or on the tissues, cells or their derivatives, and on the compliance with the relevant general safety and performance requirements and the specific risk management in relation to the substance or tissues, cells or their derivatives, as well as evidence for the added value of incorporation of such constituents in relation to the clinical benefit and/or safety of the device.2.7.A list detailing the fulfilment of the relevant general safety and performance requirements set out in Annex I, including the standards and CS applied, in full or in part, as well as a description of the solutions for fulfilling the relevant general safety and performance requirements, in so far as those standards and CS have not or have only been partly fulfilled or are lacking.2.8.A detailed description of the clinical procedures and diagnostic tests used in the course of the clinical investigation and in particular information on any deviation from normal clinical practice.3.Clinical Investigation PlanThe clinical investigation plan (CIP) shall set out the rationale, objectives, design methodology, monitoring, conduct, record-keeping and the method of analysis for the clinical investigation. It shall contain in particular the information as laid down in this Annex. If part of this information is submitted in a separate document, it shall be referenced in the CIP.3.1.General3.1.1.Single identification number of the clinical investigation, as referred to in Article 70(1).3.1.2.Identification of the sponsor — name, address and contact details of the sponsor and, where applicable, the name, address and contact details of the sponsor's contact person or legal representative in accordance with Article 62(2) established in the Union.3.1.3.Information on the principal investigator at each investigational site, the coordinating investigator for the investigation, the address details for each investigational site and the emergency contact details for the principal investigator at each site. The roles, responsibilities and qualifications of the various kinds of investigators shall be specified in the CIP.3.1.4.A brief description of how the clinical investigation is financed and a brief description of the agreement between the sponsor and the site.3.1.5.Overall synopsis of the clinical investigation, in an official Union language determined by the Member State concerned.3.2.Identification and description of the device, including its intended purpose, its manufacturer, its traceability, the target population, materials coming into contact with the human body, the medical or surgical procedures involved in its use and the necessary training and experience for its use, background literature review, the current state of the art in clinical care in the relevant field of application and the proposed benefits of the new device.3.3.Risks and clinical benefits of the device to be examined, with justification of the corresponding expected clinical outcomes in the clinical investigation plan.3.4.Description of the relevance of the clinical investigation in the context of the state of the art of clinical practice.3.5.Objectives and hypotheses of the clinical investigation.3.6.Design of the clinical investigation with evidence of its scientific robustness and validity.3.6.1.General information such as type of investigation with rationale for choosing it, for its endpoints and for its variables as set out in the clinical evaluation plan.3.6.2.Information on the investigational device, on any comparator and on any other device or medication to be used in the clinical investigation.3.6.3.Information on subjects, selection criteria, size of investigation population, representativeness of investigation population in relation to target population and, if applicable, information on vulnerable subjects involved such as children, pregnant women, immuno-compromised or, elderly subjects.3.6.4.Details of measures to be taken to minimise bias, such as randomisation, and management of potential confounding factors.3.6.5.Description of the clinical procedures and diagnostic methods relating to the clinical investigation and in particular highlighting any deviation from normal clinical practice.3.6.6.Monitoring plan.3.7.Statistical considerations, with justification, including a power calculation for the sample size, if applicable.3.8.Data management.3.9.Information about any amendments to the CIP.3.10.Policy regarding follow-up and management of any deviations from the CIP at the investigational site and clear prohibition of use of waivers from the CIP.3.11.Accountability regarding the device, in particular control of access to the device, follow-up in relation to the device used in the clinical investigation and the return of unused, expired or malfunctioning devices.3.12.Statement of compliance with the recognised ethical principles for medical research involving humans, and the principles of good clinical practice in the field of clinical investigations of devices, as well as with the applicable regulatory requirements.3.13.Description of the Informed consent process.3.14.Safety reporting, including definitions of adverse events and serious adverse events, device deficiencies, procedures and timelines for reporting.3.15.Criteria and procedures for follow-up of subjects following the end, temporary halt or early termination of an investigation, for follow-up of subjects who have withdrawn their consent and procedures for subjects lost to follow-up. Such procedures shall for implantable devices, cover as a minimum traceability.3.16.A description of the arrangements for taking care of the subjects after their participation in the clinical investigation has ended, where such additional care is necessary because of the subjects' participation in the clinical investigation and where it differs from that normally expected for the medical condition in question.3.17.Policy as regards the establishment of the clinical investigation report and publication of results in accordance with the legal requirements and the ethical principles referred to in Section 1 of Chapter I.3.18.List of the technical and functional features of the device, with specific mention of those covered by the investigation.3.19.Bibliography.4.Other information4.1.A signed statement by the natural or legal person responsible for the manufacture of the investigational device that the device in question conforms to the general safety and performance requirements apart from the aspects covered by the clinical investigation and that, with regard to those aspects, every precaution has been taken to protect the health and safety of the subject.4.2.Where applicable according to national law, copy of the opinion or opinions of the ethics committee or committees concerned. Where according to national law the opinion or opinions of the ethics committee or committees is not required at the time of the submission of the application, a copy of the opinion or opinions shall be submitted as soon as available.4.3.Proof of insurance cover or indemnification of subjects in case of injury, pursuant to Article 69 and the corresponding national law.4.4.Documents to be used to obtain informed consent, including the patient information sheet and the informed consent document.4.5.Description of the arrangements to comply with the applicable rules on the protection and confidentiality of personal data, in particular:organisational and technical arrangements that will be implemented to avoid unauthorised access, disclosure, dissemination, alteration or loss of information and personal data processed;a description of measures that will be implemented to ensure confidentiality of records and personal data of subjects; anda description of measures that will be implemented in case of a data security breach in order to mitigate the possible adverse effects.4.6.Full details of the available technical documentation, for example detailed risk analysis/management documentation or specific test reports, shall, upon request, be submitted to the competent authority reviewing an application.CHAPTER IIIOTHER OBLIGATIONS OF THE SPONSOR1.The sponsor shall undertake to keep available for the competent national authorities any documentation necessary to provide evidence for the documentation referred to in Chapter II of this Annex. If the sponsor is not the natural or legal person responsible for the manufacture of the investigational device, that obligation may be fulfilled by that person on behalf of the sponsor.2.The Sponsor shall have an agreement in place to ensure that any serious adverse events or any other event as referred to in Article 80(2) are reported by the investigator or investigators to the sponsor in a timely manner.3.The documentation mentioned in this Annex shall be kept for a period of at least 10 years after the clinical investigation with the device in question has ended, or, in the event that the device is subsequently placed on the market, at least 10 years after the last device has been placed on the market. In the case of implantable devices, the period shall be at least 15 years.Each Member State shall require that this documentation is kept at the disposal of the competent authorities for the period referred to in the first subparagraph in case the sponsor, or its contact person or legal representative as referred to in Article 62(2) established within its territory, goes bankrupt or ceases its activity prior to the end of this period.4.The Sponsor shall appoint a monitor that is independent from the investigational site to ensure that the investigation is conducted in accordance with the CIP, the principles of good clinical practice and this Regulation.5.The Sponsor shall complete the follow-up of investigation subjects.6.The Sponsor shall provide evidence that the investigation is being conducted in line with good clinical practice, for instance through internal or external inspection.7.The Sponsor shall prepare a clinical investigation report which includes at least the following:Cover/introductory page or pages indicating the title of the investigation, the investigational device, the single identification number, the CIP number and the details with signatures of the coordinating investigators and the principal investigators from each investigational site.Details of the author and date of the report.A summary of the investigation covering the title, purpose of the investigation, description of the investigation, investigational design and methods used, the results of the investigation and conclusion of the investigation. The completion date of the investigation, and in particular details of early termination, temporary halts or suspensions of investigations.Investigational device description, in particular clearly defined intended purpose.A summary of the clinical investigation plan covering objectives, design, ethical aspects, monitoring and quality measures, selection criteria, target patient populations, sample size, treatment schedules, follow-up duration, concomitant treatments, statistical plan, including hypothesis, sample size calculation and analysis methods, as well as a justification.Results of the clinical investigation covering, with rationale and justification, subject demographics, analysis of results related to chosen endpoints, details of subgroup analysis, as well as compliance with the CIP, and covering follow-up of missing data and of patients withdrawing from the clinical investigation, or lost to follow-up.Summary of serious adverse events, adverse device effects, device deficiencies and any relevant corrective actions.Discussion and overall conclusions covering safety and performance results, assessment of risks and clinical benefits, discussion of clinical relevance in accordance with clinical state of the art, any specific precautions for specific patient populations, implications for the investigational device, limitations of the investigation.

The full entry, with the citation mapping v1 = 32017R0745, v2 = 02017R0745-20170505, is committed at eu/32017R0745/CHANGELOG.md.

02017R0745-2025011002017R0745-20260101

in force 2026-01-01 · detected 2026-08-10

5 provisions touched — 5 substantive, 0 date-only, 5 disputed · 0 sentences quoted verbatim by the gate, 4 changes shipped without an explanation

MODIFIED SCT 10.4.1 · applies from unknown

Disputed — seen by the instruction parse, not by the structural diff, corpus metadata.

No explanation shipped — the structural diff did not see this change, so it carries no text; another signal named the unit and the disagreement ships as `disputed`.

text before / after

No text on either side: this unit was named by a signal that carries no text, and only the structural diff carries any.

MODIFIED SCT 10.4.2 · applies from unknown

Disputed — seen by the instruction parse, not by the structural diff, corpus metadata.

No explanation shipped — the structural diff did not see this change, so it carries no text; another signal named the unit and the disagreement ships as `disputed`.

text before / after

No text on either side: this unit was named by a signal that carries no text, and only the structural diff carries any.

MODIFIED SCT 10.4.3 · applies from unknown

Disputed — seen by the instruction parse, not by the structural diff, corpus metadata.

No explanation shipped — the structural diff did not see this change, so it carries no text; another signal named the unit and the disagreement ships as `disputed`.

text before / after

No text on either side: this unit was named by a signal that carries no text, and only the structural diff carries any.

MODIFIED SCT 10.4.4 · applies from unknown

Disputed — seen by the instruction parse, not by the structural diff, corpus metadata.

No explanation shipped — the structural diff did not see this change, so it carries no text; another signal named the unit and the disagreement ships as `disputed`.

text before / after

No text on either side: this unit was named by a signal that carries no text, and only the structural diff carries any.

MODIFIED Annex I — ANNEX I · applies from unknown (the text changed beyond its dates; the applicability binding is prose)

Disputed — seen by the structural diff, corpus metadata, not by the instruction parse.

Section 10.4.1(b) no longer defines endocrine-disrupting substances solely by reference to identification under Article 59 of Regulation (EC) No 1907/2006 or a future Article 5(3) delegated act under Regulation (EU) No 528/2012; it now also includes substances classified as Category 1 endocrine disruptors for human health under Part 3 of Annex VI to Regulation (EC) No 1272/2008, alongside substances identified under the REACH Article 59 procedure or under Regulation (EU) No 528/2012 directly, without reference to a delegated act. Annex I, v2 Annex I, v1

Section 10.4.2(d) now refers to "the latest relevant guidelines" rather than "the latest relevant scientific committee guidelines," and Sections 10.4.3 and 10.4.4 replace the prior mandate to a scientific committee, with its 26 May 2018 and 26 May 2020 deadlines, with a mechanism whereby the Commission requests the European Chemicals Agency to prepare and update phthalate guidelines when appropriate but at least every five years, and to prepare guidelines on other CMR and endocrine-disrupting substances following the same process. Annex I, v1 Annex I, v2

The revised Section 10.4.3 also adds that the European Chemicals Agency shall consult the Committee for Risk Assessment and the Committee for Socio-economic Analysis when appropriate or when requested by the Commission, a step absent from the earlier text. Annex I, v2

text before / after

02017R0745-2025011002017R0745-20260101

ANNEX IGENERAL SAFETY AND PERFORMANCE REQUIREMENTSCHAPTER IGENERAL REQUIREMENTS1.Devices shall achieve the performance intended by their manufacturer and shall be designed and manufactured in such a way that, during normal conditions of use, they are suitable for their intended purpose. They shall be safe and effective and shall not compromise the clinical condition or the safety of patients, or the safety and health of users or, where applicable, other persons, provided that any risks which may be associated with their use constitute acceptable risks when weighed against the benefits to the patient and are compatible with a high level of protection of health and safety, taking into account the generally acknowledged state of the art.2.The requirement in this Annex to reduce risks as far as possible means the reduction of risks as far as possible without adversely affecting the benefit-risk ratio.3.Manufacturers shall establish, implement, document and maintain a risk management system.Risk management shall be understood as a continuous iterative process throughout the entire lifecycle of a device, requiring regular systematic updating. In carrying out risk management manufacturers shall:(a)establish and document a risk management plan for each device;(b)identify and analyse the known and foreseeable hazards associated with each device;(c)estimate and evaluate the risks associated with, and occurring during, the intended use and during reasonably foreseeable misuse;(d)eliminate or control the risks referred to in point (c) in accordance with the requirements of Section 4;(e)evaluate the impact of information from the production phase and, in particular, from the post-market surveillance system, on hazards and the frequency of occurrence thereof, on estimates of their associated risks, as well as on the overall risk, benefit-risk ratio and risk acceptability; and(f)based on the evaluation of the impact of the information referred to in point (e), if necessary amend control measures in line with the requirements of Section 4.4.Risk control measures adopted by manufacturers for the design and manufacture of the devices shall conform to safety principles, taking account of the generally acknowledged state of the art. To reduce risks, Manufacturers shall manage risks so that the residual risk associated with each hazard as well as the overall residual risk is judged acceptable. In selecting the most appropriate solutions, manufacturers shall, in the following order of priority:(a)eliminate or reduce risks as far as possible through safe design and manufacture;(b)where appropriate, take adequate protection measures, including alarms if necessary, in relation to risks that cannot be eliminated; and(c)provide information for safety (warnings/precautions/contra-indications) and, where appropriate, training to users.Manufacturers shall inform users of any residual risks.5.In eliminating or reducing risks related to use error, the manufacturer shall:(a)reduce as far as possible the risks related to the ergonomic features of the device and the environment in which the device is intended to be used (design for patient safety), and(b)give consideration to the technical knowledge, experience, education, training and use environment, where applicable, and the medical and physical conditions of intended users (design for lay, professional, disabled or other users).6.The characteristics and performance of a device shall not be adversely affected to such a degree that the health or safety of the patient or the user and, where applicable, of other persons are compromised during the lifetime of the device, as indicated by the manufacturer, when the device is subjected to the stresses which can occur during normal conditions of use and has been properly maintained in accordance with the manufacturer's instructions.7.Devices shall be designed, manufactured and packaged in such a way that their characteristics and performance during their intended use are not adversely affected during transport and storage, for example, through fluctuations of temperature and humidity, taking account of the instructions and information provided by the manufacturer.8.All known and foreseeable risks, and any undesirable side-effects, shall be minimised and be acceptable when weighed against the evaluated benefits to the patient and/or user arising from the achieved performance of the device during normal conditions of use.9.For the devices referred to in Annex XVI, the general safety requirements set out in Sections 1 and 8 shall be understood to mean that the device, when used under the conditions and for the purposes intended, does not present a risk at all or presents a risk that is no more than the maximum acceptable risk related to the product's use which is consistent with a high level of protection for the safety and health of persons.CHAPTER IIREQUIREMENTS REGARDING DESIGN AND MANUFACTURE10.Chemical, physical and biological properties10.1.Devices shall be designed and manufactured in such a way as to ensure that the characteristics and performance requirements referred to in Chapter I are fulfilled. Particular attention shall be paid to:(a)the choice of materials and substances used, particularly as regards toxicity and, where relevant, flammability;(b)the compatibility between the materials and substances used and biological tissues, cells and body fluids, taking account of the intended purpose of the device and, where relevant, absorption, distribution, metabolism and excretion;(c)the compatibility between the different parts of a device which consists of more than one implantable part;(d)the impact of processes on material properties;(e)where appropriate, the results of biophysical or modelling research the validity of which has been demonstrated beforehand;(f)the mechanical properties of the materials used, reflecting, where appropriate, matters such as strength, ductility, fracture resistance, wear resistance and fatigue resistance;(g)surface properties; and(h)the confirmation that the device meets any defined chemical and/or physical specifications.10.2.Devices shall be designed, manufactured and packaged in such a way as to minimise the risk posed by contaminants and residues to patients, taking account of the intended purpose of the device, and to the persons involved in the transport, storage and use of the devices. Particular attention shall be paid to tissues exposed to those contaminants and residues and to the duration and frequency of exposure.10.3.Devices shall be designed and manufactured in such a way that they can be used safely with the materials and substances, including gases, with which they enter into contact during their intended use; if the devices are intended to administer medicinal products they shall be designed and manufactured in such a way as to be compatible with the medicinal products concerned in accordance with the provisions and restrictions governing those medicinal products and that the performance of both the medicinal products and of the devices is maintained in accordance with their respective indications and intended use.10.4.Substances10.4.1.Design and manufacture of devicesDevices shall be designed and manufactured in such a way as to reduce as far as possible the risks posed by substances or particles, including wear debris, degradation products and processing residues, that may be released from the device.Devices, or those parts thereof or those materials used therein that:are invasive and come into direct contact with the human body,(re)administer medicines, body liquids or other substances, including gases, to/from the body, ortransport or store such medicines, body fluids or substances, including gases, to be (re)administered to the body,shall only contain the following substances in a concentration that is above 0,1 % weight by weight (w/w) where justified pursuant to Section 10.4.2:(a)substances which are carcinogenic, mutagenic or toxic to reproduction (CMR), of category 1A or 1B, in accordance with Part 3 of Annex VI to Regulation (EC) No 1272/2008 of the European Parliament and of the CouncilRegulation (EC) No 1272/2008 of the European Parliament and of the Council of 16 December 2008 on classification, labelling and packaging of substances and mixtures, amending and repealing Directives 67/548/EEC and 1999/45/EC, and amending Regulation (EC) No 1907/2006 ( OJ L 353, 31.12.2008, p. 1)., or(b)substances which are classified as endocrine disruptors for human health, of Category 1, in accordance with Part 3 of Annex VI to Regulation (EC) No 1272/2008 of the European Parliament and of the CouncilRegulation (EC) No 1272/2008 of the European Parliament and of the Council of 16 December 2008 on classification, labelling and packaging of substances and mixtures, amending and repealing Directives 67/548/EEC and 1999/45/EC, and amending Regulation (EC) No 1907/2006 (OJ L 353, 31.12.2008, p. 1, ELI: http://data.europa.eu/eli/reg/2008/1272/oj). and substances having endocrine-disrupting properties for which there is scientific evidence of probable serious effects to human health and which are identified either in accordance with the procedure set out in Article 59 of Regulation (EC) No 1907/2006 of the European Parliament and of the CouncilRegulation (EC) No 1907/2006 of the European Parliament and of the Council of 18 December 2006 concerning the Registration, Evaluation, Authorisation and Restriction of Chemicals (REACH) (REACH), establishing a European Chemicals Agency, amending Directive 1999/45/EC and repealing Council Regulation (EEC) No 793/93 and Commission Regulation (EC) No 1488/94 as well as Council Directive 76/769/EEC and Commission Directives 91/155/EEC, 93/67/EEC, 93/105/EC and 2000/21/EC (OJ L 396, 30.12.2006, p. 1). or, once a delegated act has been adopted by the Commission pursuant 1, ELI: http://data.europa.eu/eli/reg/2006/1907/oj). or substances having endocrine-disrupting properties relevant to the first subparagraph of Article 5(3) of human health identified in accordance with Regulation (EU) No 528/2012 of the European Parliament and the CouncilRegulation (EU) No 528/2012 of the European Parliament and the Council of 22 May 2012 concerning the making available on the market of and use of biocidal products (OJ L 167, 27.6.2012, p. 1)., in accordance with the criteria that are relevant to human health amongst the criteria established therein.10.4.2.Justification 1, ELI: http://data.europa.eu/eli/reg/2012/528/oj)..10.4.2.Justification regarding the presence of CMR and/or endocrine-disrupting substancesThe justification for the presence of such substances shall be based upon:(a)an analysis and estimation of potential patient or user exposure to the substance;(b)an analysis of possible alternative substances, materials or designs, including, where available, information about independent research, peer-reviewed studies, scientific opinions from relevant scientific committees and an analysis of the availability of such alternatives;(c)argumentation as to why possible substance and/ or material substitutes, if available, or design changes, if feasible, are inappropriate in relation to maintaining the functionality, performance and the benefit-risk ratios of the product; including taking into account if the intended use of such devices includes treatment of children or treatment of pregnant or breastfeeding women or treatment of other patient groups considered particularly vulnerable to such substances and/or materials; and(d)where applicable and available, the latest relevant scientific committee guidelines in accordance with Sections 10.4.3. 10.4.3 and 10.4.4.10.4.3.Guidelines on phthalatesFor phthalatesWhen deemed appropriate based on the purposes of Section 10.4., latest scientific evidence, but at least every 5 years, the Commission shall, as soon as possible and by 26 May 2018, provide shall request the relevant scientific committee with a mandate European Chemicals Agency (ECHA) to prepare and update guidelines that shall be ready before 26 May 2020. The mandate for on the committee shall encompass at least a benefit-risk assessment of the presence of phthalates which belong to either of the groups of substances referred to in Section 10.4.1, points (a) and (b) of Section 10.4.1. (b). The benefit-risk assessment shall take into account consider the intended purpose and context of the use of the device, as well as any available alternative substances and alternative materials, designs or medical treatments. When deemed treatments.When appropriate on or when requested by the basis of Commission, the latest scientific evidence, but at least every five years, ECHA shall consult the guidelines shall be updated.10.4.4.Guidelines Committee for Risk Assessment and the Committee for Socio-economic Analysis.10.4.4.Guidelines on other CMR and endocrine-disrupting substancesSubsequently, substancesIn addition to the Commission shall mandate the relevant scientific committee to prepare guidelines as referred to in Section 10.4.3. also 10.4.3, the Commission shall request the ECHA to prepare such guidelines for other substances referred to in Section 10.4.1, points (a) and (b) of (b), where appropriate. Such guidelines shall be prepared in accordance with the process described in Section 10.4.1., where appropriate.10.4.5.LabellingWhere 10.4.3.10.4.5.LabellingWhere devices, parts thereof or materials used therein as referred to in Section 10.4.1. contain substances referred to in points (a) or (b) of Section 10.4.1. in a concentration above 0,1 % weight by weight (w/w), the presence of those substances shall be labelled on the device itself and/or on the packaging for each unit or, where appropriate, on the sales packaging, with the list of such substances. If the intended use of such devices includes treatment of children or treatment of pregnant or breastfeeding women or treatment of other patient groups considered particularly vulnerable to such substances and/or materials, information on residual risks for those patient groups and, if applicable, on appropriate precautionary measures shall be given in the instructions for use.10.5.Devices shall be designed and manufactured in such a way as to reduce as far as possible the risks posed by the unintentional ingress of substances into the device taking into account the device and the nature of the environment in which it is intended to be used.10.6.Devices shall be designed and manufactured in such a way as to reduce as far as possible the risks linked to the size and the properties of particles which are or can be released into the patient's or user's body, unless they come into contact with intact skin only. Special attention shall be given to nanomaterials.11.Infection and microbial contamination11.1.Devices and their manufacturing processes shall be designed in such a way as to eliminate or to reduce as far as possible the risk of infection to patients, users and, where applicable, other persons. The design shall:(a)reduce as far as possible and appropriate the risks from unintended cuts and pricks, such as needle stick injuries,(b)allow easy and safe handling,(c)reduce as far as possible any microbial leakage from the device and/or microbial exposure during use, and(d)prevent microbial contamination of the device or its content such as specimens or fluids.11.2.Where necessary devices shall be designed to facilitate their safe cleaning, disinfection, and/or re-sterilisation.11.3.Devices labelled as having a specific microbial state shall be designed, manufactured and packaged to ensure that they remain in that state when placed on the market and remain so under the transport and storage conditions specified by the manufacturer.11.4.Devices delivered in a sterile state shall be designed, manufactured and packaged in accordance with appropriate procedures, to ensure that they are sterile when placed on the market and that, unless the packaging which is intended to maintain their sterile condition is damaged, they remain sterile, under the transport and storage conditions specified by the manufacturer, until that packaging is opened at the point of use. It shall be ensured that the integrity of that packaging is clearly evident to the final user.11.5.Devices labelled as sterile shall be processed, manufactured, packaged and, sterilised by means of appropriate, validated methods.11.6.Devices intended to be sterilised shall be manufactured and packaged in appropriate and controlled conditions and facilities.11.7.Packaging systems for non-sterile devices shall maintain the integrity and cleanliness of the product and, where the devices are to be sterilised prior to use, minimise the risk of microbial contamination; the packaging system shall be suitable taking account of the method of sterilisation indicated by the manufacturer.11.8.The labelling of the device shall distinguish between identical or similar devices placed on the market in both a sterile and a non-sterile condition additional to the symbol used to indicate that devices are sterile.12.Devices incorporating a substance considered to be a medicinal product and devices that are composed of substances or of combinations of substances that are absorbed by or locally dispersed in the human body.12.1.In the case of devices referred to in the first subparagraph of Article 1(8), the quality, safety and usefulness of the substance which, if used separately, would be considered to be a medicinal product within the meaning of point (2) of Article 1 of Directive 2001/83/EC, shall be verified by analogy with the methods specified in Annex I to Directive 2001/83/EC, as required by the applicable conformity assessment procedure under this Regulation.12.2.Devices that are composed of substances or of combinations of substances that are intended to be introduced into the human body, and that are absorbed by or locally dispersed in the human body shall comply, where applicable and in a manner limited to the aspects not covered by this Regulation, with the relevant requirements laid down in Annex I to Directive 2001/83/EC for the evaluation of absorption, distribution, metabolism, excretion, local tolerance, toxicity, interaction with other devices, medicinal products or other substances and potential for adverse reactions, as required by the applicable conformity assessment procedure under this Regulation.13.Devices incorporating materials of biological origin13.1.For devices manufactured utilising derivatives of tissues or cells of human origin which are non-viable or are rendered non-viable covered by this Regulation in accordance with point (g) of Article 1(6), the following shall apply:(a)donation, procurement and testing of the tissues and cells shall be done in accordance with Directive 2004/23/EC;(b)processing, preservation and any other handling of those tissues and cells or their derivatives shall be carried out so as to provide safety for patients, users and, where applicable, other persons. In particular, safety with regard to viruses and other transmissible agents shall be addressed by appropriate methods of sourcing and by implementation of validated methods of elimination or inactivation in the course of the manufacturing process;(c)the traceability system for those devices shall be complementary and compatible with the traceability and data protection requirements laid down in Directive 2004/23/EC and in Directive 2002/98/EC.13.2.For devices manufactured utilising tissues or cells of animal origin, or their derivatives, which are non-viable or rendered non-viable the following shall apply:(a)where feasible taking into account the animal species, tissues and cells of animal origin, or their derivatives, shall originate from animals that have been subjected to veterinary controls that are adapted to the intended use of the tissues. Information on the geographical origin of the animals shall be retained by manufacturers;(b)sourcing, processing, preservation, testing and handling of tissues, cells and substances of animal origin, or their derivatives, shall be carried out so as to provide safety for patients, users and, where applicable, other persons. In particular safety with regard to viruses and other transmissible agents shall be addressed by implementation of validated methods of elimination or viral inactivation in the course of the manufacturing process, except when the use of such methods would lead to unacceptable degradation compromising the clinical benefit of the device;(c)in the case of devices manufactured utilising tissues or cells of animal origin, or their derivatives, as referred to in Regulation (EU) No 722/2012 the particular requirements laid down in that Regulation shall apply.13.3.For devices manufactured utilising non-viable biological substances other than those referred to in Sections 13.1 and 13.2, the processing, preservation, testing and handling of those substances shall be carried out so as to provide safety for patients, users and, where applicable, other persons, including in the waste disposal chain. In particular, safety with regard to viruses and other transmissible agents shall be addressed by appropriate methods of sourcing and by implementation of validated methods of elimination or inactivation in the course of the manufacturing process.14.Construction of devices and interaction with their environment14.1.If the device is intended for use in combination with other devices or equipment the whole combination, including the connection system shall be safe and shall not impair the specified performance of the devices. Any restrictions on use applying to such combinations shall be indicated on the label and/or in the instructions for use. Connections which the user has to handle, such as fluid, gas transfer, electrical or mechanical coupling, shall be designed and constructed in such a way as to minimise all possible risks, such as misconnection.14.2.Devices shall be designed and manufactured in such a way as to remove or reduce as far as possible:(a)the risk of injury, in connection with their physical features, including the volume/pressure ratio, dimensional and where appropriate ergonomic features;(b)risks connected with reasonably foreseeable external influences or environmental conditions, such as magnetic fields, external electrical and electromagnetic effects, electrostatic discharge, radiation associated with diagnostic or therapeutic procedures, pressure, humidity, temperature, variations in pressure and acceleration or radio signal interferences;(c)the risks associated with the use of the device when it comes into contact with materials, liquids, and substances, including gases, to which it is exposed during normal conditions of use;(d)the risks associated with the possible negative interaction between software and the IT environment within which it operates and interacts;(e)the risks of accidental ingress of substances into the device;(f)the risks of reciprocal interference with other devices normally used in the investigations or for the treatment given; and(g)risks arising where maintenance or calibration are not possible (as with implants), from ageing of materials used or loss of accuracy of any measuring or control mechanism.14.3.Devices shall be designed and manufactured in such a way as to minimise the risks of fire or explosion during normal use and in single fault condition. Particular attention shall be paid to devices the intended use of which includes exposure to or use in association with flammable or explosive substances or substances which could cause combustion.14.4.Devices shall be designed and manufactured in such a way that adjustment, calibration, and maintenance can be done safely and effectively.14.5.Devices that are intended to be operated together with other devices or products shall be designed and manufactured in such a way that the interoperability and compatibility are reliable and safe.14.6Any measurement, monitoring or display scale shall be designed and manufactured in line with ergonomic principles, taking account of the intended purpose, users and the environmental conditions in which the devices are intended to be used.14.7.Devices shall be designed and manufactured in such a way as to facilitate their safe disposal and the safe disposal of related waste substances by the user, patient or other person. To that end, manufacturers shall identify and test procedures and measures as a result of which their devices can be safely disposed after use. Such procedures shall be described in the instructions for use.15.Devices with a diagnostic or measuring function15.1.Diagnostic devices and devices with a measuring function, shall be designed and manufactured in such a way as to provide sufficient accuracy, precision and stability for their intended purpose, based on appropriate scientific and technical methods. The limits of accuracy shall be indicated by the manufacturer.15.2.The measurements made by devices with a measuring function shall be expressed in legal units conforming to the provisions of Council Directive 80/181/EECCouncil Directive 80/181/EEC of 20 December 1979 on the approximation of the laws of the Member States relating to units of measurement and on the repeal of Directive 71/354/EEC (OJ L 39, 15.2.1980, p. 40)..16.Protection against radiation16.1.General(a)Devices shall be designed, manufactured and packaged in such a way that exposure of patients, users and other persons to radiation is reduced as far as possible, and in a manner that is compatible with the intended purpose, whilst not restricting the application of appropriate specified levels for therapeutic and diagnostic purposes.(b)The operating instructions for devices emitting hazardous or potentially hazardous radiation shall contain detailed information as to the nature of the emitted radiation, the means of protecting the patient and the user, and on ways of avoiding misuse and of reducing the risks inherent to installation as far as possible and appropriate. Information regarding the acceptance and performance testing, the acceptance criteria, and the maintenance procedure shall also be specified.16.2.Intended radiation(a)Where devices are designed to emit hazardous, or potentially hazardous, levels of ionizing and/or non-ionizing radiation necessary for a specific medical purpose the benefit of which is considered to outweigh the risks inherent to the emission, it shall be possible for the user to control the emissions. Such devices shall be designed and manufactured to ensure reproducibility of relevant variable parameters within an acceptable tolerance.(b)Where devices are intended to emit hazardous, or potentially hazardous, ionizing and/or non-ionizing radiation, they shall be fitted, where possible, with visual displays and/or audible warnings of such emissions.16.3.Devices shall be designed and manufactured in such a way that exposure of patients, users and other persons to the emission of unintended, stray or scattered radiation is reduced as far as possible. Where possible and appropriate, methods shall be selected which reduce the exposure to radiation of patients, users and other persons who may be affected.16.4.Ionising radiation(a)Devices intended to emit ionizing radiation shall be designed and manufactured taking into account the requirements of the Directive 2013/59/Euratom laying down basic safety standards for protection against the dangers arising from exposure to ionising radiation.(b)Devices intended to emit ionising radiation shall be designed and manufactured in such a way as to ensure that, where possible, taking into account the intended use, the quantity, geometry and quality of the radiation emitted can be varied and controlled, and, if possible, monitored during treatment.(c)Devices emitting ionising radiation intended for diagnostic radiology shall be designed and manufactured in such a way as to achieve an image and/or output quality that are appropriate to the intended medical purpose whilst minimising radiation exposure of the patient and user.(d)Devices that emit ionising radiation and are intended for therapeutic radiology shall be designed and manufactured in such a way as to enable reliable monitoring and control of the delivered dose, the beam type, energy and, where appropriate, the quality of radiation.17.Electronic programmable systems — devices that incorporate electronic programmable systems and software that are devices in themselves17.1.Devices that incorporate electronic programmable systems, including software, or software that are devices in themselves, shall be designed to ensure repeatability, reliability and performance in line with their intended use. In the event of a single fault condition, appropriate means shall be adopted to eliminate or reduce as far as possible consequent risks or impairment of performance.17.2.For devices that incorporate software or for software that are devices in themselves, the software shall be developed and manufactured in accordance with the state of the art taking into account the principles of development life cycle, risk management, including information security, verification and validation.17.3.Software referred to in this Section that is intended to be used in combination with mobile computing platforms shall be designed and manufactured taking into account the specific features of the mobile platform (e.g. size and contrast ratio of the screen) and the external factors related to their use (varying environment as regards level of light or noise).17.4.Manufacturers shall set out minimum requirements concerning hardware, IT networks characteristics and IT security measures, including protection against unauthorised access, necessary to run the software as intended.18.Active devices and devices connected to them18.1.For non-implantable active devices, in the event of a single fault condition, appropriate means shall be adopted to eliminate or reduce as far as possible consequent risks.18.2.Devices where the safety of the patient depends on an internal power supply shall be equipped with a means of determining the state of the power supply and an appropriate warning or indication for when the capacity of the power supply becomes critical. If necessary, such warning or indication shall be given prior to the power supply becoming critical.18.3.Devices where the safety of the patient depends on an external power supply shall include an alarm system to signal any power failure.18.4.Devices intended to monitor one or more clinical parameters of a patient shall be equipped with appropriate alarm systems to alert the user of situations which could lead to death or severe deterioration of the patient's state of health.18.5.Devices shall be designed and manufactured in such a way as to reduce as far as possible the risks of creating electromagnetic interference which could impair the operation of the device in question or other devices or equipment in the intended environment.18.6.Devices shall be designed and manufactured in such a way as to provide a level of intrinsic immunity to electromagnetic interference such that is adequate to enable them to operate as intended.18.7.Devices shall be designed and manufactured in such a way as to avoid, as far as possible, the risk of accidental electric shocks to the patient, user or any other person, both during normal use of the device and in the event of a single fault condition in the device, provided the device is installed and maintained as indicated by the manufacturer.18.8.Devices shall be designed and manufactured in such a way as to protect, as far as possible, against unauthorised access that could hamper the device from functioning as intended.19.Particular requirements for active implantable devices19.1.Active implantable devices shall be designed and manufactured in such a way as to remove or minimize as far as possible:(a)risks connected with the use of energy sources with particular reference, where electricity is used, to insulation, leakage currents and overheating of the devices,(b)risks connected with medical treatment, in particular those resulting from the use of defibrillators or high-frequency surgical equipment, and(c)risks which may arise where maintenance and calibration are impossible, including:excessive increase of leakage currents,ageing of the materials used,excess heat generated by the device,decreased accuracy of any measuring or control mechanism.19.2.Active implantable devices shall be designed and manufactured in such a way as to ensureif applicable, the compatibility of the devices with the substances they are intended to administer, andthe reliability of the source of energy.19.3.Active implantable devices and, if appropriate, their component parts shall be identifiable to allow any necessary measure to be taken following the discovery of a potential risk in connection with the devices or their component parts.19.4.Active implantable devices shall bear a code by which they and their manufacturer can be unequivocally identified (particularly with regard to the type of device and its year of manufacture); it shall be possible to read this code, if necessary, without the need for a surgical operation.20.Protection against mechanical and thermal risks20.1.Devices shall be designed and manufactured in such a way as to protect patients and users against mechanical risks connected with, for example, resistance to movement, instability and moving parts.20.2.Devices shall be designed and manufactured in such a way as to reduce to the lowest possible level the risks arising from vibration generated by the devices, taking account of technical progress and of the means available for limiting vibrations, particularly at source, unless the vibrations are part of the specified performance.20.3.Devices shall be designed and manufactured in such a way as to reduce to the lowest possible level the risks arising from the noise emitted, taking account of technical progress and of the means available to reduce noise, particularly at source, unless the noise emitted is part of the specified performance.20.4.Terminals and connectors to the electricity, gas or hydraulic and pneumatic energy supplies which the user or other person has to handle, shall be designed and constructed in such a way as to minimise all possible risks.20.5.Errors likely to be made when fitting or refitting certain parts which could be a source of risk shall be made impossible by the design and construction of such parts or, failing this, by information given on the parts themselves and/or their housings.The same information shall be given on moving parts and/or their housings where the direction of movement needs to be known in order to avoid a risk.20.6.Accessible parts of devices (excluding the parts or areas intended to supply heat or reach given temperatures) and their surroundings shall not attain potentially dangerous temperatures under normal conditions of use.21.Protection against the risks posed to the patient or user by devices supplying energy or substances21.1.Devices for supplying the patient with energy or substances shall be designed and constructed in such a way that the amount to be delivered can be set and maintained accurately enough to ensure the safety of the patient and of the user.21.2.Devices shall be fitted with the means of preventing and/or indicating any inadequacies in the amount of energy delivered or substances delivered which could pose a danger. Devices shall incorporate suitable means to prevent, as far as possible, the accidental release of dangerous levels of energy or substances from an energy and/or substance source.21.3.The function of the controls and indicators shall be clearly specified on the devices. Where a device bears instructions required for its operation or indicates operating or adjustment parameters by means of a visual system, such information shall be understandable to the user and, as appropriate, the patient.22.Protection against the risks posed by medical devices intended by the manufacturer for use by lay persons22.1.Devices for use by lay persons shall be designed and manufactured in such a way that they perform appropriately for their intended purpose taking into account the skills and the means available to lay persons and the influence resulting from variation that can be reasonably anticipated in the lay person's technique and environment. The information and instructions provided by the manufacturer shall be easy for the lay person to understand and apply.22.2.Devices for use by lay persons shall be designed and manufactured in such a way as to:ensure that the device can be used safely and accurately by the intended user at all stages of the procedure, if necessary after appropriate training and/or information,reduce, as far as possible and appropriate, the risk from unintended cuts and pricks such as needle stick injuries, andreduce as far as possible the risk of error by the intended user in the handling of the device and, if applicable, in the interpretation of the results.22.3.Devices for use by lay persons shall, where appropriate, include a procedure by which the lay person:can verify that, at the time of use, the device will perform as intended by the manufacturer, andif applicable, is warned if the device has failed to provide a valid result.CHAPTER IIIREQUIREMENTS REGARDING THE INFORMATION SUPPLIED WITH THE DEVICE23.Label and instructions for use23.1.General requirements regarding the information supplied by the manufacturerEach device shall be accompanied by the information needed to identify the device and its manufacturer, and by any safety and performance information relevant to the user, or any other person, as appropriate. Such information may appear on the device itself, on the packaging or in the instructions for use, and shall, if the manufacturer has a website, be made available and kept up to date on the website, taking into account the following:(a)The medium, format, content, legibility, and location of the label and instructions for use shall be appropriate to the particular device, its intended purpose and the technical knowledge, experience, education or training of the intended user(s). In particular, instructions for use shall be written in terms readily understood by the intended user and, where appropriate, supplemented with drawings and diagrams.(b)The information required on the label shall be provided on the device itself. If this is not practicable or appropriate, some or all of the information may appear on the packaging for each unit, and/or on the packaging of multiple devices.(c)Labels shall be provided in a human-readable format and may be supplemented by machine-readable information, such as radio-frequency identification (RFID) or bar codes.(d)Instructions for use shall be provided together with devices. By way of exception, instructions for use shall not be required for class I and class IIa devices if such devices can be used safely without any such instructions and unless otherwise provided for elsewhere in this Section.(e)Where multiple devices are supplied to a single user and/or location, a single copy of the instructions for use may be provided if so agreed by the purchaser who in any case may request further copies to be provided free of charge.(f)Instructions for use may be provided to the user in non-paper format (e.g. electronic) to the extent, and only under the conditions, set out in Regulation (EU) No 207/2012 or in any subsequent implementing rules adopted pursuant to this Regulation.(g)Residual risks which are required to be communicated to the user and/or other person shall be included as limitations, contra-indications, precautions or warnings in the information supplied by the manufacturer.(h)Where appropriate, the information supplied by the manufacturer shall take the form of internationally recognised symbols. Any symbol or identification colour used shall conform to the harmonised standards or CS. In areas for which no harmonised standards or CS exist, the symbols and colours shall be described in the documentation supplied with the device.23.2.Information on the labelThe label shall bear all of the following particulars:(a)the name or trade name of the device;(b)the details strictly necessary for a user to identify the device, the contents of the packaging and, where it is not obvious for the user, the intended purpose of the device;(c)the name, registered trade name or registered trade mark of the manufacturer and the address of its registered place of business;(d)if the manufacturer has its registered place of business outside the Union, the name of the authorised representative and address of the registered place of business of the authorised representative;(e)where applicable, an indication that the device contains or incorporates:a medicinal substance, including a human blood or plasma derivative, ortissues or cells, or their derivatives, of human origin, ortissues or cells of animal origin, or their derivatives, as referred to in Regulation (EU) No 722/2012;(f)where applicable, information labelled in accordance with Section 10.4.5.;(g)the lot number or the serial number of the device preceded by the words LOT NUMBER or SERIAL NUMBER or an equivalent symbol, as appropriate;(h)the UDI carrier referred to in Article 27(4) and Part C of Annex VI;(i)an unambiguous indication of t the time limit for using or implanting the device safely, expressed at least in terms of year and month, where this is relevant;(j)where there is no indication of the date until when it may be used safely, the date of manufacture. This date of manufacture may be included as part of the lot number or serial number, provided the date is clearly identifiable;(k)an indication of any special storage and/or handling condition that applies;(l)if the device is supplied sterile, an indication of its sterile state and the sterilisation method;(m)warnings or precautions to be taken that need to be brought to the immediate attention of the user of the device, and to any other person. This information may be kept to a minimum in which case more detailed information shall appear in the instructions for use, taking into account the intended users;(n)if the device is intended for single use, an indication of that fact. A manufacturer's indication of single use shall be consistent across the Union;(o)if the device is a single-use device that has been reprocessed, an indication of that fact, the number of reprocessing cycles already performed, and any limitation as regards the number of reprocessing cycles;(p)if the device is custom-made, the words custom-made device;(q)an indication that the device is a medical device. If the device is intended for clinical investigation only, the words exclusively for clinical investigation;(r)in the case of devices that are composed of substances or of combinations of substances that are intended to be introduced into the human body via a body orifice or applied to the skin and that are absorbed by or locally dispersed in the human body, the overall qualitative composition of the device and quantitative information on the main constituent or constituents responsible for achieving the principal intended action;(s)for active implantable devices, the serial number, and for other implantable devices, the serial number or the lot number.23.3.Information on the packaging which maintains the sterile condition of a device (sterile packaging)The following particulars shall appear on the sterile packaging:(a)an indication permitting the sterile packaging to be recognised as such,(b)a declaration that the device is in a sterile condition,(c)the method of sterilisation,(d)the name and address of the manufacturer,(e)a description of the device,(f)if the device is intended for clinical investigations, the words exclusively for clinical investigations,(g)if the device is custom-made, the words custom-made device,(h)the month and year of manufacture,(i)an unambiguous indication of the time limit for using or implanting the device safely expressed at least in terms of year and month, and(j)an instruction to check the instructions for use for what to do if the sterile packaging is damaged or unintentionally opened before use.23.4.Information in the instructions for useThe instructions for use shall contain all of the following particulars:(a)the particulars referred to in points (a), (c), (e), (f), (k), (l), (n) and (r) of Section 23.2;(b)the device's intended purpose with a clear specification of indications, contra-indications, the patient target group or groups, and of the intended users, as appropriate;(c)where applicable, a specification of the clinical benefits to be expected.(d)where applicable, links to the summary of safety and clinical performance referred to in Article 32;(e)the performance characteristics of the device;(f)where applicable, information allowing the healthcare professional to verify if the device is suitable and select the corresponding software and accessories;(g)any residual risks, contra-indications and any undesirable side-effects, including information to be conveyed to the patient in this regard;(h)specifications the user requires to use the device appropriately, e.g. if the device has a measuring function, the degree of accuracy claimed for it;(i)details of any preparatory treatment or handling of the device before it is ready for use or during its use, such as sterilisation, final assembly, calibration, etc., including the levels of disinfection required to ensure patient safety and all available methods for achieving those levels of disinfection;(j)any requirements for special facilities, or special training, or particular qualifications of the device user and/or other persons;(k)the information needed to verify whether the device is properly installed and is ready to perform safely and as intended by the manufacturer, together with, where relevant:details of the nature, and frequency, of preventive and regular maintenance, and of any preparatory cleaning or disinfection,identification of any consumable components and how to replace them,information on any necessary calibration to ensure that the device operates properly and safely during its intended lifetime, andmethods for eliminating the risks encountered by persons involved in installing, calibrating or servicing devices;(l)if the device is supplied sterile, instructions in the event of the sterile packaging being damaged or unintentionally opened before use;(m)if the device is supplied non-sterile with the intention that it is sterilised before use, the appropriate instructions for sterilisation;(n)if the device is reusable, information on the appropriate processes for allowing reuse, including cleaning, disinfection, packaging and, where appropriate, the validated method of re-sterilisation appropriate to the Member State or Member States in which the device has been placed on the market. Information shall be provided to identify when the device should no longer be reused, e.g. signs of material degradation or the maximum number of allowable reuses;(o)an indication, if appropriate, that a device can be reused only if it is reconditioned under the responsibility of the manufacturer to comply with the general safety and performance requirements;(p)if the device bears an indication that it is for single use, information on known characteristics and technical factors known to the manufacturer that could pose a risk if the device were to be re-used. This information shall be based on a specific section of the manufacturer's risk management documentation, where such characteristics and technical factors shall be addressed in detail. If in accordance with point (d) of Section 23.1. no instructions for use are required, this information shall be made available to the user upon request;(q)for devices intended for use together with other devices and/or general purpose equipment:information to identify such devices or equipment, in order to obtain a safe combination, and/orinformation on any known restrictions to combinations of devices and equipment;(r)if the device emits radiation for medical purposes:detailed information as to the nature, type and where appropriate, the intensity and distribution of the emitted radiation,the means of protecting the patient, user, or other person from unintended radiation during use of the device;(s)information that allows the user and/or patient to be informed of any warnings, precautions, contra-indications, measures to be taken and limitations of use regarding the device. That information shall, where relevant, allow the user to brief the patient about any warnings, precautions, contra-indications, measures to be taken and limitations of use regarding the device. The information shall cover, where appropriate:warnings, precautions and/or measures to be taken in the event of malfunction of the device or changes in its performance that may affect safety,warnings, precautions and/or measures to be taken as regards the exposure to reasonably foreseeable external influences or environmental conditions, such as magnetic fields, external electrical and electromagnetic effects, electrostatic discharge, radiation associated with diagnostic or therapeutic procedures, pressure, humidity, or temperature,warnings, precautions and/or measures to be taken as regards the risks of interference posed by the reasonably foreseeable presence of the device during specific diagnostic investigations, evaluations, or therapeutic treatment or other procedures such as electromagnetic interference emitted by the device affecting other equipment,if the device is intended to administer medicinal products, tissues or cells of human or animal origin, or their derivatives, or biological substances, any limitations or incompatibility in the choice of substances to be delivered,warnings, precautions and/or limitations related to the medicinal substance or biological material that is incorporated into the device as an integral part of the device; andprecautions related to materials incorporated into the device that contain or consist of CMR substances or endocrine-disrupting substances, or that could result in sensitisation or an allergic reaction by the patient or user;(t)in the case of devices that are composed of substances or of combinations of substances that are intended to be introduced into the human body and that are absorbed by or locally dispersed in the human body, warnings and precautions, where appropriate, related to the general profile of interaction of the device and its products of metabolism with other devices, medicinal products and other substances as well as contra-indications, undesirable side-effects and risks relating to overdose;(u)in the case of implantable devices, the overall qualitative and quantitative information on the materials and substances to which patients can be exposed;(v)warnings or precautions to be taken in order to facilitate the safe disposal of the device, its accessories and the consumables used with it, if any. This information shall cover, where appropriate:infection or microbial hazards such as explants, needles or surgical equipment contaminated with potentially infectious substances of human origin, andphysical hazards such as from sharps.If in accordance with the point (d) of Section 23.1 no instructions for use are required, this information shall be made available to the user upon request;(w)for devices intended for use by lay persons, the circumstances in which the user should consult a healthcare professional;(x)for the devices covered by this Regulation pursuant to Article 1(2), information regarding the absence of a clinical benefit and the risks related to use of the device;(y)date of issue of the instructions for use or, if they have been revised, date of issue and identifier of the latest revision of the instructions for use;(z)a notice to the user and/or patient that any serious incident that has occurred in relation to the device should be reported to the manufacturer and the competent authority of the Member State in which the user and/or patient is established;(aa)information to be supplied to the patient with an implanted device in accordance with Article 18;(ab)for devices that incorporate electronic programmable systems, including software, or software that are devices in themselves, minimum requirements concerning hardware, IT networks characteristics and IT security measures, including protection against unauthorised access, necessary to run the software as intended.

The full entry, with the citation mapping v1 = 02017R0745-20250110, v2 = 02017R0745-20260101, is committed at eu/32017R0745/CHANGELOG.md.

02017R0745-2024070902017R0745-20250110

in force 2025-01-01, 2025-01-10 · detected 2026-08-10

2 provisions touched — 2 substantive, 0 date-only, 0 disputed · 0 sentences quoted verbatim by the gate, 0 changes shipped without an explanation

INSERTED Art. 10a — Obligations in case of interruption or discontinuation of supply of certain devices · applies from unknown (an inserted provision states its own application date only in prose)

A new Article 10a is added, requiring a manufacturer that anticipates an interruption or discontinuation of supply of a device (other than a custom-made device) to notify the competent authority of the Member State where it or its authorised representative is established, as well as the economic operators, health institutions and healthcare professionals it directly supplies, when serious harm or a risk of serious harm to patients or public health is reasonably foreseeable, with that notice due at least six months beforehand except in exceptional circumstances and including the reasons for the interruption or discontinuation. Art. 10a, v2

The article also newly requires the receiving competent authority to inform, without undue delay, the competent authorities of other Member States and the Commission, and requires economic operators who received such information to pass it on without undue delay to other economic operators, health institutions and healthcare professionals they directly supply. Art. 10a, v2

text before / after

inserted text (02017R0745-20250110)

Article 10aObligations in case of interruption or discontinuation of supply of certain devices1.Where a manufacturer anticipates an interruption or a discontinuation of the supply of a device, other than a custom-made device, and where it is reasonably foreseeable that such interruption or discontinuation could result in serious harm or a risk of serious harm to patients or public health in one or more Member States, the manufacturer shall inform the competent authority of the Member State where it or its authorised representative is established, as well as the economic operators, health institutions and healthcare professionals to whom it directly supplies the device, of the anticipated interruption or discontinuation.The information referred to in the first subparagraph shall, other than in exceptional circumstances, be provided at least 6 months before the anticipated interruption or discontinuation. The manufacturer shall specify the reasons for the interruption or discontinuation in the information provided to the competent authority.2.The competent authority that has received the information referred to in paragraph 1 shall, without undue delay, inform the competent authorities of the other Member States and the Commission of the anticipated interruption or discontinuation.3.The economic operators who have received the information from the manufacturer in accordance with paragraph 1 or from another economic operator in the supply chain shall, without undue delay, inform any other economic operators, health institutions and healthcare professionals to whom they directly supply the device, of the anticipated interruption or discontinuation.

MODIFIED Art. 106 — Provision of scientific, technical and clinical opinions and advice · applies from unknown (the text changed beyond its dates; the applicability binding is prose)

Paragraph 14 now describes the fees under paragraph 13 as "established" rather than "payable to the Commission," and the cross-reference to Annex IX is reformatted from "point (c) of Section 5.1" to "Section 5.1, point (c)". Art. 106, v1 Art. 106, v2

Paragraph 14 adds two new sentences specifying that fees related to advice from expert panels are payable to EMA under Article 30, point (f), of Regulation (EU) 2022/123, while fees related to advice from expert laboratories remain payable to the Commission. Art. 106, v2

text before / after

02017R0745-2024070902017R0745-20250110

Article 106Provision of scientific, technical and clinical opinions and advice1.The Commission shall, by means of implementing acts and in consultation with the MDCG, make provision for expert panels to be designated for the assessment of the clinical evaluation in relevant medical fields as referred to in paragraph 9 of this Article and to provide views in accordance with Article 48(6) of Regulation (EU) 2017/746 on the performance evaluation of certain in vitro diagnostic medical devices and, where necessary, for categories or groups of devices, or for specific hazards relating to categories or groups of devices, observing the principles of highest scientific competence, impartiality, independence and transparency. The same principles shall apply where the Commission decides to appoint expert laboratories in accordance with paragraph 7 of this Article.2.Expert panels and expert laboratories may be designated in areas where the Commission, in consultation with the MDCG, has identified a need for the provision of consistent scientific, technical and/or clinical advice or laboratory expertise in relation to the implementation of this Regulation. Expert panels and expert laboratories may be appointed on a standing or temporary basis.3.Expert panels shall consist of advisors appointed by the Commission on the basis of up-to-date clinical, scientific or technical expertise in the field and with a geographical distribution that reflects the diversity of scientific and clinical approaches in the Union. The Commission shall determine the number of members of each panel in accordance with the requisite needs.The members of expert panels shall perform their tasks with impartiality and objectivity. They shall neither seek nor take instructions from notified bodies or manufacturers. Each member shall draw up a declaration of interests, which shall be made publicly available.The Commission shall establish systems and procedures to actively manage and prevent potential conflicts of interest.4.Expert panels shall take into account relevant information provided by stakeholders including patients' organisations and healthcare professionals when preparing their scientific opinions.5.The Commission, following consultation with the MDCG, may appoint advisors to expert panels following publication in the Official Journal of the European Union and on the Commission website following a call for expressions of interest. Depending on the type of task and the need for specific expertise, advisors may be appointed to the expert panels for a maximum period of three years and their appointment may be renewed.6.The Commission, following consultation with the MDCG, may include advisors on a central list of available experts who, whilst not being formally appointed to a panel, are available to provide advice and to support the work of the expert panel as needed. That list shall be published on the Commission website.7.The Commission may, by means of implementing acts and following consultation with the MDCG, designate expert laboratories, on the basis of their expertise in:physico-chemical characterisation, ormicrobiological, biocompatibility, mechanical, electrical, electronic or non-clinical biological and toxicological testingof specific devices, categories or groups of devices.The Commission shall only designate expert laboratories for which a Member State or the Joint Research Centre has submitted an application for designation.8.Expert laboratories shall satisfy the following criteria:(a)have adequate and appropriately qualified staff with adequate knowledge and experience in the field of the devices for which they are designated;(b)possess the necessary equipment to carry out the tasks assigned to them;(c)have the necessary knowledge of international standards and best practices;(d)have an appropriate administrative organisation and structure;(e)ensure that their staff observe the confidentiality of information and data obtained in carrying out their tasks.9.Expert panels appointed for clinical evaluation in relevant medical fields shall fulfil the tasks provided for in Article 54(1) and Article 61(2) and Section 5.1 of Annex IX or Section 6 of Annex X, as applicable.10.Expert panels and expert laboratories may have the following tasks, depending on the requisite needs:(a)to provide scientific, technical and clinical assistance to the Commission and the MDCG in relation to the implementation of this Regulation;(b)to contribute to the development and maintenance of appropriate guidance and CS for:clinical investigations,clinical evaluation and PMCF,performance studies,performance evaluation and post-market performance follow-up,physico-chemical characterisation, andmicrobiological, biocompatibility, mechanical, electrical, electronic or non-clinical toxicological testingfor specific devices, or a category or group of devices, or for specific hazards related to a category or group of devices;(c)to develop and review clinical evaluation guidance and performance evaluation guidance for performance of conformity assessment in line with the state of the art with regard to clinical evaluation, performance evaluation, physico-chemical characterisation, and microbiological, biocompatibility, mechanical, electrical, electronic or non-clinical toxicological testing;(d)to contribute to the development of standards at international level, ensuring that such standards reflect the state of the art;(e)to provide opinions in response to consultations by manufacturers in accordance with Article 61(2), notified bodies and Member States in accordance with paragraphs 11 to 13 of this Article.(f)to contribute to identification of concerns and emerging issues on the safety and performance of medical devices;(g)to provide views in accordance with Article 48(4) of Regulation (EU) 2017/746 on the performance evaluation of certain in vitro diagnostic medical devices.11.The Commission, shall facilitate the access of Member States and notified bodies and manufacturers to advice provided by expert panels and expert laboratories concerning, inter alia, the criteria for an appropriate data set for assessment of the conformity of a device, in particular with regard to the clinical data required for clinical evaluation, with regard to physico-chemical characterisation, and with regard to microbiological, biocompatibility, mechanical, electrical, electronic and non-clinical toxicological testing.12.When adopting its scientific opinion in accordance with paragraph 9, the members of the expert panels shall use their best endeavours to reach consensus. If consensus cannot be reached, the expert panels shall decide by a majority of their members, and the scientific opinion shall mention the divergent positions and the grounds on which they are based.The Commission shall publish the scientific opinion and advice delivered in accordance with paragraphs 9 and 11 of this Article, ensuring consideration of aspects of confidentiality as set out in Article 109. The clinical evaluation guidance referred to in point (c) of paragraph 10 shall be published following consultation with the MDCG.13.The Commission may require manufacturers and notified bodies to pay fees for the advice provided by expert panels and expert laboratories. The structure and the level of fees as well as the scale and structure of recoverable costs shall be adopted by the Commission by means of implementing acts, taking into account the objectives of the adequate implementation of this Regulation, protection of health and safety, support of innovation and cost-effectiveness and the necessity to achieve active participation in the expert panels. Those implementing acts shall be adopted in accordance with the examination procedure referred to in Article 114(3).14.The fees payable to the Commission established in accordance with the procedure under paragraph 13 of this Article shall be set in a transparent manner and on the basis of the costs for the services provided. The fees payable shall be reduced in the case of a clinical evaluation consultation procedure initiated in accordance with Section 5.1, point (c) of Section 5.1 (c), of Annex IX involving a manufacturer who is a micro, small or medium-sized enterprise within the meaning of Recommendation 2003/361/EC.15.The 2003/361/EC.The fees related to the advice provided by expert panels are payable to EMA pursuant to Article 30, point (f), of Regulation (EU) 2022/123 of the European Parliament and of the CouncilRegulation (EU) 2022/123 of the European Parliament and of the Council of 25 January 2022 on a reinforced role for the European Medicines Agency in crisis preparedness and management for medicinal products and medical devices (OJ L 20, 31.1.2022, p. 1)..The fees related to the advice provided by expert laboratories are payable to the Commission.15.The Commission is empowered to adopt delegated acts in accordance with Article 115 to amend the tasks of expert panels and expert laboratories referred to in paragraph 10 of this Article.

The full entry, with the citation mapping v1 = 02017R0745-20240709, v2 = 02017R0745-20250110, is committed at eu/32017R0745/CHANGELOG.md.

02017R0745-2023032002017R0745-20240709

in force 2024-07-09 · detected 2026-08-10

6 provisions touched — 6 substantive, 0 date-only, 1 disputed · 0 sentences quoted verbatim by the gate, 1 changes shipped without an explanation

INSERTED Art. 10a · applies from unknown

Disputed — seen by the instruction parse, not by the structural diff, corpus metadata.

No explanation shipped — the structural diff did not see this change, so it carries no text; another signal named the unit and the disagreement ships as `disputed`.

text before / after

No text on either side: this unit was named by a signal that carries no text, and only the structural diff carries any.

MODIFIED Art. 34 — Functionality of Eudamed · applies from unknown (the text changed beyond its dates; the applicability binding is prose)

The sentence in paragraph 1 setting a target date of 25 March 2021 for full Eudamed functionality and referencing the deadlines in Article 123 of this Regulation and Article 113 of Regulation (EU) 2017/746 has been removed, leaving only the plan-drafting requirement by 26 May 2018. Art. 34, v1 Art. 34, v2

Paragraph 2 now has the Commission inform the MDCG based on independent audit reports (plural) that one or more of the electronic systems referred to in Article 33(2) are functional, rather than reporting on a single independent audit report confirming that Eudamed as a whole has achieved full functionality. Art. 34, v1 Art. 34, v2

text before / after

02017R0745-2023032002017R0745-20240709

Article 34Functionality of Eudamed1.The Commission shall, in collaboration with the MDCG, draw up the functional specifications for Eudamed. The Commission shall draw up a plan for the implementation of those specifications by 26 May 2018. That plan 2018.2.The Commission shall seek to ensure that Eudamed is fully functional at a date that allows inform the Commission to publish the notice referred to in paragraph 3 of this Article by 25 March 2021 and that all other relevant deadlines laid down in Article 123 of this Regulation and in Article 113 of Regulation (EU) 2017/746 are met.2.The Commission shall, MDCG when, on the basis of an independent audit report, inform the MDCG when reports, it has verified that Eudamed has achieved full functionality one or more of the electronic systems referred to in Article 33(2) are functional and Eudamed meets meet the functional specifications drawn up pursuant to paragraph 1.3.The 1 of this Article.3.The Commission shall, after consultation with the MDCG and when it is satisfied that the conditions referred to in paragraph 2 have been fulfilled, publish a notice to that effect in the Official Journal of the European Union.

MODIFIED Art. 78 — Coordinated assessment procedure for clinical investigations · applies from unknown (the text changed beyond its dates; the applicability binding is prose)

Paragraph 14 no longer sets fixed calendar dates of 25 May 2027 and 26 May 2027 for the transition to mandatory use of the coordinated assessment procedure. Art. 78, v1 Art. 78, v2

Instead, the revised text ties full mandatory application by all Member States to a date calculated as five years from publication of the notice referred to in Article 34(3) confirming that the electronic system is functional, and it provides that before that date, and at the earliest six months after that notice's publication, the procedure applies only to Member States that have agreed to apply it. Art. 78, v2

text before / after

02017R0745-2023032002017R0745-20240709

Article 78Coordinated assessment procedure for clinical investigations1.By means of the electronic system referred to in Article 73, the sponsor of a clinical investigation to be conducted in more than one Member State may submit, for the purpose of Article 70, a single application that, upon receipt, is transmitted electronically to all Member States in which the clinical investigation is to be conducted.2.The sponsor shall propose in the single application referred to in paragraph 1 that one of the Member States in which the clinical investigation is to be conducted acts as coordinating Member State. The Member States in which the clinical investigation is to be conducted shall, within six days of submission of the application, agree on one of them taking the role of the coordinating Member State. If they do not agree on a coordinating Member State, the coordinating Member State proposed by the sponsor shall assume that role.3.Under the direction of the coordinating Member State referred to in paragraph 2, the Member States concerned shall coordinate their assessment of the application, in particular of the documentation referred to in Chapter II of Annex XV.However, the completeness of the documentation referred to in Sections 1.13, 3.1.3, 4.2, 4.3 and 4.4 of Chapter II of Annex XV shall be assessed separately by each Member State concerned in accordance with Article 70(1) to (5).4.With regard to documentation other than that referred to in the second subparagraph of paragraph 3, the coordinating Member State shall:(a)within six days of receipt of the single application, notify the sponsor that it is the coordinating Member State (notification date);(b)for the purpose of the validation of the application, take into account any considerations submitted within seven days of the notification date by any Member State concerned;(c)within 10 days of the notification date, assess whether the clinical investigation falls within the scope of this Regulation and whether the application is complete, and shall notify the sponsor accordingly. Article 70(1) and (3) to (5) shall apply to the coordinating Member State in relation to that assessment;(d)establish the results of its assessment in a draft assessment report to be transmitted within 26 days of the validation date to the Member States concerned. By day 38 after the validation date, the other Member States concerned shall transmit their comments and proposals on the draft assessment report and the underlying application to the coordinating Member State which shall take due account of those comments and proposals in its finalisation of the final assessment report, to be transmitted within 45 days of the validation date to the sponsor and the other Member States concerned.The final assessment report shall be taken into account by all Member States concerned when deciding on the sponsor's application in accordance with Article 70(7).5.As regards the assessment of the documentation referred to in the second subparagraph of paragraph 3, each Member State concerned may request, on a single occasion, additional information from the sponsor. The sponsor shall submit the requested additional information within the period set by the Member State concerned, which shall not exceed 12 days from the receipt of the request. The expiry of the last deadline pursuant to point (d) of paragraph 4 shall be suspended from the date of the request until such time as the additional information has been received.6.For class IIb and class III devices, the coordinating Member State may also extend the periods referred to in paragraph 4 by a further 50 days, for the purpose of consulting with experts.7.The Commission may, by means of implementing acts, further specify the procedures and timescales for coordinated assessments to be taken into account by Member States concerned when deciding on the sponsor's application. Such implementing acts may also set out the procedures and timescales for coordinated assessment in the case of substantial modifications pursuant to paragraph 12 of this Article, in the case of reporting of adverse events pursuant to Article 80(4) and in the case of clinical investigations of combination products between medical devices and medicinal products, where the latter are under a concurrent coordinated assessment of a clinical trial under Regulation (EU) No 536/2014. Those implementing acts shall be adopted in accordance with the examination procedure referred to in Article 114(3).8.Where the conclusion of the coordinating Member State concerning the area of coordinated assessment is that the conduct of the clinical investigation is acceptable or acceptable subject to compliance with specific conditions, that conclusion shall be deemed to be the conclusion of all Member States concerned.Notwithstanding the first subparagraph, a Member State concerned may only disagree with the conclusion of the coordinating Member State concerning the area of coordinated assessment on the following grounds:(a)when it considers that participation in the clinical investigation would lead to a subject receiving treatment inferior to that received in normal clinical practice in that Member State concerned;(b)infringement of national law; or(c)considerations as regards subject safety and data reliability and robustness submitted under point (d) of paragraph 4.Where one of the Member States concerned disagrees with the conclusion on the basis of the second subparagraph of this paragraph, it shall communicate its disagreement, together with a detailed justification, through the electronic system referred to in Article 73, to the Commission, to all other Member States concerned and to the sponsor.9.Where the conclusion of the coordinating Member State concerning the area of coordinated assessment is that the clinical investigation is not acceptable, that conclusion shall be deemed to be the conclusion of all Member States concerned.10.A Member State concerned shall refuse to authorise a clinical investigation if it disagrees with the conclusion of the coordinating Member State as regards any of the grounds referred to in the second subparagraph of paragraph 8, or if it finds, on duly justified grounds, that the aspects addressed in Sections 1.13, 3.1.3, 4.2, 4.3 and 4.4 of Chapter II of Annex XV are not complied with, or where an ethics committee has issued a negative opinion in relation to that clinical investigation, which is valid, in accordance with national law, for that entire Member State. That Member State shall provide for an appeal procedure in respect of such refusal.11.Each Member State concerned shall notify the sponsor through the electronic system referred to in Article 73 as to whether the clinical investigation is authorised, whether it is authorised subject to conditions, or whether authorisation has been refused. Notification shall be done by way of one single decision within five days of the transmission, pursuant to point (d) of paragraph 4, by the coordinating Member State of the final assessment report. Where an authorisation of a clinical investigation is subject to conditions, those conditions may only be such that, by their nature, they cannot be fulfilled at the time of that authorisation.12.Any substantial modifications as referred to in Article 75 shall be notified to the Member States concerned by means of the electronic system referred to in Article 73. Any assessment as to whether there are grounds for disagreement as referred to in the second subparagraph of paragraph 8 of this Article shall be carried out under the direction of the coordinating Member State, except for substantial modifications concerning Sections 1.13, 3.1.3, 4.2, 4.3 and 4.4 of Chapter II of Annex XV, which shall be assessed separately by each Member State concerned.13.The Commission shall provide administrative support to the coordinating Member State in the accomplishment of its tasks under this Chapter.14.The Chapter.14.All Member States shall be required to apply the procedure set out in this Article shall, until 25 May 2027, from the date corresponding to 5 years from the date of publication of the notice referred to in Article 34(3), informing that the electronic system referred to in Article 33(2), point (e), is functional and meets the functional specifications drawn up pursuant to Article 34(1).Before the date set out in the first subparagraph of this paragraph and at the earliest 6 months from the date of publication of the notice referred to in that subparagraph, the procedure set out in this Article shall be applied only by those of the Member States in which the clinical investigation is to be conducted which have agreed to apply it. From 26 May 2027, all Member States shall be required to apply that procedure.

MODIFIED Art. 120 — Transitional provisions · applies from unchanged

The after text adds a new paragraph 13 to Article 120, which states that Article 10a shall also apply to devices referred to in paragraphs 3a and 3b of this Article. Art. 120, v2

This paragraph does not appear in the before version of Article 120, which ends at paragraph 12. Art. 120, v1

text before / after

02017R0745-2023032002017R0745-20240709

Article 120Transitional provisions1.From 26 May 2021, any publication of a notification in respect of a notified body in accordance with Directives 90/385/EEC and 93/42/EEC shall become void.2.Certificates issued by notified bodies in accordance with Directives 90/385/EEC and 93/42/EEC prior to 25 May 2017 shall remain valid until the end of the period indicated on the certificate, except for certificates issued in accordance with Annex 4 to Directive 90/385/EEC or Annex IV to Directive 93/42/EEC which shall become void at the latest on 27 May 2022.Certificates issued by notified bodies in accordance with Directives 90/385/EEC and 93/42/EEC from 25 May 2017 that were still valid on 26 May 2021 and that have not been withdrawn afterwards shall remain valid after the end of the period indicated on the certificate until the date set out in paragraph 3a of this Article applicable for the relevant risk class of the devices. Certificates issued by notified bodies in accordance with those Directives from 25 May 2017 that were still valid on 26 May 2021 and that have expired before 20 March 2023 shall be considered to be valid until the dates set out in paragraph 3a of this Article only if one of the following conditions is fulfilled:(a)before the date of expiry of the certificate, the manufacturer and a notified body have signed a written agreement in accordance with Section 4.3, second subparagraph, of Annex VII to this Regulation for the conformity assessment in respect of the device covered by the expired certificate or in respect of a device intended to substitute that device;(b)a competent authority of a Member State has granted a derogation from the applicable conformity assessment procedure in accordance with Article 59(1) of this Regulation or has required the manufacturer, in accordance with Article 97(1) of this Regulation, to carry out the applicable conformity assessment procedure.3.By way of derogation from Article 5 and provided the conditions set out in paragraph 3c of this Article are met, devices referred to in paragraphs 3a and 3b of this Article may be placed on the market or put into service until the dates set out in those paragraphs.3a.Devices which have a certificate that was issued in accordance with Directive 90/385/EEC or Directive 93/42/EEC and that is valid by virtue of paragraph 2 of this Article may be placed on the market or put into service until the following dates:(a)31 December 2027, for all class III devices, and for class IIb implantable devices except sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors;(b)31 December 2028, for class IIb devices other than those covered by point (a) of this paragraph, for class IIa devices, and for class I devices placed on the market in sterile condition or having a measuring function.3b.Devices for which the conformity assessment procedure pursuant to Directive 93/42/EEC did not require the involvement of a notified body, for which the declaration of conformity was drawn up prior to 26 May 2021 and for which the conformity assessment procedure pursuant to this Regulation requires the involvement of a notified body, may be placed on the market or put into service until 31 December 2028.3c.Devices referred to in paragraphs 3a and 3b of this Article may be placed on the market or put into service until the dates referred to in those paragraphs only if the following conditions are met:(a)those devices continue to comply with Directive 90/385/EEC or Directive 93/42/EEC, as applicable;(b)there are no significant changes in the design and intended purpose;(c)the devices do not present an unacceptable risk to the health or safety of patients, users or other persons, or to other aspects of the protection of public health;(d)no later than 26 May 2024, the manufacturer has put in place a quality management system in accordance with Article 10(9);(e)no later than 26 May 2024, the manufacturer or the authorised representative has lodged a formal application with a notified body in accordance with Section 4.3, first subparagraph, of Annex VII for conformity assessment in respect of a device referred to in paragraph 3a or 3b of this Article or in respect of a device intended to substitute that device, and, no later than 26 September 2024, the notified body and the manufacturer have signed a written agreement in accordance with Section 4.3, second subparagraph, of Annex VII.3d.By way of derogation from paragraph 3 of this Article, the requirements of this Regulation relating to post-market surveillance, market surveillance, vigilance, registration of economic operators and of devices shall apply to devices referred to in paragraphs 3a and 3b of this Article in place of the corresponding requirements in Directives 90/385/EEC and 93/42/EEC.3e.Without prejudice to Chapter IV and paragraph 1 of this Article, the notified body that issued the certificate referred to in paragraph 3a of this Article shall continue to be responsible for the appropriate surveillance in respect of the applicable requirements relating to the devices it has certified, unless the manufacturer has agreed with a notified body designated in accordance with Article 42 that the latter shall carry out such surveillance.No later than 26 September 2024, the notified body that has signed the written agreement referred to in paragraph 3c, point (e), of this Article shall be responsible for the surveillance in respect of the devices covered by the written agreement. Where the written agreement covers a device intended to substitute a device which has a certificate that was issued in accordance with Directive 90/385/EEC or Directive 93/42/EEC, the surveillance shall be conducted in respect of the device that is being substituted.The arrangements for the transfer of the surveillance from the notified body that issued the certificate to the notified body designated in accordance with Article 42 shall be clearly defined in an agreement between the manufacturer and the notified body designated in accordance with Article 42 and, where practicable, the notified body that issued the certificate. The notified body designated in accordance with Article 42 shall not be responsible for conformity assessment activities carried out by the notified body that issued the certificate.3f.By way of derogation from Article 5, class III custom-made implantable devices may be placed on the market or put into service until 26 May 2026 without a certificate issued by a notified body in accordance with the conformity assessment procedure referred to in Article 52(8), second subparagraph, provided that no later than 26 May 2024, the manufacturer or the authorised representative has lodged a formal application with a notified body in accordance with Section 4.3, first subparagraph, of Annex VII for conformity assessment, and, no later than 26 September 2024, the notified body and the manufacturer have signed a written agreement in accordance with Section 4.3, second subparagraph, of Annex VII.4.Devices lawfully placed on the market pursuant to Directives 90/385/EEC and 93/42/EEC prior to 26 May 2021, and devices lawfully placed on the market from 26 May 2021 pursuant to paragraphs 3, 3a, 3b and 3f of this Article, may continue to be made available on the market or put into service.5.By way of derogation from Directives 90/385/EEC and 93/42/EEC, devices which comply with this Regulation may be placed on the market prior to 26 May 2021.6.By way of derogation from Directives 90/385/EEC and 93/42/EEC, conformity assessment bodies which comply with this Regulation may be designated and notified prior to 26 May 2021. Notified bodies which are designated and notified in accordance with this Regulation may carry out the conformity assessment procedures laid down in this Regulation and issue certificates in accordance with this Regulation prior to 26 May 2021.7.As regards devices subject to the consultation procedure laid down in Article 54, paragraph 5 of this Article shall apply provided that the necessary appointments to the MDCG and expert panels have been made.8.By way of derogation from Article 10a, point (a) of Article 10b(1) and Article 11(5) of Directive 90/385/EEC and Article 14(1) and (2), points (a) and (b) of Article 14a(1) and Article 16(5) of Directive 93/42/EEC, manufacturers, authorised representatives, importers and notified bodies which, during the period starting on the later of the dates referred to in point (d) of Article 123(3) and ending 18 months later, comply with Articles 29(4), 31(1) and 56(5) of this Regulation shall be considered to comply with the laws and regulations adopted by Member States in accordance with, respectively, Article 10a of Directive 90/385/EEC or Article 14(1) and (2) of Directive 93/42/EEC, with, respectively, point (a) of Article 10b(1) of Directive 90/385/EEC or points (a) and (b) of Article 14a(1) of Directive 93/42/EEC and with, respectively, Article 11(5) of Directive 90/385/EEC or Article 16(5) of Directive 93/42/EEC, as specified in Decision 2010/227/EU.9.Authorisations granted by the competent authorities of the Member States in accordance with Article 9(9) of Directive 90/385/EEC or Article 11(13) of Directive 93/42/EEC shall keep the validity indicated in the authorisation.10.Devices falling within the scope of this Regulation in accordance with point (g) of Article 1(6) which have been legally placed on the market or put into service in accordance with the rules in force in the Member States prior to 26 May 2021 may continue to be placed on the market and put into service in the Member States concerned.11.Clinical investigations which have started to be conducted in accordance with Article 10 of Directive 90/385/EEC or Article 15 of Directive 93/42/EEC prior to 26 May 2021 may continue to be conducted. As of 26 May 2021, however, the reporting of serious adverse events and device deficiencies shall be carried out in accordance with this Regulation.12.Until the Commission has designated, pursuant to Article 27(2), issuing entities, GS1, HIBCC and ICCBBA shall be considered to be designated issuing entities. entities.13.Article 10a shall also apply to devices referred to in paragraphs 3a and 3b of this Article.

MODIFIED Art. 122 — Repeal · applies from unchanged

The list of exceptions to the repeal of Directives 90/385/EEC and 93/42/EEC was restructured, merging what had been four separate bullet points into two, combining the vigilance and clinical investigation provisions of both Directives into one entry and the registration and certificate notification provisions of both Directives into another. Art. 122, v2

Where the earlier version tied the repeal effect of these provisions to the later of the dates referred to in Article 123(3), point (d), the later version ties it to that same date but expressed as applicable to specific electronic systems referenced in Article 33(2), points (e) and (f) for the vigilance and clinical investigation group, and points (a) to (d) for the registration and certificate notification group. Art. 122, v1 Art. 122, v2

The separate 18-months-after clause that previously governed the registration and certificate notification provisions was removed, and those provisions are now tied directly to the date in Article 123(3), point (d), in the same manner as the vigilance provisions. Art. 122, v1 Art. 122, v2

text before / after

02017R0745-2023032002017R0745-20240709

Article 122RepealWithout prejudice to Article 120(3) to (3e) and (4) of this Regulation, and without prejudice to the obligations of the Member States and manufacturers as regards vigilance and to the obligations of manufacturers as regards the making available of documentation, under Directives 90/385/EEC and 93/42/EEC, those Directives are repealed with effect from 26 May 2021, with the exception of:Articles 8 and 10, Article 10b(1), points (b) and (c) of Article 10b(1), (c), and Article 10b(2) and Article 10b(3) (3) of Directive 90/385/EEC, and the obligations relating to vigilance and clinical investigations provided for in the corresponding Annexes, which are repealed with effect from the later of the dates referred to in point (d) of Article 123(3) of this Regulation;Article 10a, point (a) of 10, Article 10b(1) and Article 11(5) of Directive 90/385/EEC, and the obligations relating to registration of devices and economic operators, and to certificate notifications, provided for in the corresponding Annexes, which are repealed with effect from 18 months after the later of the dates referred to in point (d) of Article 123(3) of this Regulation;Article 10, 14a(1), points (c) and (d) of (d), Article 14a(1), Article 14a(2), Article 14a(3) 14a(2) and (3) and Article 15 of Directive 93/42/EEC, and the obligations relating to vigilance and clinical investigations provided for in the corresponding Annexes, Annexes to those Directives, which are repealed repealed, as applicable, with effect from the later of the dates date referred to in Article 123(3), point (d) of Article 123(3) (d), of this Regulation; andArticle Regulation in respect of the application of the obligations and requirements that relate to the electronic systems referred to in Article 33(2), points (e) and (f), respectively, of this Regulation;Article 10a, Article 10b(1), point (a), and Article 11(5) of Directive 90/385/EEC, Article 14(1) and (2), Article 14a(1), points (a) and (b) of Article 14a(1) (b), and Article 16(5) of Directive 93/42/EEC, and the obligations relating to registration of devices and economic operators, and to certificate notifications, provided for in the corresponding Annexes, Annexes to those Directives, which are repealed repealed, as applicable, with effect from 18 months after the later of the dates date referred to in Article 123(3), point (d) of Article 123(3) (d), of this Regulation; andArticle Regulation in respect of the application of the obligations and requirements that relate to the electronic systems referred to in Article 33(2), points (a) to (d), respectively, of this Regulation;Article 9(9) of Directive 90/385/EEC and Article 11(13) of Directive 93/42/EEC, which are repealed with effect from 24 April 2020.As regards the devices referred to in Article 120(3) to (3e) and (4) of this Regulation, the Directives referred to in the first paragraph of this Article shall continue to apply to the extent necessary for the application of those paragraphs.Notwithstanding the first paragraph, Regulations (EU) No 207/2012 and (EU) No 722/2012 shall remain in force and continue to apply unless and until repealed by implementing acts adopted by the Commission pursuant to this Regulation.References to the repealed Directives shall be understood as references to this Regulation and shall be read in accordance with the correlation table laid down in Annex XVII to this Regulation.

MODIFIED Art. 123 — Entry into force and date of application · applies from unchanged

Point (d) no longer conditions the delayed application of Eudamed-related obligations on Eudamed not being 'fully functional' due to unforeseen circumstances, and instead ties the six-month delay to publication of a notice confirming that the relevant electronic system referred to in Article 33(2) is functional and meets its functional specifications, adding Article 56(5) and Article 78(1) to (13) to the listed provisions and rephrasing the transitional rule on continued application of the Directives to run until the date of application of those listed provisions. Art. 123, v1 Art. 123, v2

Point (e), previously a short rule applying Article 29(4) and Article 56(5) from 18 months after the date in point (d), is replaced by a new obligation on manufacturers to enter Eudamed information under Article 29 within 12 months of the notice for the electronic system in Article 33(2)(a) and (b), covering devices listed in new sub-points (i) and (ii). Art. 123, v1 Art. 123, v2

Three new points, (ea), (eb) and (ec), are added, setting out separate obligations for notified bodies to enter certificate information under Article 56(5) within 18 months, for manufacturers to upload summaries of safety and clinical performance and notify competent authorities once certificates are entered under point (ea), and for manufacturers to register devices in the electronic systems under Article 33(2)(a) and (b) when submitting PSURs or vigilance reports through the system under Article 33(2)(f). Art. 123, v2

text before / after

02017R0745-2023032002017R0745-20240709

Article 123Entry into force and date of application1.This Regulation shall enter into force on the twentieth day following that of its publication in the Official Journal of the European Union.2.It shall apply from 26 May 2021.3.By way of derogation from paragraph 2:(a)Articles 35 to 50 shall apply from 26 November 2017. However, from that date until 26 May 2021, the obligations on notified bodies pursuant to Articles 35 to 50 shall apply only to those bodies which submit an application for designation in accordance with Article 38;(b)Articles 101 and 103 shall apply from 26 November 2017;(c)Article 102 shall apply from 26 May 2018;(d)without prejudice to the obligations on of the Commission pursuant to Article 34, where, due to circumstances that could not reasonably have been foreseen when drafting the plan referred to in Article 34(1), Eudamed is not fully functional on 26 May 2021, the obligations and requirements that relate to Eudamed any of the electronic systems referred to in Article 33(2) shall apply from the date corresponding to six 6 months after from the date of publication of the notice referred to in Article 34(3). 34(3), informing that the relevant electronic system is functional and meets the functional specifications drawn up pursuant to Article 34(1). The provisions referred to in the preceding sentence are:Article 29,Article 31,Article 32,Article 33(4),the second sentence of Article 40(2),Article 42(10),Article 43(2),the second subparagraph of Article 44(12),points (d) and (e) of Article 46(7),Article 53(2),Article 54(3),Article 55(1),Articles 55(1),Article 56(5),Articles 70 to 77,paragraphs 1 77,Article 78(1) to 13 of (13), without prejudice to Article 78,Articles 78(14),Articles 79 to 82,Article 86(2),Articles 87 and 88,Article 89(5) and (7), and the third subparagraph of Article 89(8),Article 90,Article 93(4), (7) and (8),Article 95(2) and (4),the last sentence of Article 97(2),Article 99(4),Article 120(3d).Until Eudamed is fully functional, the date of application of the provisions referred to in the first subparagraph of this point, the corresponding provisions of Directives 90/385/EEC and 93/42/EEC shall continue to apply for the purpose of meeting the obligations laid down in the provisions listed in the first paragraph of this point regarding exchange of information including, and in particular, information regarding on vigilance reporting, clinical investigations, registration of devices and economic operators, and certificate notifications.(e)Article 29(4) notifications shall continue to apply.(e)no later than 12 months from the date of publication of the notice referred to in Article 34(3) in respect of the electronic system referred to in Article 33(2), points (a) and (b), manufacturers shall ensure that the information to be entered in Eudamed in accordance with Article 29 is entered in that electronic system, including regarding the following devices, provided that those devices are also placed on the market from 6 months from the date of publication of that notice:(i)devices, other than custom-made devices, for which the manufacturer has undertaken a conformity assessment in accordance with Article 52;(ii)devices, other than custom-made devices, placed on the market pursuant to Article 120(3), (3a) or (3b), unless the device, for which the manufacturer has undertaken a conformity assessment in accordance with Article 52, is already registered in Eudamed;(ea)no later than 18 months from the date of publication of the notice referred to in Article 34(3) in respect of the electronic system referred to in Article 33(2), point (d), notified bodies shall ensure that the information to be entered in Eudamed in accordance with Article 56(5) shall apply from 18 months after the later of the dates is entered in that electronic system, including regarding devices referred to in point (d);(f)for (e)(i) of this paragraph; for those devices, only the latest relevant certificate and, where applicable, any subsequent decision taken by the notified body related to such certificate shall be entered;(eb)by way of derogation from point (d), first subparagraph, of this paragraph, the obligations to upload the summary of safety and clinical performance in accordance with Article 32(1) and to notify competent authorities in accordance with Article 55(1), through the electronic system referred in Article 33(2), point (d), shall apply to devices referred to in point (e) of this paragraph when the certificate is entered in Eudamed in accordance with point (ea) of this paragraph;(ec)without prejudice to point (d), first subparagraph, of this paragraph, when a manufacturer is to submit a PSUR in accordance with Article 86(2) of this Regulation, to report a serious incident or a field safety corrective action in accordance with Article 87 of this Regulation, or to submit a trend report in accordance with Article 88 of this Regulation through the electronic system referred to in Article 33(2), point (f), of this Regulation, it shall also register the device, which is the subject of the PSUR or the vigilance reporting, in the electronic system referred to in Article 33(2), points (a) and (b), of this Regulation, except if such device was placed on the market in accordance with Directive 90/385/EEC or 93/42/EEC;(f)for implantable devices and for class III devices Article 27(4) shall apply from 26 May 2021. For class IIa and class IIb devices Article 27(4) shall apply from 26 May 2023. For class I devices Article 27(4) shall apply from 26 May 2025;(g)with regard to reusable devices that are required to bear the UDI carrier on the device itself, Article 27(4) shall apply to:(i)implantable devices and class III devices from 26 May 2023;(ii)class IIa and class IIb devices from 26 May 2025;(iii)class I devices from 26 May 2027;(h)The procedure set out in Article 78 shall apply from 26 May 2027, without prejudice to Article 78(14);(i)Article 120(12) shall apply from 26 May 2019;(j)Article 59 shall apply from 24 April 2020.

The full entry, with the citation mapping v1 = 02017R0745-20230320, v2 = 02017R0745-20240709, is committed at eu/32017R0745/CHANGELOG.md.

02017R0745-2023031102017R0745-20230320

in force 2023-03-20 · detected 2026-08-10

3 provisions touched — 3 substantive, 0 date-only, 0 disputed · 0 sentences quoted verbatim by the gate, 1 changes shipped without an explanation

MODIFIED Art. 120 — Transitional provisions · applies from unknown (the text changed beyond its dates; the applicability binding is prose)

The certificate validity rule in paragraph 2 replaces the single fixed expiry of 27 May 2024 with conditional extended validity tied to new risk-class-based dates set out in a new paragraph 3a, and adds conditions concerning certificates that expired before 20 March 2023, including a written agreement under Annex VII or a competent authority derogation or requirement. Art. 120, v2

Paragraph 3 is restructured and new paragraphs 3a through 3f are inserted, introducing risk-class-specific placing-on-the-market deadlines of 31 December 2027 and 31 December 2028, a 26 May 2026 deadline for certain custom-made implantable devices, and new conditions on quality management systems, formal applications and written agreements with notified bodies by 26 May 2024 and 26 September 2024. Art. 120, v2

Paragraph 4 is changed so that continued availability on the market or putting into service of devices is no longer capped at 26 May 2025 and instead refers to devices lawfully placed on the market under the newly added paragraphs 3, 3a, 3b and 3f. Art. 120, v2 Art. 120, v1

text before / after

02017R0745-2023031102017R0745-20230320

Article 120Transitional provisions1.From 26 May 2021, any publication of a notification in respect of a notified body in accordance with Directives 90/385/EEC and 93/42/EEC shall become void.2.Certificates issued by notified bodies in accordance with Directives 90/385/EEC and 93/42/EEC prior to 25 May 2017 shall remain valid until the end of the period indicated on the certificate, except for certificates issued in accordance with Annex 4 to Directive 90/385/EEC or Annex IV to Directive 93/42/EEC which shall become void at the latest on 27 May 2022.Certificates issued by notified bodies in accordance with Directives 90/385/EEC and 93/42/EEC from 25 May 2017 that were still valid on 26 May 2021 and that have not been withdrawn afterwards shall remain valid until after the end of the period indicated on the certificate until the date set out in paragraph 3a of this Article applicable for the relevant risk class of the devices. Certificates issued by notified bodies in accordance with those Directives from 25 May 2017 that were still valid on 26 May 2021 and that have expired before 20 March 2023 shall be considered to be valid until the dates set out in paragraph 3a of this Article only if one of the following conditions is fulfilled:(a)before the date of expiry of the certificate, which shall not exceed five years the manufacturer and a notified body have signed a written agreement in accordance with Section 4.3, second subparagraph, of Annex VII to this Regulation for the conformity assessment in respect of the device covered by the expired certificate or in respect of a device intended to substitute that device;(b)a competent authority of a Member State has granted a derogation from its issuance. They shall however become void at the latest on 27 May 2024.3.By applicable conformity assessment procedure in accordance with Article 59(1) of this Regulation or has required the manufacturer, in accordance with Article 97(1) of this Regulation, to carry out the applicable conformity assessment procedure.3.By way of derogation from Article 5 and provided the conditions set out in paragraph 3c of this Regulation, Article are met, devices referred to in paragraphs 3a and 3b of this Article may be placed on the market or put into service until the dates set out in those paragraphs.3a.Devices which have a device which certificate that was issued in accordance with Directive 90/385/EEC or Directive 93/42/EEC and that is a valid by virtue of paragraph 2 of this Article may be placed on the market or put into service until the following dates:(a)31 December 2027, for all class III devices, and for class IIb implantable devices except sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors;(b)31 December 2028, for class IIb devices other than those covered by point (a) of this paragraph, for class IIa devices, and for class I device devices placed on the market in sterile condition or having a measuring function.3b.Devices for which the conformity assessment procedure pursuant to Directive 93/42/EEC, 93/42/EEC did not require the involvement of a notified body, for which the declaration of conformity was drawn up prior to 26 May 2021 and for which the conformity assessment procedure pursuant to this Regulation requires the involvement of a notified body, or which has a certificate that was issued in accordance with Directive 90/385/EEC or Directive 93/42/EEC and that is valid by virtue of paragraph 2 of this Article, may be placed on the market or put into service until 26 May 2024, provided that from 26 May 2021 it continues 31 December 2028.3c.Devices referred to in paragraphs 3a and 3b of this Article may be placed on the market or put into service until the dates referred to in those paragraphs only if the following conditions are met:(a)those devices continue to comply with either of those Directives, and provided there Directive 90/385/EEC or Directive 93/42/EEC, as applicable;(b)there are no significant changes in the design and intended purpose. However, purpose;(c)the devices do not present an unacceptable risk to the health or safety of patients, users or other persons, or to other aspects of the protection of public health;(d)no later than 26 May 2024, the manufacturer has put in place a quality management system in accordance with Article 10(9);(e)no later than 26 May 2024, the manufacturer or the authorised representative has lodged a formal application with a notified body in accordance with Section 4.3, first subparagraph, of Annex VII for conformity assessment in respect of a device referred to in paragraph 3a or 3b of this Article or in respect of a device intended to substitute that device, and, no later than 26 September 2024, the notified body and the manufacturer have signed a written agreement in accordance with Section 4.3, second subparagraph, of Annex VII.3d.By way of derogation from paragraph 3 of this Article, the requirements of this Regulation relating to post-market surveillance, market surveillance, vigilance, registration of economic operators and of devices shall apply to devices referred to in paragraphs 3a and 3b of this Article in place of the corresponding requirements in those Directives.Without Directives 90/385/EEC and 93/42/EEC.3e.Without prejudice to Chapter IV and paragraph 1 of this Article, the notified body that issued the certificate referred to in the first subparagraph paragraph 3a of this Article shall continue to be responsible for the appropriate surveillance in respect of all of the applicable requirements relating to the devices it has certified.4.Devices certified, unless the manufacturer has agreed with a notified body designated in accordance with Article 42 that the latter shall carry out such surveillance.No later than 26 September 2024, the notified body that has signed the written agreement referred to in paragraph 3c, point (e), of this Article shall be responsible for the surveillance in respect of the devices covered by the written agreement. Where the written agreement covers a device intended to substitute a device which has a certificate that was issued in accordance with Directive 90/385/EEC or Directive 93/42/EEC, the surveillance shall be conducted in respect of the device that is being substituted.The arrangements for the transfer of the surveillance from the notified body that issued the certificate to the notified body designated in accordance with Article 42 shall be clearly defined in an agreement between the manufacturer and the notified body designated in accordance with Article 42 and, where practicable, the notified body that issued the certificate. The notified body designated in accordance with Article 42 shall not be responsible for conformity assessment activities carried out by the notified body that issued the certificate.3f.By way of derogation from Article 5, class III custom-made implantable devices may be placed on the market or put into service until 26 May 2026 without a certificate issued by a notified body in accordance with the conformity assessment procedure referred to in Article 52(8), second subparagraph, provided that no later than 26 May 2024, the manufacturer or the authorised representative has lodged a formal application with a notified body in accordance with Section 4.3, first subparagraph, of Annex VII for conformity assessment, and, no later than 26 September 2024, the notified body and the manufacturer have signed a written agreement in accordance with Section 4.3, second subparagraph, of Annex VII.4.Devices lawfully placed on the market pursuant to Directives 90/385/EEC and 93/42/EEC prior to 26 May 2021, and devices lawfully placed on the market from 26 May 2021 pursuant to paragraph 3 paragraphs 3, 3a, 3b and 3f of this Article, may continue to be made available on the market or put into service until 26 May 2025.5.By service.5.By way of derogation from Directives 90/385/EEC and 93/42/EEC, devices which comply with this Regulation may be placed on the market prior to 26 May 2021.6.By way of derogation from Directives 90/385/EEC and 93/42/EEC, conformity assessment bodies which comply with this Regulation may be designated and notified prior to 26 May 2021. Notified bodies which are designated and notified in accordance with this Regulation may carry out the conformity assessment procedures laid down in this Regulation and issue certificates in accordance with this Regulation prior to 26 May 2021.7.As regards devices subject to the consultation procedure laid down in Article 54, paragraph 5 of this Article shall apply provided that the necessary appointments to the MDCG and expert panels have been made.8.By way of derogation from Article 10a, point (a) of Article 10b(1) and Article 11(5) of Directive 90/385/EEC and Article 14(1) and (2), points (a) and (b) of Article 14a(1) and Article 16(5) of Directive 93/42/EEC, manufacturers, authorised representatives, importers and notified bodies which, during the period starting on the later of the dates referred to in point (d) of Article 123(3) and ending 18 months later, comply with Articles 29(4), 31(1) and 56(5) of this Regulation shall be considered to comply with the laws and regulations adopted by Member States in accordance with, respectively, Article 10a of Directive 90/385/EEC or Article 14(1) and (2) of Directive 93/42/EEC, with, respectively, point (a) of Article 10b(1) of Directive 90/385/EEC or points (a) and (b) of Article 14a(1) of Directive 93/42/EEC and with, respectively, Article 11(5) of Directive 90/385/EEC or Article 16(5) of Directive 93/42/EEC, as specified in Decision 2010/227/EU.9.Authorisations granted by the competent authorities of the Member States in accordance with Article 9(9) of Directive 90/385/EEC or Article 11(13) of Directive 93/42/EEC shall keep the validity indicated in the authorisation.10.Devices falling within the scope of this Regulation in accordance with point (g) of Article 1(6) which have been legally placed on the market or put into service in accordance with the rules in force in the Member States prior to 26 May 2021 may continue to be placed on the market and put into service in the Member States concerned.11.Clinical investigations which have started to be conducted in accordance with Article 10 of Directive 90/385/EEC or Article 15 of Directive 93/42/EEC prior to 26 May 2021 may continue to be conducted. As of 26 May 2021, however, the reporting of serious adverse events and device deficiencies shall be carried out in accordance with this Regulation.12.Until the Commission has designated, pursuant to Article 27(2), issuing entities, GS1, HIBCC and ICCBBA shall be considered to be designated issuing entities.

MODIFIED Art. 122 — Repeal · applies from unknown (the text changed beyond its dates; the applicability binding is prose)

The cross-references to Article 120 have been broadened from citing paragraphs (3) and (4) to citing paragraphs (3) to (3e) and (4), both in the opening clause and in the paragraph on continued application of the repealed Directives. Art. 122, v2

The paragraph on continued application of the Directives for devices under Article 120 no longer specifies the end date of 27 May 2025 that appeared in the earlier version. Art. 122, v1 Art. 122, v2

That same paragraph now refers to "the first paragraph of this Article" instead of simply "the first paragraph". Art. 122, v2

text before / after

02017R0745-2023031102017R0745-20230320

Article 122RepealWithout prejudice to Articles Article 120(3) to (3e) and (4) of this Regulation, and without prejudice to the obligations of the Member States and manufacturers as regards vigilance and to the obligations of manufacturers as regards the making available of documentation, under Directives 90/385/EEC and 93/42/EEC, those Directives are repealed with effect from 26 May 2021, with the exception of:Articles 8 and 10, points (b) and (c) of Article 10b(1), Article 10b(2) and Article 10b(3) of Directive 90/385/EEC, and the obligations relating to vigilance and clinical investigations provided for in the corresponding Annexes, which are repealed with effect from the later of the dates referred to in point (d) of Article 123(3) of this Regulation;Article 10a, point (a) of Article 10b(1) and Article 11(5) of Directive 90/385/EEC, and the obligations relating to registration of devices and economic operators, and to certificate notifications, provided for in the corresponding Annexes, which are repealed with effect from 18 months after the later of the dates referred to in point (d) of Article 123(3) of this Regulation;Article 10, points (c) and (d) of Article 14a(1), Article 14a(2), Article 14a(3) and Article 15 of Directive 93/42/EEC, and the obligations relating to vigilance and clinical investigations provided for in the corresponding Annexes, which are repealed with effect from the later of the dates referred to in point (d) of Article 123(3) of this Regulation; andArticle 14(1) and (2), points (a) and (b) of Article 14a(1) and Article 16(5) of Directive 93/42/EEC, and the obligations relating to registration of devices and economic operators, and to certificate notifications, provided for in the corresponding Annexes, which are repealed with effect from 18 months after the later of the dates referred to in point (d) of Article 123(3) of this Regulation; andArticle 9(9) of Directive 90/385/EEC and Article 11(13) of Directive 93/42/EEC, which are repealed with effect from 24 April 2020.As regards the devices referred to in Article 120 (3) 120(3) to (3e) and (4) of this Regulation, the Directives referred to in the first paragraph of this Article shall continue to apply until 27 May 2025 to the extent necessary for the application of those paragraphs.Notwithstanding the first paragraph, Regulations (EU) No 207/2012 and (EU) No 722/2012 shall remain in force and continue to apply unless and until repealed by implementing acts adopted by the Commission pursuant to this Regulation.References to the repealed Directives shall be understood as references to this Regulation and shall be read in accordance with the correlation table laid down in Annex XVII to this Regulation.

MODIFIED Art. 123 — Entry into force and date of application · applies from unchanged

No explanation shipped — UnexpectedModelBehavior: Exceeded maximum output retries (1).

text before / after

02017R0745-2023031102017R0745-20230320

Article 123Entry into force and date of application1.This Regulation shall enter into force on the twentieth day following that of its publication in the Official Journal of the European Union.2.It shall apply from 26 May 2021.3.By way of derogation from paragraph 2:(a)Articles 35 to 50 shall apply from 26 November 2017. However, from that date until 26 May 2021, the obligations on notified bodies pursuant to Articles 35 to 50 shall apply only to those bodies which submit an application for designation in accordance with Article 38;(b)Articles 101 and 103 shall apply from 26 November 2017;(c)Article 102 shall apply from 26 May 2018;(d)without prejudice to the obligations on the Commission pursuant to Article 34, where, due to circumstances that could not reasonably have been foreseen when drafting the plan referred to in Article 34(1), Eudamed is not fully functional on 26 May 2021, the obligations and requirements that relate to Eudamed shall apply from the date corresponding to six months after the date of publication of the notice referred to in Article 34(3). The provisions referred to in the preceding sentence are:Article 29,Article 31,Article 32,Article 33(4),the second sentence of Article 40(2),Article 42(10),Article 43(2),the second subparagraph of Article 44(12),points (d) and (e) of Article 46(7),Article 53(2),Article 54(3),Article 55(1),Articles 70 to 77,paragraphs 1 to 13 of Article 78,Articles 79 to 82,Article 86(2),Articles 87 and 88,Article 89(5) and (7), and the third subparagraph of Article 89(8),Article 90,Article 93(4), (7) and (8),Article 95(2) and (4),the last sentence of Article 97(2),Article 99(4),the second sentence of the first subparagraph of Article 120(3).Until 99(4),Article 120(3d).Until Eudamed is fully functional, the corresponding provisions of Directives 90/385/EEC and 93/42/EEC shall continue to apply for the purpose of meeting the obligations laid down in the provisions listed in the first paragraph of this point regarding exchange of information including, and in particular, information regarding vigilance reporting, clinical investigations, registration of devices and economic operators, and certificate notifications.(e)Article 29(4) and Article 56(5) shall apply from 18 months after the later of the dates referred to in point (d);(f)for implantable devices and for class III devices Article 27(4) shall apply from 26 May 2021. For class IIa and class IIb devices Article 27(4) shall apply from 26 May 2023. For class I devices Article 27(4) shall apply from 26 May 2025;(g)with regard to reusable devices that are required to bear the UDI carrier on the device itself, Article 27(4) shall apply to:(i)implantable devices and class III devices from 26 May 2023;(ii)class IIa and class IIb devices from 26 May 2025;(iii)class I devices from 26 May 2027;(h)The procedure set out in Article 78 shall apply from 26 May 2027, without prejudice to Article 78(14);(i)Article 120(12) shall apply from 26 May 2019;(j)Article 59 shall apply from 24 April 2020.

The full entry, with the citation mapping v1 = 02017R0745-20230311, v2 = 02017R0745-20230320, is committed at eu/32017R0745/CHANGELOG.md.

02017R0745-2020042402017R0745-20230311

in force 2023-03-11 · detected 2026-08-10

2 provisions touched — 2 substantive, 0 date-only, 2 disputed · 0 sentences quoted verbatim by the gate, 1 changes shipped without an explanation

MODIFIED Art. 1 · applies from unknown

Disputed — seen by the instruction parse, not by the structural diff, corpus metadata.

No explanation shipped — the structural diff did not see this change, so it carries no text; another signal named the unit and the disagreement ships as `disputed`.

text before / after

No text on either side: this unit was named by a signal that carries no text, and only the structural diff carries any.

MODIFIED Art. 44 — Monitoring and re-assessment of notified bodies · applies from unknown (the text changed beyond its dates; the applicability binding is prose)

Disputed — seen by the structural diff, corpus metadata, not by the instruction parse.

The interval for the complete re-assessment of a notified body changes from three years after notification and every fourth year thereafter to five years after notification and every fifth year thereafter, and the reference to the procedure moves from Articles 38 and 39 to Article 39 alone. Art. 44, v1 Art. 44, v2

The revised text adds that the authority responsible for notified bodies may conduct a complete re-assessment earlier than the scheduled dates, either at the notified body's request or based on concerns arising from the annual assessments described in paragraph 4. Art. 44, v2

A new passage also addresses complete re-assessments already underway before 11 March 2023, stating they continue unless the authority decides to suspend or terminate them after weighing resources spent and annual assessment results, and after hearing the notified body concerned. Art. 44, v2

text before / after

02017R0745-2020042402017R0745-20230311

Article 44Monitoring and re-assessment of notified bodies1.Notified bodies shall, without delay, and at the latest within 15 days, inform the authority responsible for notified bodies of relevant changes which may affect their compliance with the requirements set out in Annex VII or their ability to conduct the conformity assessment activities relating to the devices for which they have been designated.2.The authorities responsible for notified bodies shall monitor the notified bodies established on their territory and their subsidiaries and subcontractors to ensure ongoing compliance with the requirements and the fulfilment of its obligations set out in this Regulation. Notified bodies shall, upon request by their authority responsible for notified bodies, supply all relevant information and documents, required to enable the authority, the Commission and other Member States to verify compliance.3.Where the Commission or the authority of a Member State submits a request to a notified body established on the territory of another Member State relating to a conformity assessment carried out by that notified body, it shall send a copy of that request to the authority responsible for notified bodies of that other Member State. The notified body concerned shall respond without delay and within 15 days at the latest to the request. The authority responsible for notified bodies of the Member State in which the body is established shall ensure that requests submitted by authorities of any other Member State or by the Commission are resolved by the notified body unless there is a legitimate reason for not doing so in which case the matter may be referred to the MDCG.4.At least once a year, the authorities responsible for notified bodies shall re-assess whether the notified bodies established on their respective territory and, where appropriate, the subsidiaries and subcontractors under the responsibility of those notified bodies still satisfy the requirements and fulfil their obligations set out in Annex VII. That review shall include an on-site audit of each notified body and, where necessary, of its subsidiaries and subcontractors.The authority responsible for notified bodies shall conduct its monitoring and assessment activities according to an annual assessment plan to ensure that it can effectively monitor the continued compliance of the notified body with the requirements of this Regulation. That plan shall provide a reasoned schedule for the frequency of assessment of the notified body and, in particular, associated subsidiaries and subcontractors. The authority shall submit its annual plan for monitoring or assessment for each notified body for which it is responsible to the MDCG and to the Commission.5.The monitoring of notified bodies by the authority responsible for notified bodies shall include observed audits of notified body personnel, including where necessary any personnel from subsidiaries and subcontractors, as that personnel is in the process of conducting quality management system assessments at a manufacturer's facility.6.The monitoring of notified bodies conducted by the authority responsible for notified bodies shall consider data arising from market surveillance, vigilance and post-market surveillance to help guide its activities.The authority responsible for notified bodies shall provide for a systematic follow-up of complaints and other information, including from other Member States, which may indicate non-fulfilment of the obligations by a notified body or its deviation from common or best practice.7.The authority responsible for notified bodies may in addition to regular monitoring or on-site assessments conduct short-notice, unannounced or for-cause reviews if needed to address a particular issue or to verify compliance.8.The authority responsible for notified bodies shall review the assessments by notified bodies of manufacturers' technical documentation, in particular the clinical evaluation documentation as further outlined in Article 45.9.The authority responsible for notified bodies shall document and record any findings regarding non-compliance of the notified body with the requirements set out in Annex VII and shall monitor the timely implementation of corrective and preventive actions.10.Three actions.10.5 years after notification of a notified body, and again every fourth year 5 years thereafter, a complete re-assessment to determine whether the notified body still satisfies the requirements set out in Annex VII shall be conducted by the authority responsible for notified bodies of the Member State in which the body is established and by a joint assessment team appointed for the purpose of in accordance with the procedure described in Articles 38 Article 39.The authority responsible for notified bodies of the Member State in which the notified body is established may conduct a complete re-assessment prior to the dates referred to in the first subparagraph, upon request by the notified body or where, based on the results of the annual assessments conducted in accordance with paragraph 4 of this Article, it has concerns regarding the continued fulfilment by the notified body of the requirements set out in Annex VII.Complete re-assessments that have already started prior to 11 March 2023 shall continue to be conducted, unless the authority responsible for notified bodies of the Member State in which the notified body is established decides to suspend or terminate the ongoing complete re-assessment, taking into account its own resources and 39.11.The the resources of the notified body already spent on the re-assessment, as well as the results of the annual assessments conducted in accordance with paragraph 4 of this Article. Before suspending or terminating an ongoing complete re-assessment, the authority responsible for notified bodies shall hear the notified body concerned.11.The Commission is empowered to adopt delegated acts in accordance with Article 115 in order to amend paragraph 10 to modify the frequency at which the complete re-assessment referred to in that paragraph is to be carried out.12.The Member States shall report to the Commission and to the MDCG, at least once a year, on their monitoring and on-site assessment activities regarding notified bodies and, where applicable, subsidiaries and subcontractors. The report shall provide details of the outcome of those activities, including activities pursuant to paragraph 7, and shall be treated as confidential by the MDCG and the Commission; however it shall contain a summary which shall be made publicly available.The summary of the report shall be uploaded to the electronic system referred to in Article 57.

The full entry, with the citation mapping v1 = 02017R0745-20200424, v2 = 02017R0745-20230311, is committed at eu/32017R0745/CHANGELOG.md.

02017R0745-2017050502017R0745-20200424

in force 2020-04-24 · detected 2026-08-10

9 provisions touched — 3 substantive, 6 date-only, 0 disputed · 0 sentences quoted verbatim by the gate, 0 changes shipped without an explanation

DEFERRED Art. 1 — Subject matter and scope · applies from 2021-05-26

In paragraph 2, the deadline by which the necessary common specifications must be adopted is changed from 26 May 2020 to 26 May 2021. Art. 1, v2

The related reference date used to determine when those common specifications apply is likewise changed from 26 May 2020 to 26 May 2021. Art. 1, v2

text before / after

02017R0745-2017050502017R0745-20200424

Article 1Subject matter and scope1.This Regulation lays down rules concerning the placing on the market, making available on the market or putting into service of medical devices for human use and accessories for such devices in the Union. This Regulation also applies to clinical investigations concerning such medical devices and accessories conducted in the Union.2.This Regulation shall also apply, as from the date of application of common specifications adopted pursuant to Article 9, to the groups of products without an intended medical purpose that are listed in Annex XVI, taking into account the state of the art, and in particular existing harmonised standards for analogous devices with a medical purpose, based on similar technology. The common specifications for each of the groups of products listed in Annex XVI shall address, at least, application of risk management as set out in Annex I for the group of products in question and, where necessary, clinical evaluation regarding safety.The necessary common specifications shall be adopted by 26 May 2020. 2021. They shall apply as from six months after the date of their entry into force or from 26 May 2020, 2021, whichever is the latest.Notwithstanding Article 122, Member States' measures regarding the qualification of the products covered by Annex XVI as medical devices pursuant to Directive 93/42/EEC shall remain valid until the date of application, as referred to in the first subparagraph, of the relevant common specifications for that group of products.This Regulation also applies to clinical investigations conducted in the Union concerning the products referred to in the first subparagraph.3.Devices with both a medical and a non-medical intended purpose shall fulfil cumulatively the requirements applicable to devices with an intended medical purpose and those applicable to devices without an intended medical purpose.4.For the purposes of this Regulation, medical devices, accessories for medical devices, and products listed in Annex XVI to which this Regulation applies pursuant to paragraph 2 shall hereinafter be referred to as devices.5.Where justified on account of the similarity between a device with an intended medical purpose placed on the market and a product without an intended medical purpose in respect of their characteristics and risks, the Commission is empowered to adopt delegated acts in accordance with Article 115 to amend the list in Annex XVI, by adding new groups of products, in order to protect the health and safety of users or other persons or other aspects of public health.6.This Regulation does not apply to:(a)in vitro diagnostic medical devices covered by Regulation (EU) 2017/746;(b)medicinal products as defined in point 2 of Article 1 of Directive 2001/83/EC. In deciding whether a product falls under Directive 2001/83/EC or under this Regulation, particular account shall be taken of the principal mode of action of the product;(c)advanced therapy medicinal products covered by Regulation (EC) No 1394/2007;(d)human blood, blood products, plasma or blood cells of human origin or devices which incorporate, when placed on the market or put into service, such blood products, plasma or cells, except for devices referred to in paragraph 8 of this Article;(e)cosmetic products covered by Regulation (EC) No 1223/2009;(f)transplants, tissues or cells of animal origin, or their derivatives, or products containing or consisting of them; however this Regulation does apply to devices manufactured utilising tissues or cells of animal origin, or their derivatives, which are non-viable or are rendered non-viable;(g)transplants, tissues or cells of human origin, or their derivatives, covered by Directive 2004/23/EC, or products containing or consisting of them; however this Regulation does apply to devices manufactured utilising derivatives of tissues or cells of human origin which are non-viable or are rendered non-viable;(h)products, other than those referred to in points (d), (f) and (g), that contain or consist of viable biological material or viable organisms, including living micro-organisms, bacteria, fungi or viruses in order to achieve or support the intended purpose of the product;(i)food covered by Regulation (EC) No 178/2002.7.Any device which, when placed on the market or put into service, incorporates as an integral part an in vitro diagnostic medical device as defined in point 2 of Article 2 of Regulation (EU) 2017/746, shall be governed by this Regulation. The requirements of Regulation (EU) 2017/746 shall apply to the in vitro diagnostic medical device part of the device.8.Any device which, when placed on the market or put into service, incorporates, as an integral part, a substance which, if used separately, would be considered to be a medicinal product as defined in point 2 of Article 1 of Directive 2001/83/EC, including a medicinal product derived from human blood or human plasma as defined in point 10 of Article 1 of that Directive, and that has an action ancillary to that of the device, shall be assessed and authorised in accordance with this Regulation.However, if the action of that substance is principal and not ancillary to that of the device, the integral product shall be governed by Directive 2001/83/EC or Regulation (EC) No 726/2004 of the European Parliament and of the CouncilRegulation (EC) No 726/2004 of the European Parliament and of the Council of 31 March 2004 laying down Community procedures for the authorisation and supervision of medicinal products for human and veterinary use and establishing a European Medicines Agency (OJ L 136, 30.4.2004, p. 1)., as applicable. In that case, the relevant general safety and performance requirements set out in Annex I to this Regulation shall apply as far as the safety and performance of the device part are concerned.9.Any device which is intended to administer a medicinal product as defined in point 2 of Article 1 of Directive 2001/83/EC shall be governed by this Regulation, without prejudice to the provisions of that Directive and of Regulation (EC) No 726/2004 with regard to the medicinal product.However, if the device intended to administer a medicinal product and the medicinal product are placed on the market in such a way that they form a single integral product which is intended exclusively for use in the given combination and which is not reusable, that single integral product shall be governed by Directive 2001/83/EC or Regulation (EC) No 726/2004, as applicable. In that case, the relevant general safety and performance requirements set out in Annex I to this Regulation shall apply as far as the safety and performance of the device part of the single integral product are concerned.10.Any device which, when placed on the market or put into service, incorporates, as an integral part, non-viable tissues or cells of human origin or their derivatives that have an action ancillary to that of the device shall be assessed and authorised in accordance with this Regulation. In that case, the provisions for donation, procurement and testing laid down in Directive 2004/23/EC shall apply.However, if the action of those tissues or cells or their derivatives is principal and not ancillary to that of the device and the product is not governed by Regulation (EC) No 1394/2007, the product shall be governed by Directive 2004/23/EC. In that case, the relevant general safety and performance requirements set out in Annex I to this Regulation shall apply as far as the safety and performance of the device part are concerned.11.This Regulation is specific Union legislation within the meaning of Article 2(3) of Directive 2014/30/EU.12.Devices that are also machinery within the meaning of point (a) of the second paragraph of Article 2 of Directive 2006/42/EC of the European Parliament and of the CouncilDirective 2006/42/EC of the European Parliament and of the Council of 17 May 2006 on machinery, and amending Directive 95/16/EC (OJ L 157, 9.6.2006, p. 24). shall, where a hazard relevant under that Directive exists, also meet the essential health and safety requirements set out in Annex I to that Directive to the extent to which those requirements are more specific than the general safety and performance requirements set out in Chapter II of Annex I to this Regulation.13.This Regulation shall not affect the application of Directive 2013/59/Euratom.14.This Regulation shall not affect the right of a Member State to restrict the use of any specific type of device in relation to aspects not covered by this Regulation.15.This Regulation shall not affect national law concerning the organisation, delivery or financing of health services and medical care, such as the requirement that certain devices may only be supplied on a medical prescription, the requirement that only certain health professionals or healthcare institutions may dispense or use certain devices or that their use be accompanied by specific professional counselling.16.Nothing in this Regulation shall restrict the freedom of the press or the freedom of expression in the media in so far as those freedoms are guaranteed in the Union and in the Member States, in particular under Article 11 of the Charter of Fundamental Rights of the European Union.

DEFERRED Art. 17 — Single-use devices and their reprocessing · applies from 2021-05-26

The two dates of 26 May 2020 in paragraph 5, marking the deadline for adopting common specifications and the fallback trigger date, are changed to 26 May 2021. Art. 17, v2

The date of 26 May 2020 in paragraph 6, which marks the cutoff for devices placed on the market under Directive 93/42/EEC that may still be reprocessed, is likewise changed to 26 May 2021. Art. 17, v2

text before / after

02017R0745-2017050502017R0745-20200424

Article 17Single-use devices and their reprocessing1.Reprocessing and further use of single-use devices may only take place where permitted by national law and only in accordance with this Article.2.Any natural or legal person who reprocesses a single-use device to make it suitable for further use within the Union shall be considered to be the manufacturer of the reprocessed device and shall assume the obligations incumbent on manufacturers laid down in this Regulation, which include obligations relating to the traceability of the reprocessed device in accordance with Chapter III of this Regulation. The reprocessor of the device shall be considered to be a producer for the purpose of Article 3(1) of Directive 85/374/EEC.3.By way of derogation from paragraph 2, as regards single-use devices that are reprocessed and used within a health institution, Member States may decide not to apply all of the rules relating to manufacturers' obligations laid down in this Regulation provided that they ensure that:(a)the safety and performance of the reprocessed device is equivalent to that of the original device and the requirements in points (a), (b), (d), (e), (f), (g) and (h) of Article 5(5) are complied with;(b)the reprocessing is performed in accordance with CS detailing the requirements concerning:risk management, including the analysis of the construction and material, related properties of the device (reverse engineering) and procedures to detect changes in the design of the original device as well as of its planned application after reprocessing,the validation of procedures for the entire process, including cleaning steps,the product release and performance testing,the quality management system,the reporting of incidents involving devices that have been reprocessed, andthe traceability of reprocessed devices.Member States shall encourage, and may require, health institutions to provide information to patients on the use of reprocessed devices within the health institution and, where appropriate, any other relevant information on the reprocessed devices that patients are treated with.Member States shall notify the Commission and the other Member States of the national provisions introduced pursuant to this paragraph and the grounds for introducing them. The Commission shall keep the information publicly available.4.Member States may choose to apply the provisions referred to in paragraph 3 also as regards single-use devices that are reprocessed by an external reprocessor at the request of a health institution, provided that the reprocessed device in its entirety is returned to that health institution and the external reprocessor complies with the requirements referred to in points (a) and (b) of paragraph 3.5.The Commission shall adopt, in accordance with Article 9(1), the necessary CS referred to in point (b) of paragraph 3 by 26 May 2020. 2021. Those CS shall be consistent with the latest scientific evidence and shall address the application of the general requirements on safety and performance laid down in in this Regulation. In the event that those CS are not adopted by 26 May 2020, 2021, reprocessing shall be performed in accordance with any relevant harmonised standards and national provisions that cover the aspects outlined in point (b) of paragraph 3. Compliance with CS or, in the absence of CS, with any relevant harmonised standards and national provisions, shall be certified by a notified body.6.Only single-use devices that have been placed on the market in accordance with this Regulation, or prior to 26 May 2020 2021 in accordance with Directive 93/42/EEC, may be reprocessed.7.Only reprocessing of single-use devices that is considered safe according to the latest scientific evidence may be carried out.8.The name and address of the legal or natural person referred to in paragraph 2 and the other relevant information referred to in Section 23 of Annex I shall be indicated on the label and, where applicable, in the instructions for use of the reprocessed device.The name and address of the manufacturer of the original single-use device shall no longer appear on the label, but shall be mentioned in the instructions for use of the reprocessed device.9.A Member State that permits reprocessing of single-use devices may maintain or introduce national provisions that are stricter than those laid down in this Regulation and which restrict or prohibit, within its territory, the following:(a)the reprocessing of single-use devices and the transfer of single-use devices to another Member State or to a third country with a view to their reprocessing;(b)the making available or further use of reprocessed single-use devices.Member States shall notify the Commission and the other Member States of those national provisions. The Commission shall make such information publicly available.10.The Commission shall by 27 May 2024 draw up a report on the operation of this Article and submit it to the European Parliament and to the Council. On the basis of that report, the Commission shall, if appropriate, make proposals for amendments to this Regulation.

DEFERRED Art. 34 — Functionality of Eudamed · applies from 2021-03-25

The date by which the Commission's implementation plan must allow publication of the notice referred to in paragraph 3 is changed from 25 March 2020 to 25 March 2021. Art. 34, v1 Art. 34, v2

text before / after

02017R0745-2017050502017R0745-20200424

Article 34Functionality of Eudamed1.The Commission shall, in collaboration with the MDCG, draw up the functional specifications for Eudamed. The Commission shall draw up a plan for the implementation of those specifications by 26 May 2018. That plan shall seek to ensure that Eudamed is fully functional at a date that allows the Commission to publish the notice referred to in paragraph 3 of this Article by 25 March 2020 2021 and that all other relevant deadlines laid down in Article 123 of this Regulation and in Article 113 of Regulation (EU) 2017/746 are met.2.The Commission shall, on the basis of an independent audit report, inform the MDCG when it has verified that Eudamed has achieved full functionality and Eudamed meets the functional specifications drawn up pursuant to paragraph 1.3.The Commission shall, after consultation with the MDCG and when it is satisfied that the conditions referred to in paragraph 2 have been fulfilled, publish a notice to that effect in the Official Journal of the European Union.

MODIFIED Art. 59 — Derogation from the conformity assessment procedures · applies from unknown (the text changed beyond its dates; the applicability binding is prose)

Paragraph 1 now adds a derogation for the period from 24 April 2020 to 25 May 2021 from Article 9(1) and (2) of Directive 90/385/EEC or Article 11(1) to (6) of Directive 93/42/EEC, alongside the existing derogation from Article 52, and refers to procedures under 'those Articles' rather than 'that Article'. Art. 59, v2

Paragraph 2 adds a sentence allowing the Member State to inform the Commission and other Member States of authorisations granted under Article 9(9) of Directive 90/385/EEC or Article 11(13) of Directive 93/42/EEC before 24 April 2020. Art. 59, v2

Paragraph 3 now also allows extension of the validity of authorisations granted before 24 April 2020 under Article 9(9) of Directive 90/385/EEC or Article 11(13) of Directive 93/42/EEC, in addition to authorisations granted under paragraph 1. Art. 59, v2

text before / after

02017R0745-2017050502017R0745-20200424

Article 59Derogation from the conformity assessment procedures1.By way of derogation from Article 52, 52 of this Regulation or, for the period from 24 April 2020 to 25 May 2021, by way of derogation from Article 9(1) and (2) of Directive 90/385/EEC or from Article 11(1) to (6) of Directive 93/42/EEC, any competent authority may authorise, on a duly justified request, the placing on the market or putting into service within the territory of the Member State concerned, of a specific device for which the applicable procedures referred to in that Article those Articles have not been carried out but use of which is in the interest of public health or patient safety or health.2.The Member State shall inform the Commission and the other Member States of any decision to authorise the placing on the market or putting into service of a device in accordance with paragraph 1 where such authorisation is granted for use other than for a single patient.3.Following patient.The Member State may inform the Commission and the other Member States of any authorisation granted in accordance with Article 9(9) of Directive 90/385/EEC or Article 11(13) of Directive 93/42/EEC before 24 April 2020.3.Following a notification pursuant to paragraph 2 of this Article, the Commission, in exceptional cases relating to public health or patient safety or health, may, by means of implementing acts, extend for a limited period of time the validity of an authorisation granted by a Member State in accordance with paragraph 1 of this Article or, when granted before 24 April 2020, in accordance with Article 9(9) of Directive 90/385/EEC or Article 11(13) of Directive 93/42/EEC to the territory of the Union and set the conditions under which the device may be placed on the market or put into service. Those implementing acts shall be adopted in accordance with the examination procedure referred to in Article 114(3).On duly justified imperative grounds of urgency relating to the health and safety of humans, the Commission shall adopt immediately applicable implementing acts in accordance with the procedure referred to in Article 114(4).

DEFERRED Art. 113 — Penalties · applies from 2021-02-25

The date by which Member States must notify the Commission of penalty rules and measures was changed from 25 February 2020 to 25 February 2021. Art. 113, v1 Art. 113, v2

text before / after

02017R0745-2017050502017R0745-20200424

Article 113PenaltiesThe Member States shall lay down the rules on penalties applicable for infringement of the provisions of this Regulation and shall take all measures necessary to ensure that they are implemented. The penalties provided for shall be effective, proportionate, and dissuasive. The Member States shall notify the Commission of those rules and of those measures by 25 February 2020 2021 and shall notify it, without delay, of any subsequent amendment affecting them.

DEFERRED Art. 120 — Transitional provisions · applies from 2021-05-26

Every reference to 26 May 2020 in paragraphs 1, 3, 4, 5, 6, 10 and 11 has been replaced with 26 May 2021, while the other dates in the article remain unchanged. Art. 120, v2 Art. 120, v1

text before / after

02017R0745-2017050502017R0745-20200424

Article 120Transitional provisions1.From 26 May 2020, 2021, any publication of a notification in respect of a notified body in accordance with Directives 90/385/EEC and 93/42/EEC shall become void.2.Certificates issued by notified bodies in accordance with Directives 90/385/EEC and 93/42/EEC prior to 25 May 2017 shall remain valid until the end of the period indicated on the certificate, except for certificates issued in accordance with Annex 4 to Directive 90/385/EEC or Annex IV to Directive 93/42/EEC which shall become void at the latest on 27 May 2022.Certificates issued by notified bodies in accordance with Directives 90/385/EEC and 93/42/EEC from 25 May 2017 shall remain valid until the end of the period indicated on the certificate, which shall not exceed five years from its issuance. They shall however become void at the latest on 27 May 2024.3.By way of derogation from Article 5 of this Regulation, a device which is a class I device pursuant to Directive 93/42/EEC, for which the declaration of conformity was drawn up prior to 26 May 2020 2021 and for which the conformity assessment procedure pursuant to this Regulation requires the involvement of a notified body, or which has a certificate that was issued in accordance with Directive 90/385/EEC or Directive 93/42/EEC and that is valid by virtue of paragraph 2 of this Article, may be placed on the market or put into service until 26 May 2024, provided that from 26 May 2020 2021 it continues to comply with either of those Directives, and provided there are no significant changes in the design and intended purpose. However, the requirements of this Regulation relating to post-market surveillance, market surveillance, vigilance, registration of economic operators and of devices shall apply in place of the corresponding requirements in those Directives.Without prejudice to Chapter IV and paragraph 1 of this Article, the notified body that issued the certificate referred to in the first subparagraph shall continue to be responsible for the appropriate surveillance in respect of all of the applicable requirements relating to the devices it has certified.4.Devices lawfully placed on the market pursuant to Directives 90/385/EEC and 93/42/EEC prior to 26 May 2020, 2021, and devices placed on the market from 26 May 2020 2021 pursuant to paragraph 3 of this Article, may continue to be made available on the market or put into service until 26 May 2025.5.By way of derogation from Directives 90/385/EEC and 93/42/EEC, devices which comply with this Regulation may be placed on the market prior to 26 May 2020.6.By 2021.6.By way of derogation from Directives 90/385/EEC and 93/42/EEC, conformity assessment bodies which comply with this Regulation may be designated and notified prior to 26 May 2020. 2021. Notified bodies which are designated and notified in accordance with this Regulation may carry out the conformity assessment procedures laid down in this Regulation and issue certificates in accordance with this Regulation prior to 26 May 2020.7.As 2021.7.As regards devices subject to the consultation procedure laid down in Article 54, paragraph 5 of this Article shall apply provided that the necessary appointments to the MDCG and expert panels have been made.8.By way of derogation from Article 10a, point (a) of Article 10b(1) and Article 11(5) of Directive 90/385/EEC and Article 14(1) and (2), points (a) and (b) of Article 14a(1) and Article 16(5) of Directive 93/42/EEC, manufacturers, authorised representatives, importers and notified bodies which, during the period starting on the later of the dates referred to in point (d) of Article 123(3) and ending 18 months later, comply with Articles 29(4), 31(1) and 56(5) of this Regulation shall be considered to comply with the laws and regulations adopted by Member States in accordance with, respectively, Article 10a of Directive 90/385/EEC or Article 14(1) and (2) of Directive 93/42/EEC, with, respectively, point (a) of Article 10b(1) of Directive 90/385/EEC or points (a) and (b) of Article 14a(1) of Directive 93/42/EEC and with, respectively, Article 11(5) of Directive 90/385/EEC or Article 16(5) of Directive 93/42/EEC, as specified in Decision 2010/227/EU.9.Authorisations granted by the competent authorities of the Member States in accordance with Article 9(9) of Directive 90/385/EEC or Article 11(13) of Directive 93/42/EEC shall keep the validity indicated in the authorisation.10.Devices falling within the scope of this Regulation in accordance with point (g) of Article 1(6) which have been legally placed on the market or put into service in accordance with the rules in force in the Member States prior to 26 May 2020 2021 may continue to be placed on the market and put into service in the Member States concerned.11.Clinical investigations which have started to be conducted in accordance with Article 10 of Directive 90/385/EEC or Article 15 of Directive 93/42/EEC prior to 26 May 2020 2021 may continue to be conducted. As of 26 May 2020, 2021, however, the reporting of serious adverse events and device deficiencies shall be carried out in accordance with this Regulation.12.Until the Commission has designated, pursuant to Article 27(2), issuing entities, GS1, HIBCC and ICCBBA shall be considered to be designated issuing entities.

MODIFIED Art. 122 — Repeal · applies from unknown (the text changed beyond its dates; the applicability binding is prose)

The general repeal date for Directives 90/385/EEC and 93/42/EEC was changed from 26 May 2020 to 26 May 2021. Art. 122, v1 Art. 122, v2

A new exception was added covering Article 9(9) of Directive 90/385/EEC and Article 11(13) of Directive 93/42/EEC, which are stated to be repealed with effect from 24 April 2020. Art. 122, v2

text before / after

02017R0745-2017050502017R0745-20200424

Article 122RepealWithout prejudice to Articles 120(3) and (4) of this Regulation, and without prejudice to the obligations of the Member States and manufacturers as regards vigilance and to the obligations of manufacturers as regards the making available of documentation, under Directives 90/385/EEC and 93/42/EEC, those Directives are repealed with effect from 26 May 2020, 2021, with the exception of:Articles 8 and 10, points (b) and (c) of Article 10b(1), Article 10b(2) and Article 10b(3) of Directive 90/385/EEC, and the obligations relating to vigilance and clinical investigations provided for in the corresponding Annexes, which are repealed with effect from the later of the dates referred to in point (d) of Article 123(3) of this Regulation;Article 10a, point (a) of Article 10b(1) and Article 11(5) of Directive 90/385/EEC, and the obligations relating to registration of devices and economic operators, and to certificate notifications, provided for in the corresponding Annexes, which are repealed with effect from 18 months after the later of the dates referred to in point (d) of Article 123(3) of this Regulation;Article 10, points (c) and (d) of Article 14a(1), Article 14a(2), Article 14a(3) and Article 15 of Directive 93/42/EEC, and the obligations relating to vigilance and clinical investigations provided for in the corresponding Annexes, which are repealed with effect from the later of the dates referred to in point (d) of Article 123(3) of this Regulation; andArticle 14(1) and (2), points (a) and (b) of Article 14a(1) and Article 16(5) of Directive 93/42/EEC, and the obligations relating to registration of devices and economic operators, and to certificate notifications, provided for in the corresponding Annexes, which are repealed with effect from 18 months after the later of the dates referred to in point (d) of Article 123(3) of this Regulation.As Regulation; andArticle 9(9) of Directive 90/385/EEC and Article 11(13) of Directive 93/42/EEC, which are repealed with effect from 24 April 2020.As regards the devices referred to in Article 120 (3) and (4) of this Regulation, the Directives referred to in the first paragraph shall continue to apply until 27 May 2025 to the extent necessary for the application of those paragraphs.Notwithstanding the first paragraph, Regulations (EU) No 207/2012 and (EU) No 722/2012 shall remain in force and continue to apply unless and until repealed by implementing acts adopted by the Commission pursuant to this Regulation.References to the repealed Directives shall be understood as references to this Regulation and shall be read in accordance with the correlation table laid down in Annex XVII to this Regulation.

MODIFIED Art. 123 — Entry into force and date of application · applies from unknown (the text changed beyond its dates; the applicability binding is prose)

The general date of application in paragraph 2 and the corresponding date in point (a) and (d) were moved from 26 May 2020 to 26 May 2021. Art. 123, v1 Art. 123, v2

Point (g) was rewritten from a single rule tying the UDI-carrier deadline for reusable devices to two years after the point (f) dates, into three separate sub-points setting distinct dates of 26 May 2023, 26 May 2025 and 26 May 2027 for implantable/class III, class IIa/IIb, and class I devices respectively. Art. 123, v1 Art. 123, v2

A new point (j) was added stating that Article 59 shall apply from 24 April 2020. Art. 123, v2

text before / after

02017R0745-2017050502017R0745-20200424

Article 123Entry into force and date of application1.This Regulation shall enter into force on the twentieth day following that of its publication in the Official Journal of the European Union.2.It shall apply from 26 May 2020.3.By 2021.3.By way of derogation from paragraph 2:(a)Articles 35 to 50 shall apply from 26 November 2017. However, from that date until 26 May 2020, 2021, the obligations on notified bodies pursuant to Articles 35 to 50 shall apply only to those bodies which submit an application for designation in accordance with Article 38;(b)Articles 101 and 103 shall apply from 26 November 2017;(c)Article 102 shall apply from 26 May 2018;(d)without prejudice to the obligations on the Commission pursuant to Article 34, where, due to circumstances that could not reasonably have been foreseen when drafting the plan referred to in Article 34(1), Eudamed is not fully functional on 26 May 2020, 2021, the obligations and requirements that relate to Eudamed shall apply from the date corresponding to six months after the date of publication of the notice referred to in Article 34(3). The provisions referred to in the preceding sentence are:Article 29,Article 31,Article 32,Article 33(4),the second sentence of Article 40(2),Article 42(10),Article 43(2),the second subparagraph of Article 44(12),points (d) and (e) of Article 46(7),Article 53(2),Article 54(3),Article 55(1),Articles 70 to 77,paragraphs 1 to 13 of Article 78,Articles 79 to 82,Article 86(2),Articles 87 and 88,Article 89(5) and (7), and the third subparagraph of Article 89(8),Article 90,Article 93(4), (7) and (8),Article 95(2) and (4),the last sentence of Article 97(2),Article 99(4),the second sentence of the first subparagraph of Article 120(3).Until Eudamed is fully functional, the corresponding provisions of Directives 90/385/EEC and 93/42/EEC shall continue to apply for the purpose of meeting the obligations laid down in the provisions listed in the first paragraph of this point regarding exchange of information including, and in particular, information regarding vigilance reporting, clinical investigations, registration of devices and economic operators, and certificate notifications.(e)Article 29(4) and Article 56(5) shall apply from 18 months after the later of the dates referred to in point (d);(f)for implantable devices and for class III devices Article 27(4) shall apply from 26 May 2021. For class IIa and class IIb devices Article 27(4) shall apply from 26 May 2023. For class I devices Article 27(4) shall apply from 26 May 2025;(g)for 2025;(g)with regard to reusable devices that shall are required to bear the UDI carrier on the device itself, Article 27(4) shall apply to:(i)implantable devices and class III devices from two years after the date referred to in point (f) of this paragraph for the respective 26 May 2023;(ii)class IIa and class of IIb devices in that point;(h)The from 26 May 2025;(iii)class I devices from 26 May 2027;(h)The procedure set out in Article 78 shall apply from 26 May 2027, without prejudice to Article 78(14);(i)Article 120(12) shall apply from 26 May 2019. 2019;(j)Article 59 shall apply from 24 April 2020.

DEFERRED Annex IX — ANNEX IX · applies from 2021-05-26

In section 5.1(h), the deadline by which the Commission must provide guidance to expert panels on interpreting the criteria in point (c) was changed from 26 May 2020 to 26 May 2021. Annex IX, v2

text before / after

02017R0745-2017050502017R0745-20200424

ANNEX IXCONFORMITY ASSESSMENT BASED ON A QUALITY MANAGEMENT SYSTEM AND ON ASSESSMENT OF TECHNICAL DOCUMENTATIONCHAPTER IQUALITY MANAGEMENT SYSTEM1.The manufacturer shall establish, document and implement a quality management system as described in Article 10(9) and maintain its effectiveness throughout the life cycle of the devices concerned. The manufacturer shall ensure the application of the quality management system as specified in Section 2 and shall be subject to audit, as laid down in Sections 2.3 and 2.4, and to surveillance as specified in Section 3.2.Quality management system assessment2.1.The manufacturer shall lodge an application for assessment of its quality management system with a notified body. The application shall include:the name of the manufacturer and address of its registered place of business and any additional manufacturing site covered by the quality management system, and, if the manufacturer's application is lodged by its authorised representative, the name of the authorised representative and the address of the authorised representative's registered place of business,all relevant information on the device or group of devices covered by the quality management system,a written declaration that no application has been lodged with any other notified body for the same device-related quality management system, or information about any previous application for the same device-related quality management system,a draft of an EU declaration of conformity in accordance with Article 19 and Annex IV for the device model covered by the conformity assessment procedure,the documentation on the manufacturer's quality management system,a documented description of the procedures in place to fulfil the obligations arising from the quality management system and required under this Regulation and the undertaking by the manufacturer in question to apply those procedures,a description of the procedures in place to ensure that the quality management system remains adequate and effective, and the undertaking by the manufacturer to apply those procedures,the documentation on the manufacturer's post-market surveillance system and, where applicable, on the PMCF plan, and the procedures put in place to ensure compliance with the obligations resulting from the provisions on vigilance set out in Articles 87 to 92,a description of the procedures in place to keep up to date the post-market surveillance system, and, where applicable, the PMCF plan, and the procedures ensuring compliance with the obligations resulting from the provisions on vigilance set out in Articles 87 to 92, as well as the undertaking by the manufacturer to apply those procedures,documentation on the clinical evaluation plan, anda description of the procedures in place to keep up to date the clinical evaluation plan, taking into account the state of the art.2.2.Implementation of the quality management system shall ensure compliance with this Regulation. All the elements, requirements and provisions adopted by the manufacturer for its quality management system shall be documented in a systematic and orderly manner in the form of a quality manual and written policies and procedures such as quality programmes, quality plans and quality records.Moreover, the documentation to be submitted for the assessment of the quality management system shall include an adequate description of, in particular:(a)the manufacturer's quality objectives;(b)the organisation of the business and in particular:the organisational structures with the assignment of staff responsibilities in relation to critical procedures, the responsibilities of the managerial staff and their organisational authority,the methods of monitoring whether the operation of the quality management system is efficient and in particular the ability of that system to achieve the desired design and device quality, including control of devices which fail to conform,where the design, manufacture and/or final verification and testing of the devices, or parts of any of those processes, is carried out by another party, the methods of monitoring the efficient operation of the quality management system and in particular the type and extent of control applied to the other party, andwhere the manufacturer does not have a registered place of business in a Member State, the draft mandate for the designation of an authorised representative and a letter of intention from the authorised representative to accept the mandate;(c)the procedures and techniques for monitoring, verifying, validating and controlling the design of the devices and the corresponding documentation as well as the data and records arising from those procedures and techniques. Those procedures and techniques shall specifically cover:the strategy for regulatory compliance, including processes for identification of relevant legal requirements, qualification, classification, handling of equivalence, choice of and compliance with conformity assessment procedures,identification of applicable general safety and performance requirements and solutions to fulfil those requirements, taking applicable CS and, where opted for, harmonised standards or other adequate solutions into account,risk management as referred to in Section 3 of Annex I,the clinical evaluation, pursuant to Article 61 and Annex XIV, including post-market clinical follow-up,solutions for fulfilling the applicable specific requirements regarding design and construction, including appropriate pre-clinical evaluation, in particular the requirements of Chapter II of Annex I,solutions for fulfilling the applicable specific requirements regarding the information to be supplied with the device, in particular the requirements of Chapter III of Annex I,the device identification procedures drawn up and kept up to date from drawings, specifications or other relevant documents at every stage of manufacture, andmanagement of design or quality management system changes; and(d)the verification and quality assurance techniques at the manufacturing stage and in particular the processes and procedures which are to be used, particularly as regards sterilisation and the relevant documents; and(e)the appropriate tests and trials which are to be carried out before, during and after manufacture, the frequency with which they are to take place, and the test equipment to be used; it shall be possible to trace back adequately the calibration of that test equipment.In addition, the manufacturer shall grant the notified body access to the technical documentation referred to in Annexes II and III.2.3.AuditThe notified body shall audit the quality management system to determine whether it meets the requirements referred to in Section 2.2. Where the manufacturer uses a harmonised standard or CS related to a quality management system, the notified body shall assess conformity with those standards or CS. The notified body shall assume that a quality management system which satisfies the relevant harmonised standards or CS conforms to the requirements covered by those standards or CS, unless it duly substantiates not doing so.The audit team of the notified body shall include at least one member with past experience of assessments of the technology concerned in accordance with Sections 4.3. to 4.5. of Annex VII. In circumstances where such experience is not immediately obvious or applicable, the notified body shall provide a documented rationale for the composition of that team. The assessment procedure shall include an audit on the manufacturer's premises and, if appropriate, on the premises of the manufacturer's suppliers and/or subcontractors to verify the manufacturing and other relevant processes.Moreover, in the case of class IIa and class IIb devices, the quality management system assessment shall be accompanied by the assessment of technical documentation for devices selected on a representative basis as specified in Section 4. In choosing representative samples, the notified body shall take into account the published guidance developed by the MDCG pursuant to Article 105 and in particular the novelty of the technology, similarities in design, technology, manufacturing and sterilisation methods, the intended purpose and the results of any previous relevant assessments such as with regard to physical, chemical, biological or clinical properties, that have been carried out in accordance with this Regulation. The notified body in question shall document its rationale for the samples taken.If the quality management system conforms to the relevant provisions of this Regulation, the notified body shall issue an EU quality management system certificate. The notified body shall notify the manufacturer of its decision to issue the certificate. The decision shall contain the conclusions of the audit and a reasoned report.2.4.The manufacturer in question shall inform the notified body which approved the quality management system of any plan for substantial changes to the quality management system, or the device-range covered. The notified body shall assess the changes proposed, determine the need for additional audits and verify whether after those changes the quality management system still meets the requirements referred to in Section 2.2. It shall notify the manufacturer of its decision which shall contain the conclusions of the assessment, and where applicable, conclusions of additional audits. The approval of any substantial change to the quality management system or the device-range covered shall take the form of a supplement to the EU quality management system certificate.3.Surveillance assessment3.1.The aim of surveillance is to ensure that the manufacturer duly fulfils the obligations arising from the approved quality management system.3.2.The manufacturer shall give authorisation to the notified body to carry out all the necessary audits, including on-site audits, and supply it with all relevant information, in particular:the documentation on its quality management system,documentation on any findings and conclusions resulting from the application of the post-market surveillance plan, including the PMCF plan, for a representative sample of devices, and of the provisions on vigilance set out in Articles 87 to 92,the data stipulated in the part of the quality management system relating to design, such as the results of analyses, calculations, tests and the solutions adopted regarding the risk-management as referred to in Section 4 of Annex I, andthe data stipulated in the part of the quality management system relating to manufacture, such as quality control reports and test data, calibration data, and records on the qualifications of the personnel concerned.3.3.Notified bodies shall periodically, at least once every 12 months, carry out appropriate audits and assessments to make sure that the manufacturer in question applies the approved quality management system and the post-market surveillance plan. Those audits and assessments shall include audits on the premises of the manufacturer and, if appropriate, of the manufacturer's suppliers and/or subcontractors. At the time of such on-site audits, the notified body shall, where necessary, carry out or ask for tests in order to check that the quality management system is working properly. It shall provide the manufacturer with a surveillance audit report and, if a test has been carried out, with a test report.3.4.The notified body shall randomly perform at least once every five years unannounced audits on the site of the manufacturer and, where appropriate, of the manufacturer's suppliers and/or subcontractors, which may be combined with the periodic surveillance assessment referred to in Section 3.3. or be performed in addition to that surveillance assessment. The notified body shall establish a plan for such unannounced on-site audits but shall not disclose it to the manufacturer.Within the context of such unannounced on-site audits, the notified body shall test an adequate sample of the devices produced or an adequate sample from the manufacturing process to verify that the manufactured device is in conformity with the technical documentation, with the exception of the devices referred to in the second subparagraph of Article 52(8). Prior to unannounced on-site audits, the notified body shall specify the relevant sampling criteria and testing procedure.Instead of, or in addition to, sampling referred to in the second paragraph, the notified body shall take samples of devices from the market to verify that the manufactured device is in conformity with the technical documentation, with the exception of the devices referred to in the second subparagraph of Article 52(8). Prior to the sampling, the notified body in question shall specify the relevant sampling criteria and testing procedure.The notified body shall provide the manufacturer in question with an on-site audit report which shall include, if applicable, the result of the sample test.3.5.In the case of class IIa and class IIb devices, the surveillance assessment shall also include an assessment of the technical documentation as specified in Section 4 for the device or devices concerned on the basis of further representative samples chosen in accordance with the rationale documented by the notified body in accordance with the third paragraph of Section 2.3.In the case of class III devices, the surveillance assessment shall also include a test of the approved parts and/or materials that are essential for the integrity of the device, including, where appropriate, a check that the quantities of produced or purchased parts and/or materials correspond to the quantities of finished devices.3.6.The notified body shall ensure that the composition of the assessment team is such that there is sufficient experience with the evaluation of the devices, systems and processes concerned, continuous objectivity and neutrality; this shall include a rotation of the members of the assessment team at appropriate intervals. As a general rule, a lead auditor shall neither lead nor attend audits for more than three consecutive years in respect of the same manufacturer.3.7.If the notified body finds a divergence between the sample taken from the devices produced or from the market and the specifications laid down in the technical documentation or the approved design, it shall suspend or withdraw the relevant certificate or impose restrictions on it.CHAPTER IIASSESSMENT OF THE TECHNICAL DOCUMENTATION4.Assessment of the technical documentation applicable to class III devices and to the class IIb devices referred to in the second subparagraph of Article 52(4)4.1.In addition to the obligations laid down in Section 2, the manufacturer shall lodge with the notified body an application for assessment of the technical documentation relating to the device which it plans to place on the market or put into service and which is covered by the quality management system referred to in Section 2.4.2.The application shall describe the design, manufacture and performance of the device in question. It shall include the technical documentation as referred to in Annexes II and III.4.3.The notified body shall assess the technical documentation using staff with proven knowledge and experience regarding the technology concerned and its clinical application. The notified body may require the application to be completed by having further tests carried out or requesting further evidence to be provided to allow assessment of conformity with the relevant requirements of the Regulation. The notified body shall carry out adequate physical or laboratory tests in relation to the device or request the manufacturer to carry out such tests.4.4.The notified body shall review the clinical evidence presented by the manufacturer in the clinical evaluation report and the related clinical evaluation that was conducted. The notified body shall employ device reviewers with sufficient clinical expertise and, if necessary, use external clinical experts with direct and current experience relating to the device in question or the clinical condition in which it is utilised, for the purposes of that review.4.5.The notified body shall, in circumstances in which the clinical evidence is based partly or totally on data from devices which are claimed to be equivalent to the device under assessment, assess the suitability of using such data, taking into account factors such as new indications and innovation. The notified body shall clearly document its conclusions on the claimed equivalence, and on the relevance and adequacy of the data for demonstrating conformity. For any characteristic of the device claimed as innovative by the manufacturer or for new indications, the notified body shall assess to what extent specific claims are supported by specific pre-clinical and clinical data and risk analysis.4.6.The notified body shall verify that the clinical evidence and the clinical evaluation are adequate and shall verify the conclusions drawn by the manufacturer on the conformity with the relevant general safety and performance requirements. That verification shall include consideration of the adequacy of the benefit-risk determination, the risk management, the instructions for use, the user training and the manufacturer's post-market surveillance plan, and include a review of the need for, and the adequacy of, the PMCF plan proposed, where applicable.4.7.Based on its assessment of the clinical evidence, the notified body shall consider the clinical evaluation and the benefit-risk determination, and whether specific milestones need to be defined to allow the notified body to review updates to the clinical evidence that result from post-market surveillance and PMCF data.4.8.The notified body shall clearly document the outcome of its assessment in the clinical evaluation assessment report.4.9.The notified body shall provide the manufacturer with a report on the technical documentation assessment, including a clinical evaluation assessment report. If the device conforms to the relevant provisions of this Regulation, the notified body shall issue an EU technical documentation assessment certificate. The certificate shall contain the conclusions of the technical documentation assessment, the conditions of the certificate's validity, the data needed for identification of the approved design, and, where appropriate, a description of the intended purpose of the device.4.10.Changes to the approved device shall require approval from the notified body which issued the EU technical documentation assessment certificate where such changes could affect the safety and performance of the device or the conditions prescribed for use of the device. Where the manufacturer plans to introduce any of the above-mentioned changes it shall inform the notified body which issued the EU technical documentation assessment certificate thereof. The notified body shall assess the planned changes and decide whether the planned changes require a new conformity assessment in accordance with Article 52 or whether they could be addressed by means of a supplement to the EU technical documentation assessment certificate. In the latter case, the notified body shall assess the changes, notify the manufacturer of its decision and, where the changes are approved, provide it with a supplement to the EU technical documentation assessment certificate.5.Specific additional procedures5.1.Assessment procedure for certain class III and class IIb devices(a)For class III implantable devices, and for class IIb active devices intended to administer and/or remove a medicinal product as referred to in Section 6.4. of Annex VIII (Rule 12), the notified body shall, having verified the quality of clinical data supporting the clinical evaluation report of the manufacturer referred to in Article 61(12), prepare a clinical evaluation assessment report which sets out its conclusions concerning the clinical evidence provided by the manufacturer, in particular concerning the benefit-risk determination, the consistency of that evidence with the intended purpose, including the medical indication or indications and the PMCF plan referred to in Article 10(3) and Part B of Annex XIV.The notified body shall transmit its clinical evaluation assessment report, along with the manufacturer's clinical evaluation documentation, referred to in points (c) and (d) of Section 6.1 of Annex II, to the Commission.The Commission shall immediately transmit those documents to the relevant expert panel referred to in Article 106.(b)The notified body may be requested to present its conclusions as referred to in point (a) to the expert panel concerned.(c)The expert panel shall decide, under the supervision of the Commission, on the basis of all of the following criteria:(i)the novelty of the device or of the related clinical procedure involved, and the possible major clinical or health impact thereof;(ii)a significantly adverse change in the benefit-risk profile of a specific category or group of devices due to scientifically valid health concerns in respect of components or source material or in respect of the impact on health in the case of failure of the device;(iii)a significantly increased rate of serious incidents reported in accordance with Article 87 in respect of a specific category or group of devices,whether to provide a scientific opinion on the clinical evaluation assessment report of the notified body based on the clinical evidence provided by the manufacturer, in particular concerning the benefit-risk determination, the consistency of that evidence with the medical indication or indications and the PMCF plan. That scientific opinion shall be provided within a period of 60 days, starting on the day of receipt of the documents from the Commission as referred to in point (a). The reasons for the decision to provide a scientific opinion on the basis of the criteria in points (i), (ii) and (iii) shall be included in the scientific opinion. Where the information submitted is not sufficient for the expert panel to reach a conclusion, this shall be stated in the scientific opinion.(d)The expert panel may decide, under the supervision of the Commission, on the basis of the criteria laid down in point (c) not to provide a scientific opinion, in which case it shall inform the notified body as soon as possible and in any event within 21 days of receipt of the documents as referred to in point (a) from the Commission. The expert panel shall within that time limit provide the notified body and the Commission with the reasons for its decision, whereupon the notified body may proceed with the certification procedure of that device.(e)The expert panel shall within 21 days of receipt of the documents from the Commission notify the Commission, through Eudamed whether it intends to provide a scientific opinion, pursuant to point (c), or whether it intends not to provide a scientific opinion, pursuant to point (d).(f)Where no opinion has been delivered within a period of 60 days, the notified body may proceed with the certification procedure of the device in question.(g)The notified body shall give due consideration to the views expressed in the scientific opinion of the expert panel. Where the expert panel finds that the level of clinical evidence is not sufficient or otherwise gives rise to serious concerns about the benefit-risk determination, the consistency of that evidence with the intended purpose, including the medical indication(s), and with the PMCF plan, the notified body shall, if necessary, advise the manufacturer to restrict the intended purpose of the device to certain groups of patients or certain medical indications and/or to impose a limit on the duration of validity of the certificate, to undertake specific PMCF studies, to adapt the instructions for use or the summary of safety and performance, or to impose other restrictions in its conformity assessment report, as appropriate. The notified body shall provide a full justification where it has not followed the advice of the expert panel in its conformity assessment report and the Commission shall without prejudice to Article 109 make both the scientific opinion of the expert panel and the written justification provided by the notified body publicly available via Eudamed.(h)The Commission, after consultation with the Member States and relevant scientific experts shall provide guidance for expert panels for consistent interpretation of the criteria in point (c) before 26 May 2020.5.2.Procedure 2021.5.2.Procedure in the case of devices incorporating a medicinal substance(a)Where a device incorporates, as an integral part, a substance which, if used separately, may be considered to be a medicinal product within the meaning of point 2 of Article 1 of Directive 2001/83/EC, including a medicinal product derived from human blood or human plasma and that has an action ancillary to that of the device, the quality, safety and usefulness of the substance shall be verified by analogy with the methods specified in Annex I to Directive 2001/83/EC.(b)Before issuing an EU technical documentation assessment certificate, the notified body shall, having verified the usefulness of the substance as part of the device and taking account of the intended purpose of the device, seek a scientific opinion from one of the competent authorities designated by the Member States in accordance with Directive 2001/83/EC or from the EMA, either of which to be referred to in this Section as the medicinal products authority consulted depending on which has been consulted under this point, on the quality and safety of the substance including the benefit or risk of the incorporation of the substance into the device. Where the device incorporates a human blood or plasma derivative or a substance that, if used separately, may be considered to be a medicinal product falling exclusively within the scope of the Annex to Regulation (EC) No 726/2004, the notified body shall seek the opinion of the EMA.(c)When issuing its opinion, the medicinal products authority consulted shall take into account the manufacturing process and the data relating to the usefulness of incorporation of the substance into the device as determined by the notified body.(d)The medicinal products authority consulted shall provide its opinion to the notified body within 210 days of receipt of all the necessary documentation.(e)The scientific opinion of the medicinal products authority consulted, and any possible update of that opinion, shall be included in the documentation of the notified body concerning the device. The notified body shall give due consideration to the views expressed in the scientific opinion when making its decision. The notified body shall not deliver the certificate if the scientific opinion is unfavourable and shall convey its final decision to the medicinal products authority consulted.(f)Before any change is made with respect to an ancillary substance incorporated in a device, in particular related to its manufacturing process, the manufacturer shall inform the notified body of the changes. That notified body shall seek the opinion of the medicinal products authority consulted, in order to confirm that the quality and safety of the ancillary substance remain unchanged. The medicinal products authority consulted shall take into account the data relating to the usefulness of incorporation of the substance into the device as determined by the notified body, in order to ensure that the changes have no negative impact on the risk or benefit previously established concerning the incorporation of the substance into the device. The medicinal products authority consulted shall provide its opinion within 60 days after receipt of all the necessary documentation regarding the changes. The notified body shall not deliver the supplement to the EU technical documentation assessment certificate if the scientific opinion provided by the medicinal products authority consulted is unfavourable. The notified body shall convey its final decision to the medicinal products authority consulted.(g)Where the medicinal products authority consulted obtains information on the ancillary substance, which could have an impact on the risk or benefit previously established concerning the incorporation of the substance into the device, it shall advise the notified body as to whether this information has an impact on the risk or benefit previously established concerning the incorporation of the substance into the device. The notified body shall take that advice into account in reconsidering its assessment of the conformity assessment procedure.5.3.Procedure in the case of devices manufactured utilising, or incorporating, tissues or cells of human or animal origin, or their derivatives, that are non-viable or rendered non-viable5.3.1.Tissues or cells of human origin or their derivatives(a)For devices manufactured utilising derivatives of tissues or cells of human origin that are covered by this Regulation in accordance with point (g) of Article 1(6) and for devices that incorporate, as an integral part, tissues or cells of human origin, or their derivatives, covered by Directive 2004/23/EC, that have an action ancillary to that of the device, the notified body shall, prior to issuing an EU technical documentation assessment certificate, seek a scientific opinion from one of the competent authorities designated by the Member States in accordance with Directive 2004/23/EC (human tissues and cells competent authority) on the aspects relating to the donation, procurement and testing of tissues or cells of human origin or their derivatives. The notified body shall submit a summary of the preliminary conformity assessment which provides, among other things, information about the non-viability of the human tissues or cells in question, their donation, procurement and testing and the risk or benefit of the incorporation of the tissues or cells of human origin or their derivatives into the device.(b)Within 120 days of receipt of all the necessary documentation, the human tissues and cells competent authority shall provide to the notified body its opinion.(c)The scientific opinion of the human tissues and cells competent authority, and any possible update, shall be included in the documentation of the notified body concerning the device. The notified body shall give due consideration to the views expressed in the scientific opinion of the human tissues and cells competent authority when making its decision. The notified body shall not deliver the certificate if that scientific opinion is unfavourable. It shall convey its final decision to the human tissues and cells competent authority concerned.(d)Before any change is made with respect to non-viable tissues or cells of human origin or their derivatives incorporated in a device, in particular relating to their donation, testing or procurement, the manufacturer shall inform the notified body of the intended changes. The notified body shall consult the authority that was involved in the initial consultation, in order to confirm that the quality and safety of the tissues or cells of human origin or their derivatives incorporated in the device are maintained. The human tissues and cells competent authority concerned shall take into account the data relating to the usefulness of incorporation of the tissues or cells of human origin or their derivatives into the device as determined by the notified body, in order to ensure that the changes have no negative impact on the established benefit-risk ratio of the addition of the tissues or cells of human origin or their derivatives in the device. It shall provide its opinion within 60 days of receipt of all the necessary documentation regarding the intended changes. The notified body shall not deliver a supplement to the EU technical documentation assessment certificate if the scientific opinion is unfavourable and shall convey its final decision to the human tissues and cells competent authority concerned.5.3.2.Tissues or cells of animal origin or their derivativesIn the case of devices manufactured utilising animal tissue which is rendered non-viable or utilising non-viable products derived from animal tissue, as referred to in Regulation (EU) No 722/2012, the notified body shall apply the relevant requirements laid down in that Regulation.5.4.Procedure in the case of devices that are composed of substances or of combinations of substances that are absorbed by or locally dispersed in the human body(a)The quality and safety of devices that are composed of substances or of combinations of substances that are intended to be introduced into the human body via a body orifice or applied to the skin and that are absorbed by, or locally dispersed in, the human body, shall be verified where applicable and only in respect of the requirements not covered by this Regulation, in accordance with the relevant requirements laid down in Annex I to Directive 2001/83/EC for the evaluation of absorption, distribution, metabolism, excretion, local tolerance, toxicity, interaction with other devices, medicinal products or other substances and potential for adverse reactions.(b)In addition, for devices, or their products of metabolism, that are systemically absorbed by the human body in order to achieve their intended purpose, the notified body shall seek a scientific opinion from one of the competent authorities designated by the Member States in accordance with Directive 2001/83/EC or from the EMA, either of which to be referred to in this Section as the medicinal products authority consulted depending on which has been consulted under this point, on the compliance of the device with the relevant requirements laid down in Annex I to Directive 2001/83/EC.(c)The opinion of the medicinal products authority consulted shall be drawn up within 150 days of receipt of all the necessary documentation.(d)The scientific opinion of the medicinal products authority consulted, and any possible update, shall be included in the documentation of the notified body concerning the device. The notified body shall give due consideration to the views expressed in the scientific opinion when making its decision and shall convey its final decision to the medicinal products authority consulted.6.Batch verification in the case of devices incorporating, as an integral part, a medicinal substance which, if used separately, would be considered to be a medicinal product derived from human blood or human plasma as referred to in Article 1(8)Upon completing the manufacture of each batch of devices that incorporate, as an integral part, a medicinal substance which, if used separately, would be considered to be a medicinal product derived from human blood or human plasma as referred to in the first subparagraph of Article 1(8), the manufacturer shall inform the notified body of the release of the batch of devices and send it the official certificate concerning the release of the batch of human blood or plasma derivative used in the device, issued by a Member State laboratory or a laboratory designated for that purpose by a Member State in accordance with Article 114(2) of Directive 2001/83/EC.CHAPTER IIIADMINISTRATIVE PROVISIONS7.The manufacturer or, where the manufacturer does not have a registered place of business in a Member State, its authorised representative shall, for a period ending no sooner than 10 years, and in the case of implantable devices no sooner than 15 years, after the last device has been placed on the market, keep at the disposal of the competent authorities:the EU declaration of conformity,the documentation referred to in the fifth indent of Section 2.1 and in particular the data and records arising from the procedures referred to in point (c) of the second paragraph of Section 2.2,information on the changes referred to in Section 2.4,the documentation referred to in Section 4.2, andthe decisions and reports from the notified body as referred to in this Annex.8.Each Member State shall require that the documentation referred to in Section 7 is kept at the disposal of competent authorities for the period indicated in that Section in case a manufacturer, or its authorised representative, established within its territory goes bankrupt or ceases its business activity prior to the end of that period.

The full entry, with the citation mapping v1 = 02017R0745-20170505, v2 = 02017R0745-20200424, is committed at eu/32017R0745/CHANGELOG.md.